Image quality comparison of lower extremity CTA between CT routine reconstruction algorithms and deep learning reconstruction

Abstract

Background

To evaluate the image quality of lower extremity computed tomography angiography (CTA) with deep learning–based reconstruction (DLR) compared to model-based iterative reconstruction (MBIR), hybrid-iterative reconstruction (HIR), and filtered back projection (FBP).

Methods

Fifty patients (38 males, average age 59.8 ± 19.2 years) who underwent lower extremity CTA between January and May 2021 were included. Images were reconstructed with DLR, MBIR, HIR, and FBP. The standard deviation (SD), contrast-to-noise ratio (CNR), signal-to-noise ratio (SNR), noise power spectrum (NPS) curves, and the blur effect, were calculated. The subjective image quality was independently evaluated by two radiologists. The diagnostic accuracy of DLR, MBIR, HIR, and FBP reconstruction algorithms was calculated.

Results

The CNR and SNR were significantly higher in DLR images than in the other three reconstruction algorithms, and the SD was significantly lower in DLR images of the soft tissues. The noise magnitude was the lowest with DLR. The NPS average spatial frequency (fav) values were higher using DLR than HIR. For blur effect evaluation, DLR and FBP were similar for soft tissues and the popliteal artery, which was better than HIR and worse than MBIR. In the aorta and femoral arteries, the blur effect of DLR was worse than MBIR and FBP and better than HIR. The subjective image quality score of DLR was the highest. The sensitivity and specificity of the lower extremity CTA with DLR were the highest in the four reconstruction algorithms with 98.4% and 97.2%, respectively.

Conclusions

Compared to the other three reconstruction algorithms, DLR showed better objective and subjective image quality. The blur effect of the DLR was better than that of the HIR. The diagnostic accuracy of lower extremity CTA with DLR was the best among the four reconstruction algorithms.

Peer Review reports

Background

As one of the most common modalities for evaluating lower extremity PAD, runoff computed tomography angiography (CTA) has become a robust noninvasive imaging routine due to its short acquisition time, high spatial resolution, and increased anatomical coverage of the whole vascular tree [1, 2]. CTA truly revolutionized the diagnostic approach to vascular disorders of the lower extremities thanks to the high temporal resolution with fast execution time, the coverage of long vascular territories in a few seconds, the acquisition of isotropic datasets with the possibility of different reconstruction approaches, and accurate visualization of the anatomy and pathological abnormalities to allow appropriate treatment planning [3]. As an imaging modality based on X-rays, the balance between image quality and radiation dose is an important consideration [4]. Another problem in lower extremity CTA is the influence of image quality on small vessel visualization [5]. Thus, the CT reconstruction algorithm used can be an important factor for image quality.

The CT reconstruction algorithm was developed from filtered back projection (FBP) to hybrid iterative reconstruction (HIR) and model-based iterative reconstruction (MBIR); although it improved the image quality, diagnostic performance, and reduction of radiation dose in many clinical applications of CTA [6, 7], there are still some remaining unsatisfactory aspects in image texture, spatial resolution, and lesion detectability. Recently, a new deep learning-based CT reconstruction (DLR) algorithm (Advanced Intelligent Clear-IQ Engine [AiCE], Canon Medical Systems Corporation) has become commercially available. It was trained to differentiate noise from signals with the deep convolutional neural network (DCNN) by using MBIR patient datasets acquired with high tube current as golden standard clinical reference images [8]. The DLR reconstruction process begins in the raw data domain which is modified based on the detailed scanner model information, this raw data is then reconstructed to form a seed image for the DCNN. After that, the input image is analyzed by several network layers. The neurons in the network layer learn what features to look for based on the training data. DLR improves image quality with lower image noise, better sharpness, and more accurate diagnostic performance on the phantom study, cerebral CT, chest CT, coronary CTA, and abdominal CT [8,9,10,11,12,13]. AiCE is not the only available deep learning-based reconstruction algorithm, as there is also True Fidelity (GE Healthcare System), which has been applied in the lower extremity vessels study with better quantitative image quality for DLR in comparison with HIR [14]. Based on the promising results of previous studies, we speculate that its application to runoff CTA can improve image quality and diagnostic performance.

The purpose of this study was to evaluate the image quality of vessels and soft tissue in lower extremity runoff CTA of DLR compared to HIR, MBIR, and FBP, and to assess the diagnostic performance of runoff CTA using the DLR reconstruction algorithm.

Methods

Patient population

This was a retrospective study. The review board of our institution approved the study (HS-2427) and waived the requirement for informed consent due to its retrospective status.

Fifty-five consecutive patients who underwent runoff CTA with the indication given by the vascular surgeon were recorded, and 5 patients were excluded for metal implants. Fifty patients who had undergone runoff CTA between January 22, 2021, and May 27, 2021, were included.

CT acquisition parameters

All scans were performed on a 320-row-detector CT scanner (Aquilion One GENESIS Edition, Canon Medical Systems Corporation, Japan). The scan was performed from the distal abdominal aorta to the toes in a craniocaudal direction.

Ninety milliliters of iodinated contrast agent (370 mgI/mL, Iopromide, Ultravist, Bayer Health care, Germany) at a flow rate of 4 mL/second was administered intravenously. A region of interest (ROI) was placed on the distal abdominal aorta. When the predefined threshold of 150 HU was reached, 6 s later the scan automatically initiated.

The scanning parameters included slice collimation: 80 × 0.5 mm, rotation time: 0.75 s, pitch: 0.637, automatic modulation tube current: 40–200 mA (with noise index SD = 10), and tube voltage: 100 kVp.

The volumetric CT dose index (CTDIvol) and dose-length product (DLP) were automatically saved for each examination. The effective radiation dose was estimated using the DLP multiplied by the conversion factor of 0.0058 at 100 kV reported by Saltybaeva et al. [15].

Image reconstruction

All acquisitions were reconstructed using four reconstruction techniques: DLR (AiCE Body Sharp), HIR (AIDR 3D, kernel FC08), MBIR (FIRST Body Sharp), and FBP (kernel FC08). All reconstructions had a slice thickness of 1 mm and an interval of 0.8 mm. The field of view was set to 400.4 mm with a pixel matrix of 512 × 512.

Digital subtraction angiography (DSA)

Twenty-one patients received the DSA procedure within 30 days after CTA. DSA was performed on an angiographic system (Axiom Artis, Siemens Healthcare). The pelvic, thigh, and lower leg arteries of the symptomatic leg were examined with 5 ml iodinated contrast medium (320 mg I/ml iodixanol; Visipaque; GE Healthcare) per segment (aorta-iliac, femoral-popliteal and below the knees) using the stepping DSA. The posteroanterior projections were captured first and the left and right anterior oblique projections were added if the stenosis could not be assessed from the posteroanterior projection. Lesions were considered significant with visual luminal narrowing of ≥ 50%.

CT image analysis

Quantitative analysis

One senior radiologist (with 10 years of experience) drew the vessels’ region of interest (ROI) (aorta, common femoral artery, and popliteal artery), liver, and psoas muscle based on DLR images. One medical physicist retrieved the CT attenuation and standard deviation (SD) of the ROIs using ImageJ (V1.53a, National Institutes of Health, Bethesda, Maryland) and MATLAB (R2014) from the same locations of HIR, MBIR, and FBP images (Fig. 1). Then, the contrast-to-noise ratio (CNR) and signal-to-noise ratio (SNR) were calculated as follows: CNR = (HUtarget − HUfat)/SDfat and SNR = HUtarget/SDtarget, respectively.

Fig. 1
figure 1

Flow chart of quantitative analysis

Noise power spectrum (NPS) curves [16,17,18] of three different anatomical structures (liver, the aorta, and the psoas muscle) were derived from each patient’s CT images by using imQuest software (Version 7.1, Duke University, Durham NC). Measurements were performed on 10 consecutive slices to reduce the influences of noise and artifacts, and ROIs were carefully placed in the homogeneous region to guarantee the accuracy of the measured data. The normalized NPS curves were calculated by dividing the NPS value by the maximum value of the corresponding NPS.

In addition, the blur effect which represents objective sharpness was also calculated using MATLAB. The method was introduced in a study conducted by Chankue Park et al. The “blur metric” program (Do Quoc Bao (2022). Image Blur Metric (https://www.mathworks.com/matlabcentral/fileexchange/24676-image-blur-metric)) was used to evaluate the level of image sharpness as a non-reference quantitative value [19]. Before the blur metric value calculation, one radiologist drew eight rectangular ROIs including five anatomic structures and three whole-slice sections using ImageJ (Fig. 1). The five anatomic structures included the aorta, femoral artery, popliteal artery, liver, and muscle; the three whole-slice sections included the aorta, femoral artery, and popliteal artery. A scientist retrieved the blur metric data of the ROI area for each image (DLR, MBIR, Hybrid-IR, and FBP). The blur metric program quantifies sharpness by blurring the input image, and the variation between adjacent pixels is analyzed. First, intensity variations of CT attenuation between neighboring pixels are analyzed. Second, the low-pass-filtered image is used to measure the intensity variations between pixels. Third, a comparison between original and filtered images is made to evaluate the degree of blurring. The original image is believed to be sharp when there is a large difference in intensity variation. In the end, the blur metric value ranges from 0 to 1, with lower scores denoting sharper images and higher scores representing blurrier images.

Qualitative analysis

The subjective image quality based on image noise, subjective sharpness, and natural appearance was independently evaluated by two radiologists with 10 years and 3 years of experience, respectively, (Fig. 2) with the window setting of bone (width 2000 HU; level 350 HU) and soft tissue (width 350 HU; level 40 HU). A 5-point Likert scale was used, as detailed in Table 1.

Fig. 2
figure 2

Example slice demonstrating the impact of the different reconstruction algorithms for the runoff CTA of the liver with a window setting of the soft tissue and popliteal artery with a window setting of the bone

Table 1 Description of the categories of image quality characteristics

Statistical analysis

R software (version 3.6.1; http://www.R-project.org) was used to perform the statistical analysis. The Shapiro–Wilk test was used to determine the normality of the data distribution. For continuous variables with a normal distribution, one-way repeated analysis of variance (ANOVA) was used, and paired samples t-tests with Bonferroni correction were used for subsequent multiple comparisons. For data with nonnormal distribution, the Friedman test was applied, and the Wilcoxon signed rank test with Bonferroni correction was applied for multiple comparisons. p < 0.05 was considered statistically significant. Interreader reliability was assessed using a Kendall rank correlation coefficient. The diagnostic accuracy, such as sensitivity, specificity, and positive and negative predictive values, for the detection of significant stenosis, was calculated with DSA serving as a reference standard. The comparison of sensitivity and specificity of four reconstruction algorithms was calculated using McNemar’s test. p < 0.05 was considered statistically significant.

Results

Fifty patients were enrolled in the study (38 males, mean age 59.8 ± 19.2 years, range 19–91 years; mean body mass index (BMI) 24.1 ± 3.0 kg/m2, range 18.4–31.2 kg/m2). 60 slices of CT images were used for analysis in one patient, and 3000 slices of CT images in total were analyzed in this study. Among them, 21 patients (18 males, mean age 69.2 ± 9.6 years, range 49–91 years; mean BMI 24.1 ± 2.3 kg/m2, range 20.1–27.0 kg/m2) underwent clinically indicated DSA. The CTDIvol was 3.7 ± 0.8 mGy, and the DLP was 500.0 ± 119.5 mGycm. The radiation dose was 2.9 ± 0.7 mSv.

Quantitative analysis

The results of the quantitative assessment of the image qualities (CT, SD, SNR, and CNR) are summarized in Table 2.

Table 2 Quantitative analysis

Compared to MBIR and HIR, DLR showed higher CT attenuations of the aorta (all p < 0.001). DLR images of the femoral artery and popliteal artery showed the highest CT attenuation compared with other images (all p < 0.001). For the liver and psoas muscle, FBP showed significantly different CT attenuation with other reconstructions (all p < 0.001), while DLR showed similar CT attenuations with HIR and MBIR (all p > 0.001).

For image noise, DLR showed the lowest noise of arteries imaged (aorta, femoral artery, and popliteal artery), followed by MBIR, HIR, and FBP (all p < 0.001). There was no significant difference between the HIR and MBIR in the femoral and popliteal arteries (p = 0.27 and p = 0.42, respectively). No significant differences were observed between DLR and HIR, DLR and MBIR, or HIR and MBIR (p = 1, p = 0.22, p = 0.42, respectively). For the liver and psoas muscle, DLR also showed the lowest noise, and FBP had the highest noise (all p < 0.001).

The SNR of the DLR in all structures was significantly higher than those of the other three types of images (all p < 0.001), except in the popliteal artery, where there was no significant difference between the DLR and MBIR in the popliteal artery (p = 1). In the aorta, the trend of SNR from low to high was FBP, MBIR, and HIR and was significantly different with pairwise comparison (all p < 0.001). In the femoral artery and popliteal artery, FBP showed the lowest SNR, while there was no significant difference between MBIR and HIR (p = 1, p = 0.55, respectively). In addition, there was no significant difference between the DLR and MBIR in the popliteal artery (p = 1).

Similar to the SNR, the CNR of the DLR in all structures was significantly higher than those of the other three types of images (all p < 0.001). The trend of CNR in all structures from low to high was FBP, MBIR, and HIR in all structures and was significantly different from each other with pairwise comparison (all p < 0.001).

For all anatomical structures, the noise magnitude was the lowest with DLR, followed by HIR, MBIR, and FBP. The detailed results were shown in Table 3 and Fig. 3. The NPS average spatial frequency (fav) values were higher using DLR than HIR, while FBP demonstrated the highest fav values. The normalized NPS showed that for the aorta and psoas muscle, the noise on DLR and MBIR images consisted of more high-frequency components than HIR.

Table 3 Noise magnitude and average noise power spectrum (NPS) spatial frequency obtained for images reconstructed with FBP, HIR, MBIR, and DLR
Fig. 3
figure 3

Axial images with ROIs in red are placed over different anatomical structures (a, b, c). NPS curves (d, e, f) and normalized NPS curves (g, h, i) were obtained for all reconstruction algorithms. a, d, g demonstrate the liver, b, e, h demonstrate the psoas muscle, and c, f, i demonstrate the aorta

The results of the detailed blur effect are listed in Table 4. The blur effect of DLR images was significantly lower on the edges, textured areas, or homogeneous areas than that of HIR (all p < 0.016). In the liver and psoas muscle, DLR had a blur effect value similar to that of FBP (p = 1, p = 0.9696, respectively), which means that both had similar sharpness. For larger arteries (aorta and femoral artery), the trend of blur effect from low to high was FBP, MBIR, DLR, and HIR, while the blur effect value of FBP and MBIR images were significantly lower than those of DLR and HIR (all p < 0.001). For the popliteal artery, the blur effect of the DLR was almost the same as that of the FBP (p = 1). The trend of blur effect from low to high for the whole section image (aorta level, femoral artery level, and popliteal artery level) was FBP, MBIR, DLR, and HIR, all significantly different from each other with pairwise comparison (all p < 0.001). Except at the femoral artery level, the blur effect of the MBIR image was consistent with that of FBP (p = 0.17).

Table 4 Objective sharpness analysis

Qualitative analysis

The higher the score is given by the radiologists, the better the image quality. The results of the qualitative evaluation are summarized in Table 5. Interreader reliability was good (coefficient: 0.872, p < 0.001). The image noise, subjective sharpness, and natural appearance at different levels of lower extremity arteries as well as soft tissues were evaluated. The noise of the lower extremity arteries from low to moderate with the four reconstruction algorithms is DLR, HIR, MBIR, and FBP. There were significant differences between the pairwise comparisons except for the comparison between HIR and MBIR (p = 1.00). Regarding subjective sharpness, the trend from better to moderate image quality is DLR, MBIR, HIR, and FBP. There were no significant differences between the pairwise comparison of DLR and MBIR (p = 0.049) or MBIR and HIR (p = 0.13). For a natural appearance, DLR received the highest score, and FBP received the lowest score. There was no significant difference in the pairwise comparison between DLR, MBIR, HIR, and FBP except for DLR and FBP at the popliteal artery (p = 0.011). For soft tissues, including liver and psoas muscle, the trend from high to low image quality was DLR, HIR, MBIR, and FBP for noise with a significant difference; DLR, MBIR, HIR and FBP for subjective sharpness, there was no significant difference between MBIR and HIR (p = 1.00); and DLR, HIR. For a natural appearance, there was no significant difference between MBIR and FBP (p = 0.32, p = 0.89, respectively), HIR and FBP (p = 0.19, p = 0.36, respectively) for liver and psoas muscle, nor was DLR and HIR for psoas muscle (p = 0.058).

Table 5 Qualitative analysis

Diagnostic accuracy

Using DSA as a reference (Figs. 4, 5, 6), the sensitivity, specificity, positive predictive value (PPV), and negative predictive value (NPV) for the detection of stenosis by run-off CTA using the DLR, MBIR, HIR, and FBP reconstruction algorithm are shown in Table 6. The sensitivity from high to low was DLR, MBIR, HIR, and FBP (98.4%, 96.7%, 95.1%, and 93.4%, all p > 0.05), and the specificity from high to low was DLR, HIR, MBIR, and FBP (97.2%, 96.5%, 95.7%, and 93.7%, all p > 0.05).

Fig. 4
figure 4

A 67-year-old male with intermittent claudication of the right lower limb for 1 month. There was occlusion and significant stenosis in the proximal and middle superficial femoral arteries, respectively. An aneurysmal dilatation was found immediately after significant stenosis. DSA was performed within twenty-nine days. Panel a shows the DSA image and Panel b shows the maximum intensity projection CTA image

Fig. 5
figure 5

A 70-year-old female with left gangrenous toe for 3 weeks. There was an occlusion in the left proximal superficial femoral artery. The collateral artery in the distal patent superficial femoral artery was found. Panel a shows the volume rendering CTA image, panel b shows the maximum intensity projection CTA image, and panel c shows the DSA image

Fig. 6
figure 6

A 77-year-old male with intermittent claudication of both lower limbs for 3 months. There was significant stenosis in the left external iliac artery. Panel a shows the volume rendering CTA image, panel b shows the maximum intensity projection CTA image, and panel c shows the DSA image

Table 6 Diagnostic accuracy of the four reconstruction algorithms

Discussion

This study evaluated the image quality and diagnostic accuracy in lower extremity CTA reconstructed by DLR. For conventional quantitative image quality, the noise magnitude was lowest in DLR images and resulted in the highest SNR and CNR. There were more high-frequency components in the noise of DLR and MBIR images than that of HIR. As a complement to conventional objective image quality evaluation, the blur metric which evaluates the absolute quantitative value of the blur effect, showed that DLR preserves image sharpness compared to HIR. For subjective assessment, DLR received the highest score based on image noise, subjective sharpness, and natural appearance. The diagnostic performance of DLR was the best among the four reconstruction algorithms.

This is the first study on the application of the DLR algorithm “AiCE” in lower extremity CTA. There are several studies on other anatomic parts and the phantom. The results of this study are consistent with those of previous studies on cerebral non-contrast CT [12], coronary CTA [20], pulmonary CTA [21] abdominal CT [11, 22], and phantom [8]. The image noise, CNR, and/or SNR of DLR were improved compared with those of HIR and/or MBIR, while none of the previous studies compared DLR with FBP in vivo or discussed the blur metrics of sharpness.

Another DLR named “TrueFidelity” was explored by Park et al. on the application of lower extremity CTA with the discussion of blur metrics [14]. The CNR and SNR were higher in the DLR than in the low blending factor HIR, and the blur metrics increased as the strength of the DLR was improved. However, they did not evaluate MBIR and FBP at the same time.

In our study, DLR was compared with all three kinds of other reconstruction methods, MBIR, HIR, and FBP. In general, the objective image quality, including SD, CNR, and SNR, was best in DLR for both vascular lumen and soft tissues. Based on the subjective performance of the four reconstruction algorithms, radiologists ranked the highest score for DLR for both image noise, subjective sharpness, and natural appearance. This is consistent with other previous studies in which DLR achieved a better subjective score than the MBIR and/or HIR [11, 12, 20, 22]. The diagnostic performance of lower extremity runoff CTA with routine image reconstruction was excellent [1, 23], as was DLR in this study.

No-reference blur assessment in this study showed that DLR was better than Hybrid-IR with less blur effect and closer to MBIR. The possible reason might be that the training set of DLR was high-quality MBIR, which might preserve a similar sharpness property. MBIR had higher sharpness in this study, possibly due to the sharp kernel that we applied, which improved the image sharpness. FBP showed the best sharpness, which might be due to the highest image noise. Since the blur effect metric in this study was not focused on dedicated anatomical parts or edges, the location of the rectangular parts of the image was approximately selected to evaluate the general blur effect with a multiple-filtering process and comparison. This is not consistent with the subjective perception of image sharpness and is directly related to diagnostic performance. Even so, DLR showed better sharpness than HIR.

The application of lower extremity runoff CTA is increasingly popular; however, at the same time, the radiation dose related to the scan protocol and large scan range for the lower extremities is underexplored. In previous studies, the image quality of DLR was even better despite a dose reduction of 20%-30% compared to HIR images with a reference dose [20,21,22] in cardiac CTA, abdominal CT, and pulmonary CTA, and the reduction can even reach 50% in pediatric CT [24]. Although the radiation dose was not reduced in this study, the excellent image quality performance of DLR provides the potential application of DLR in low-dose lower extremity CTA.

Compared to HIR, the CNR of MBIR was lower in all locations, and SNR was lower in the aorta and soft tissues. The results were different from the phantom study [25] and cardiac CT [26]. The subjective score of MBIR was worse for noise and better for the structure's natural appearance compared with HIR. For MBIR, the objective score of noise was similar to the subjective noise result of this study, in which noise of MBIR was higher in the aorta, liver, and psoas muscle, and was consistent with Akagi et al. [22] and Bornet et al.’s study [11]. In previous studies, the natural appearance of cerebral [12] and abdominal [11] CT was slightly worse in MBIR, and the result was different from that in this study, which may be caused by different study anatomic areas and reconstruction parameters, such as kernel selection. To keep the parameters of the DLR in line with the MBIR, we used the body sharp kernel in this study. Although the so-called “sharp” kernels usually maintain high spatial resolution, they also elevate the image noise [27, 28].

There are several limitations to this study. First, we only compared DLR with MBIR, HIR, and FBP in routine radiation doses. According to previous studies, the image quality of DLR at a low radiation dose was comparable to that of HIR [21, 24, 29]. While there was no study on a low radiation dose of DLR in lower extremity run-off CTA, further study needs to be done. Second, “body sharp” was selected as the kernel of DLR and MBIR in this study, which might increase the noise magnitude of MBIR and lower its CNR in all locations, and SNR in the aorta and soft tissues. The influence of kernels in the comparison of DLR with other reconstruction algorithms should be discussed in further studies. Third, findings in this study may be valid no longer when the CTA is performed by newly proposed sparse-view CT imaging combined with a novel deep learning method [30,31,32], and further investigations are needed. Fourth, our study is limited by the small number of patients who underwent DSA, DLR diagnostic performance analysis sub-grouped by the location of the lesion was not performed.

Conclusions

In conclusion, compared to other routine reconstruction methods, DLR showed better objective image quality in terms of CNR, SNR, and image noise for both vessels and soft tissues, as did the subjective image quality score in terms of noise, subjective sharpness, and natural appearance. The diagnostic accuracy of lower extremity CTA with DLR was the best among four reconstruction algorithms.

Availability of data and materials

The datasets used and/or analyzed during the current study are available from the corresponding author upon reasonable request.

Abbreviations

PAD:

Peripheral artery disease

CVD:

Cardiovascular disease

CTA:

Computed tomography angiography

FBP:

Filtered back projection

HIR:

Hybrid iterative reconstruction

MBIR:

Model-based iterative reconstruction

DLR:

Deep learning-based CT reconstruction

AiCE:

Advanced intelligent clear-IQ engine

DCNN:

Deep convolutional neural network

CTDIvol:

CT dose index volume

DLP:

Dose-length product

DSA:

Digital subtraction angiography

ROI:

Region of interest

SD:

Standard deviation

CNR:

Contrast-to-noise ratio

SNR:

Signal-to-noise ratio

NPS:

Noise power spectrum NPS

DICOM:

Digital imaging and communications in medicine

BMI:

Body mass index

References

  1. Met R, Bipat S, Legemate DA, Reekers JA, Koelemay MJ. Diagnostic performance of computed tomography angiography in peripheral arterial disease: a systematic review and meta-analysis. JAMA. 2009;301(4):415–24.

    Article  CAS  PubMed  Google Scholar 

  2. Hiatt MD, Fleischmann D, Hellinger JC, Rubin GD. Angiographic imaging of the lower extremities with multidetector CT. Radiol Clin N Am. 2005;43(6):1119–27.

    Article  PubMed  Google Scholar 

  3. Cellina M, Gibelli D, Martinenghi C, Oliva G, Floridi C. CT angiography of lower extremities from anatomy to traumatic and nontraumatic lesions: a pictorial review. Emerg Radiol. 2020;27(4):441–50.

    Article  PubMed  Google Scholar 

  4. Lell MM, Kachelriess M. Recent and upcoming technological developments in computed tomography: high speed, low dose, deep learning. Multienergy Investig Radiol. 2020;55(1):8–19.

    Article  Google Scholar 

  5. Buls N, de Brucker Y, Aerden D, Devos H, Van Gompel G, Boonen PT, Nieboer K, Leiner T, de Mey J. Improving the diagnosis of peripheral arterial disease in below-the-knee arteries by adding time-resolved CT scan series to conventional run-off CT angiography. First experience with a 256-slice CT scanner. Eur J Radiol. 2019;110:136–41.

    Article  PubMed  Google Scholar 

  6. Wu R, Hori M, Onishi H, Nakamoto A, Fukui H, Ota T, Nishida T, Enchi Y, Satoh K, Tomiyama N. Effects of reconstruction technique on the quality of abdominal CT angiography: a comparison between forward projected model-based iterative reconstruction solution (FIRST) and conventional reconstruction methods. Eur J Radiol. 2018;106:100–5.

    Article  PubMed  Google Scholar 

  7. Zhou WZ, Gu JP, Yin XD, Wang LP, Lu LQ, Xu H, Wu QZ, Wang TX, Zhang D, Xu Q, et al. Low-voltage, high-pitch computerized tomography angiography of the infrarenal aorta and lower extremity vessels: assessment of radiation dose, image quality with hybrid iterative reconstruction, and efficacy of test injection using a monitoring scan at knee level. J Comput Assist Tomogr. 2016;40(3):402–8.

    Article  PubMed  Google Scholar 

  8. Higaki T, Nakamura Y, Zhou J, Yu Z, Nemoto T, Tatsugami F, Awai K. Deep learning reconstruction at CT: phantom study of the image characteristics. Acad Radiol. 2020;27(1):82–7.

    Article  PubMed  Google Scholar 

  9. Greffier J, Frandon J, Si-Mohamed S, Dabli D, Hamard A, Belaouni A, Akessoul P, Besse F, Guiu B, Beregi JP. Comparison of two deep learning image reconstruction algorithms in chest CT images: a task-based image quality assessment on phantom data. Diagn Interv Imaging. 2022;103(1):21–30.

    Article  Google Scholar 

  10. Yi Y, Xu C, Xu M, Yan J, Li Y-Y, Wang J, Yang S-J, Guo Y-B, Wang Y, Li Y-M. Diagnostic improvements of deep learning-based image reconstruction for assessing calcification-related obstructive coronary artery disease. Front Cardiovasc Med. 2021;8:758793.

    Article  PubMed  PubMed Central  Google Scholar 

  11. Bornet PA, Villani N, Gillet R, Germain E, Lombard C, Blum A, Gondim Teixeira PA. Clinical acceptance of deep learning reconstruction for abdominal CT imaging: objective and subjective image quality and low-contrast detectability assessment. Eur Radiol. 2022;32:3161.

    Article  PubMed  Google Scholar 

  12. Oostveen LJ, Meijer FJA, de Lange F, Smit EJ, Pegge SA, Steens SCA, van Amerongen MJ, Prokop M, Sechopoulos I. Deep learning-based reconstruction may improve non-contrast cerebral CT imaging compared to other current reconstruction algorithms. Eur Radiol. 2021;31:5498.

    Article  PubMed  PubMed Central  Google Scholar 

  13. Akagi M, Nakamura Y, Higaki T, Narita K, Honda Y, Zhou J, Yu Z, Akino N, Awai K. Deep learning reconstruction improves image quality of abdominal ultra-high-resolution CT. Eur Radiol. 2019;29(11):6163-71.

    Article  PubMed  Google Scholar 

  14. Park C, Choo KS, Jung Y, Jeong HS, Hwang JY, Yun MS. CT iterative vs deep learning reconstruction: comparison of noise and sharpness. Eur Radiol. 2021;31(5):3156–64.

    Article  PubMed  Google Scholar 

  15. Saltybaeva N, Jafari ME, Hupfer M, Kalender WA. Estimates of effective dose for CT scans of the lower extremities. Radiology. 2014;273(1):153–9.

    Article  PubMed  Google Scholar 

  16. Dalehaug I, Bolstad KN, Aadnevik D, Flataboe S, Pettersen HES: ADMIRE vs. SAFIRE: Objective comparison of CT reconstruction algorithms and their noise properties. arXiv preprint arXiv:170809616 2017.

  17. Abadi E, Harrawood B, Sharma S, Kapadia A, Segars WP, Samei E. DukeSim: a realistic, rapid, and scanner-specific simulation framework in computed tomography. IEEE Trans Med Imaging. 2018;38(6):1457–65.

    Article  PubMed  PubMed Central  Google Scholar 

  18. Khodajou-Chokami H, Hosseini SA, Ghorbanzadeh M, Mohammadi M: QCT: A measuring tool dedicated to the estimation of image parameters for quality assurance/quality control programs of CT scanners. In: 2020 IEEE International Symposium on Medical Measurements and Applications (MeMeA): 2020; 2020: 1–6.

  19. Crete F, Dolmiere T, Ladret P, Nicolas M: The blur effect: perception and estimation with a new no-reference perceptual blur metric. In: Human vision and electronic imaging XII: 2007: International Society for Optics and Photonics; 2007: 64920I.

  20. Tatsugami F, Higaki T, Nakamura Y, Yu Z, Zhou J, Lu Y, Fujioka C, Kitagawa T, Kihara Y, Iida M, et al. Deep learning-based image restoration algorithm for coronary CT angiography. Eur Radiol. 2019;29(10):5322–9.

    Article  PubMed  Google Scholar 

  21. Lenfant M, Chevallier O, Comby PO, Secco G, Haioun K, Ricolfi F, Lemogne B, Loffroy R. Deep learning versus iterative reconstruction for CT pulmonary angiography in the emergency setting: improved image quality and reduced radiation dose. Diagnostics (Basel). 2020;10(8):558.

    Article  PubMed  Google Scholar 

  22. Akagi M, Nakamura Y, Higaki T, Narita K, Honda Y, Zhou J, Yu Z, Akino N, Awai K. Deep learning reconstruction improves image quality of abdominal ultra-high-resolution CT. Eur Radiol. 2019;29(11):6163–71.

    Article  PubMed  Google Scholar 

  23. Catalano C, Fraioli F, Laghi A, Napoli A, Bezzi M, Pediconi F, Danti M, Nofroni I, Passariello R. Infrarenal aortic and lower-extremity arterial disease: diagnostic performance of multi-detector row CT angiography. Radiology. 2004;231(2):555–63.

    Article  PubMed  Google Scholar 

  24. Brady SL, Trout AT, Somasundaram E, Anton CG, Li Y, Dillman JR. Improving image quality and reducing radiation dose for pediatric CT by using deep learning reconstruction. Radiology. 2021;298(1):180–8.

    Article  PubMed  Google Scholar 

  25. Laurent G, Villani N, Hossu G, Rauch A, Noel A, Blum A, Gondim Teixeira PA. Full model-based iterative reconstruction (MBIR) in abdominal CT increases objective image quality, but decreases subjective acceptance. Eur Radiol. 2019;29(8):4016–25.

    Article  PubMed  Google Scholar 

  26. Maeda E, Tomizawa N, Kanno S, Yasaka K, Kubo T, Ino K, Torigoe R, Ohtomo K. The feasibility of forward-projected model-based iterative reconstruction SoluTion (FIRST) for coronary 320-row computed tomography angiography: a pilot study. J Cardiovasc Comput Tomogr. 2017;11(1):40–5.

    Article  PubMed  Google Scholar 

  27. Eldevik K, Nordhøy W, Skretting A. Relationship between sharpness and noise in CT images reconstructed with different kernels. Radiat Prot Dosimetry. 2010;139(1–3):430–3.

    Article  CAS  PubMed  Google Scholar 

  28. Zhou Q, Jiang B, Dong F, Huang P, Liu H, Zhang M. Computed tomography coronary stent imaging with iterative reconstruction: a trade-off study between medium kernel and sharp kernel. J Comput Assist Tomogr. 2014;38(4):604–12.

    Article  PubMed  Google Scholar 

  29. Tamura A, Mukaida E, Ota Y, Kamata M, Abe S, Yoshioka K. Superior objective and subjective image quality of deep learning reconstruction for low-dose abdominal CT imaging in comparison with model-based iterative reconstruction and filtered back projection. Br J Radiol. 2021;94(1123):20201357.

    Article  PubMed  PubMed Central  Google Scholar 

  30. Khodajou-Chokami H, Hosseini SA, Ay MR. A deep learning method for high-quality ultra-fast CT image reconstruction from sparsely sampled projections. Nucl Instrum Methods Phys Res Sect A. 2022;1029: 166428.

    Article  CAS  Google Scholar 

  31. Khodajou-Chokami H, Hosseini S, Ay M. PARS-NET: a novel deep learning framework using parallel residual conventional neural networks for sparse-view CT reconstruction. J Instrum. 2022;17(02):P02011.

    Article  Google Scholar 

  32. Xia W, Lu Z, Huang Y, Shi Z, Liu Y, Chen H, Chen Y, Zhou J, Zhang Y. Magic: manifold and graph integrative convolutional network for low-dose ct reconstruction. IEEE Trans Med Imaging. 2021;40(12):3459–72.

    Article  PubMed  Google Scholar 

Download references

Acknowledgements

Not applicable

Funding

This research was supported by CAMS (Chinese Academy of Medical Sciences) innovation fund for medical sciences (CIFMS 2021-I2M-C&T-B-022); Peking Union Medical College Hospital Youth funding (pumch201911061); 2021 Beijing key clinical specialty construction in the interpretation of data and writing.

Author information

Authors and Affiliations

Authors

Contributions

DZ analyzed the data and drafted the manuscript. CM analyzed the data. XZ participated in image acquisition and reconstruction. JY developed the imaging processing code. MX participated in image reconstruction. YW participated in image acquisition. YW participated in the research design. HX participated research design. YC participated in research design, and image acquisition and revised the manuscript. ZJ designed the research and revised the manuscript. All authors read and approved the final manuscript.

Corresponding authors

Correspondence to Yuexin Chen or Zhengyu Jin.

Ethics declarations

Ethics approval and consent to participate

This study was approved by the Institutional Review Board of Peking Union Medical College Hospital, Chinese Academy of Medical Sciences and Peking Union Medical College (HS-2427), and the requirement for informed consent was waived by the Institutional Review Board of Peking Union Medical College Hospital, Chinese Academy of Medical Sciences and Peking Union Medical College, due to its retrospective status. All methods were performed by the relevant guidelines and regulations (for example the Declarations of Helsinki).

Consent for publication

Not applicable.

Competing interests

The authors declare that they have no competing interests.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Zhang, D., Mu, C., Zhang, X. et al. Image quality comparison of lower extremity CTA between CT routine reconstruction algorithms and deep learning reconstruction. BMC Med Imaging 23, 33 (2023). https://doi.org/10.1186/s12880-023-00988-6

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s12880-023-00988-6

Keywords