ROCKETSHIP: a flexible and modular software tool for the planning, processing and analysis of dynamic MRI studies

  • Samuel R. Barnes1,
  • Thomas S. C. Ng1, 2Email author,
  • Naomi Santa-Maria1,
  • Axel Montagne3,
  • Berislav V. Zlokovic3 and
  • Russell E. Jacobs1
Contributed equally
BMC Medical Imaging201515:19

https://doi.org/10.1186/s12880-015-0062-3

Received: 4 January 2015

Accepted: 29 May 2015

Published: 16 June 2015

Abstract

Background

Dynamic contrast-enhanced magnetic resonance imaging (DCE-MRI) is a promising technique to characterize pathology and evaluate treatment response. However, analysis of DCE-MRI data is complex and benefits from concurrent analysis of multiple kinetic models and parameters. Few software tools are currently available that specifically focuses on DCE-MRI analysis with multiple kinetic models. Here, we developed ROCKETSHIP, an open-source, flexible and modular software for DCE-MRI analysis. ROCKETSHIP incorporates analyses with multiple kinetic models, including data-driven nested model analysis.

Results

ROCKETSHIP was implemented using the MATLAB programming language. Robustness of the software to provide reliable fits using multiple kinetic models is demonstrated using simulated data. Simulations also demonstrate the utility of the data-driven nested model analysis. Applicability of ROCKETSHIP for both preclinical and clinical studies is shown using DCE-MRI studies of the human brain and a murine tumor model.

Conclusion

A DCE-MRI software suite was implemented and tested using simulations. Its applicability to both preclinical and clinical datasets is shown. ROCKETSHIP was designed to be easily accessible for the beginner, but flexible enough for changes or additions to be made by the advanced user as well. The availability of a flexible analysis tool will aid future studies using DCE-MRI.

A public release of ROCKETSHIP is available at https://github.com/petmri/ROCKETSHIP.

Keywords

DCE-MRIImagingADCParametric mappingData-driven kinetic modellingMATLABMRINested model

Background

The utility of dynamic magnetic resonance imaging (MRI) studies to diagnose diseases, characterize their progression, and evaluate treatment response is a topic of vigorous research. In particular, physiological information garnered from techniques such as diffusion MRI, blood oxygen-level dependent (BOLD) MRI, iron-oxide imaging, dynamic susceptibility MRI (DSC-MRI) and dynamic contrast-enhanced MRI (DCE-MRI) [1] are being explored as imaging biomarkers in virtually all aspects of medicine, especially in oncology and neurological diseases.

DCE-MRI involves the rapid serial acquisition of T1-weighted images before, during and after intravenous injection of a contrast agent (CA). The physiological properties of the tissue of interest are inferred by analyzing the image signal change kinetics induced by the CA within the tissue region of interest (ROI). Given the inherent leakiness of tumor blood vessels (which enhances the signal observed at the tumor site due to more CA leakage) [2], the majority of DCE-MRI studies have focused on oncology applications [3]. Studies have especially focused on using DCE-MRI to characterize tumor phenotypes or to evaluate the response of tumors to therapy [4]. The use of DCE-MRI has also been explored in other fields such as obstetrics [5] and the neurosciences [6, 7].

Although several individual studies have demonstrated the promise of DCE-MRI as an imaging biomarker, widespread clinical adoption of this technique remains limited. This is in part because implementation and standardization of DCE-MRI is challenging. Differences in hardware such as the MRI field strength, coil setups and sequence types available at different institutions can affect the quality of the images acquired and constrain the choice of certain parameters such as the volume of interest, spatial and temporal resolution. The choice of contrast agent, each with unique pharmacokinetic properties and thus behavior in vivo [8], also vary greatly between studies. These factors can significantly affect the quality of the parametric maps generated from imaging data [9].

Widespread variability also exists in the processing and analysis of DCE-MRI data. Both semi-quantitative [10, 11] and quantitative parameters [12] are used to analyze the acquired signal intensity time curves. Several studies have demonstrated the utility of either type of parameters to characterize tumor tissues or monitor treatment response [13, 14]. However, challenges remain in determining the specificity of different parameters to the underlying physiology, development of robust quantification techniques for quantitative model fitting (such as the assignment of a robust arterial input function [15] or T1 measurement of tissues [16]), and the selection of appropriate equations or models to fit the acquired data [17].

Recent studies have sought to address and understand these issues. For example, both Cramer et al. [18] and Montagne et al. [6] demonstrated that the Patlak method may accurately estimate the vascular permeability in the intact or near intact blood brain barrier (BBB); a challenge given the low permeability values being considered (Ktrans < 0.3 mL/100 g/min) [1921]. In particular, Montagne et al. were able to demonstrate significant differences in permeability within grey and white matter between human subjects with no cognitive impairment, mild cognitive impairment, and with age [6] using Patlak analysis. Two-compartment models have been shown to be more appropriate at higher permeability [18, 22]. Ewing et al. have explored the use of nested model-selection to determine the appropriate physiological models to fit both preclinical and clinical DCE-MRI data [23, 24], while Jackson et al. recently showed good correlation of semi-quantitative metrics with parameters derived from an extended Tofts model in patients with Type 2 Neurofibromatosis [13]. Importantly, these studies suggest that robust analysis of DCE-MRI data may be disease and organism specific and that a data-driven approach that considers a library of processing and analytical methods may be necessary and appropriate for DCE-MRI analysis [12].

Exploration of these different types of DCE-MRI processing and analyses would be facilitated with standardized software. Software standardization can reduce biases and errors that may arise from the type of processing algorithm used for fitting and enable more consistent comparisons between studies. Moreover, availability of a flexible, easily modifiable and extensible software suite allows the researcher/clinician to examine how various factors discussed above may apply to their specific DCE-MRI study.

To date, numerous software packages are available for DCE-MRI analysis. These include general purpose pharmacokinetic analysis packages such as PMOD (pmod.com), WinSAAM [25], JPKD (pkpd.kmu.edu.tw/jpkd) or SAAMII [26]. These packages are often complex to use, are only commercially available and/or require significant pre-processing of the data to adjust for DCE-MRI usage. DCE-MRI specific software packages also exist, including commercially available packages licensed for clinical use [27]. While these are often user friendly and directly integrated with PACS databases, details about each package’s implementation and certain options involved in the fitting algorithms may be unavailable. This can lead to a wide variability in output results [27]. Other packages available include PMI (https://sites.google.com/site/plaresmedima/), pydcemri (github.com/davidssmith/pydcemri), DCEMRI.jl [28], Jim (Xinapse Systems Ltd), BioMap (maldi-msi.org), PermGUI/PCT [29], Toppcat [30], DCEMRIS4 [31], dcetool (http://thedcetool.com/), dcemri (http://dcemri.sourceforge.net/), DATforDCEMRI [32], DCE@UrLAB [33] and UMMPerfusion [34]. Table 1 provides a summary of the capabilities of these packages. While geared for DCE-MRI, most of them are limited in the range of fitting methods, models and parameters that can be analyzed. A number of the packages (developed using IDL and C) are built around a graphical user interface (GUI) that needs to be recompiled each time a new functionality is implemented or a setting changed, which can make module development and individualized changes non-trivial for a novice user. None of the currently available packages in Table 1 enable recently developed data-driven methods, such as nested-model selection, that analyze the appropriateness of the models and parameters being used and generated [23, 24, 3537]. To circumvent these constraints, groups often develop in-house software for their studies [3841], with limited availability for other users. This hinders adoption of these methods for widespread use and testing.
Table 1

Comparison of existing DCE-MRI packages*

Software package

Language

Operating system

License

Models included

Fitting

Other MRI-relevant features

Input/output

Location

dcemriS4

R

Linux, Windows, Mac OS

BSD

- Standard Kelty- Single-compartment model- Extended Kety

- Non-linear least squares- Bayesian estimation

- motion correction and co-registration - B1 mapping - T1 mapping - AIF fitting - DWI fitting - Pixel processing - Job report for later retreival - Access to R functions

- DICOM - NIFTI - Raw data

http://dcemri.sourceforge.net

dcetool

C, Plugin for ClearCanvas

Windows

Proprietary

- Tofts - Adiabatic tissue homogeniety - Thorwarth- Extended Tofts - Semi-quantitative metrics (Slope/AUC)

Proprietary

 

- DICOM - Raw data

http://thedcetool.com/

PMI

IDL

Windows

GNU GPL

- Uptake models - Steady-state - Patlak - Model-free deconvolution - Tofts- Extended Tofts - 2CXM - 2C filtration model for kidney - Dual-inlet models for Liver - Semi-quantitative metrics (Slope/Signal enhancement)

- Non-linear Least squares - Truncated singular value decomposition

- ROI and pixel processing - AIF/time series visualization/editing - Access to IDL functions

- DICOM - Raw data

https://sites.google.com/site/plaresmedima/

UMMPerfusion

C, OsiriX plugin

Mac OS

BSD

- Model free deconvolution

- Truncated singular value decomposition

- ROI and pixel processing - Job report for later retreival - AIF/time series visualization/editing

- DICOM

http://ikrsrv1.medma.uni-heidelberg.de/redmine/projects/ummperfusion

Pydcemri

Python

Linux, Windows, Mac OS

GNU GPL

- Tofts - Extended Tofts

- Non-linear Least squares

- Access to python functions

- Raw data

https://github.com/welcheb/pydcemri

DCEMRI.jl

Julia

Linux, Windows, Mac OS

MIT

- Tofts - Extended Tofts - Plasma only

- Non-linear Least squares

- ROI and pixel processing - T1 mapping - Batch processing - Access to Julia functions

- Matlab data

https://github.com/davidssmith/DCEMRI.jl

DCE@UrLAB

IDL

Windows

BSD

- Tofts - Hoffmann - Larsson - Fast exchange limit reference region

- Non-linear Least squares

- ROI and pixel processing - Access to IDL functions

- DICOM - Bruker - Raw data

http://www2.die.upm.es/im/archives/DCEurLAB/

DATforDCEMRI

R

 

Creative Commons

- Tofts - Semi-quantitative metrics (AUC, MRT - mean residence time)

- Numerical deconvolution

- Pixel processing - Access to R functions

- R readable data formats

https://github.com/cran/DATforDCEMRI

TOPPCAT

Javascript, ImageJ plugin

Linux, Windows, Mac OS

BSD

- Patlak

- Non-linear Least squares

- T1 mapping - Access to ImageJ functions

- ImageJ readable data formats

https://dblab.duhs.duke.edu/modules/dblabs_topcat/

Jim

Java

Linux, Windows, Mac OS

Proprietary

- Tofts - Extended Tofts - One compartment - Fermi - 2CXM - Semi-quantitative metrics (AUC)

Proprietary

- ROI and pixel processing

- DICOM - Analyze - Bruker - Commercial formats - Raw data

http://www.xinapse.com/Manual/index.html

PermGUI

Matlab

Windows

Creative Commons

- Patlak

 

- T1 mapping - ROI and pixel processing

- DICOM - Analyze - NIFTI

http://www.quantilyze.com/permgui/

BioMap

IDL

Linux, Windows, Mac OS

Proprietary

- Extended Tofts

- Non-linear Least squares

- T1 fitting

- DICOM - Analyze - TIF/PNG

http://www.maldi-msi.org/

ROCKETSHIP

Matlab

Linux, Windows, Mac OS

GNU GPL

- Tofts - Extended Tofts - Fast exchange regime (FXR) - 2CXM - Tissue uptake - Nested-model selection - Patlak - Semi-quantitative metrics (AUC)

- Non-linear Least squares

- T1 mapping - ROI and pixel processing - AIF fitting/import - DWI fitting - Job report for later retreival - AIF/time series visualization/editing - Batch processing - Access to Matlab functions - Model fit comparisons with statistical metrics - Drift correction

- DICOM - Analyze - NIFTI - Raw data - Matlab data

https://github.com/petmri/ROCKETSHIP

*Excludes commerical packages for clinical use, AUC: area under curve, DWI: diffusion-weighted imaging

Here, we describe the development of a software suite, ROCKETSHIP (available for download at https://github.com/petmri/ROCKETSHIP), which is implemented in the widely available MATLAB software environment. ROCKETSHIP is designed to be modular, easily extensible and modifiable, and provides tools to process multiple types of parametric MRI datasets. In particular, ROCKETSHIP focuses on the processing and analysis of DCE-MRI data from both human and animal studies, including T1 map generation and arterial input function (AIF) processing and analyses with a library of common pharmacokinetic modeling methods as well as recently developed statistically driven nested model methods. Implementation of ROCKETSHIP in a commonly used language (MATLAB) for image analysis should allow both novice and advance users to utilize and extend the software for their specific DCE-MRI applications.

We test the robustness of this software tool using simulations and demonstrate the utility of ROCKETSHIP to analyze preclinical and clinical DCE-MRI data from different disease models.

Implementation

All components of ROCKETSHIP were implemented in the MATLAB environment, specifically MATLAB release 2014a. Detailed description of each component can be found in source code comments and documentation that is accessible at the project’s github page (https://github.com/petmri/ROCKETSHIP). In this section, we describe the general architecture of the software suite, and highlight the key aspects of the software and the algorithms currently implemented in the software.

Architecture design and GUI description

A simplified schematic outlining the design of ROCKETSHIP is shown in Fig. 1. The software is divided into discrete modules to separate processing and analysis components of the pipeline. Initially, a fitting module GUI is available for the user to generate voxel-wide parametric maps directly from imaging data. Common MRI parameters such as T2 (and T2*) and T1 relaxation times, and apparent diffusion coefficients (ADC) from diffusion MRI data can be estimated with different fitting methods. A summary of the equations used is in Appendix A. Additional user specified fitting functions can be easily implemented without major modifications to the GUI.
Fig. 1

Design outline of ROCKETSHIP. The software suite consists of a fitting module to generate T1, T2/T2* and ADC maps, and DCE-MRI module with sub-modules for each stage of DCE-MRI data processing and analysis

ROCKETSHIP is particularly focused on the processing and analysis of DCE-MRI data. Within the DCE-MRI module (Fig. 2), sub-modules are implemented that focus on:
Fig. 2

DCE-MRI processing module GUI. GUI modules reflect the schematic outlined in Fig. 2. The “root” DCE module is shown on the left, which launches each sub-module in the pipeline. a defines the sub-module that converts raw image data to concentration time curves. The data are passed to the next sub-module, which allows temporal truncation of the dynamic data and fitting or importing of the AIF (b). DCE-MRI maps are derived using the next sub-module (c). Models can be generated in real time, or the user input can be saved as a data structure job to be run in batch later. Options are provided to perform voxel-by-voxel fits as well as defined ROIs. Raw data curves can be fitted as is, or after being passed through a time smoothing filter. Finally, goodness-of-fit analysis of the fits can be performed with the final sub-module (d)

  1. 1)

    Preparation of the dynamic datasets for DCE-MRI analysis. This involves options for intensity data to concentration versus time curves conversion using T1 maps, selection of the AIF or reference region (RR) and the ROI, noise filtering and signal intensity drift correction over the dynamic time course (Fig. 2a). A summary of how these are implemented is provided in Appendix B.

     
  2. 2)

    AIF or RR processing with temporal truncation. This sub-module allows the appropriate AIF/RR to be selected. The raw or fitted AIF (currently limited to bi-exponential fitting) from an individual dataset can be used (Fig. 2b). Alternatively, multiple AIFs/RRs can be combined to form a population-averaged AIF/RR. A summary of the implementation is provided in Appendix C. As time duration and temporal resolution can be critical for analysis, options to modify these factors are included.

     
  3. 3)

    Derivation of DCE-MRI parametric maps (Fig. 2c). Multiple pharmacokinetic models and semi-quantitative metrics can be fitted using ROCKETSHIP. In the current implementation, the Tofts, extended Tofts, Patlak, shutter-speed, two-compartment exchange, tissue uptake models and the area-under-curve (AUC) metric can be calculated (Table 1). Details about these models are summarized in Appendix D. Apart from these stand-alone models, a nested model method is also implemented.

     
Nesting is based on the hierarchy inherent within the two-compartment exchange model; the algorithm was developed, tested and described in detail by Ewing et al. [12, 23, 24, 37]. First, DCE-MRI data are fitted to the zero-order model, which describes the case where there is insufficient evidence of filling of the vasculature with CA [23, 37]. Next, the initial fit is compared to the data fitting using the one parameter steady-state model:
$$ C(t)=v\ {C}_a(t) $$
(1)

Where C(t) is the concentration of the CA in the voxel of interest at time t, v is the volume fraction of the indicator distribution space and Ca(t) is the concentration of the CA in the arterial compartment at time t. Here, v can represent the plasma volume fraction (vp) in a voxel that is highly vascularized with no vascular/extravascular exchange, the extravascular volume fraction (ve) in a weakly vascularized voxel, or v = vp + ve in a fast-exchange scenario. A statistical F-test is used to evaluate the likelihood that an observed improvement of fit to the data, using a model with higher number of parameters, warrants the use of additional parameters [42]. A p-value <0.05 is used as a criterion to increase the number of parameters in the model fit.

In the current implementation, the Patlak model represents the two-parameter form of the two-compartment model with the extended Tofts model with three parameters being at the top of the nested hierarchy [23, 24, 37].

Options to smooth the dynamic signal time course and to fit specific ROIs versus voxel-by-voxel fitting are also available in the DCE-MRI sub-module (Fig. 2c).
  1. 4)

    While models belonging to the same hierarchy can be folded into the nested model fitting option, it may be desirable to compare non-nested models or make comparisons with different statistical tests. The fitting analysis sub-module allows for visual and statistical assessment of goodness-of-fit (Fig. 2d). Model fits with 95 % prediction bounds of the fit are shown graphically along with the raw data for each voxel/ROI. Fits between models can be compared using the F-test [42, 43], fraction of modeled information (FMI) and fraction of residual information (FRI) [35], and the Akaike information criterion [7, 43]. These results can be exported to an Excel (office.microsoft.com/en-us/excel) spreadsheet for offline analysis.

     

Estimation of model parameters

All curve fitting functions in ROCKETSHIP are implemented using MATLAB’s Curve Fitting Toolbox. T1, T2 and ADC signal equations can be linearized and fitted with linear regression (See Appendix A). Alternatively, these parameters can be directly fitted with non-linear methods. ROCKETSHIP uses the trust region algorithm provided in the Curve Fitting Toolbox to perform non-linear least squares regression. For T1, T2 and ADC regression, the parameters are hard-coded to have non-negative value constraints. Robust curve fitting is dependent on appropriate starting parameters for the fitting routine [44]. To facilitate this process, a preferences text file defining parameter constraints and convergence criteria, such as fitting tolerances and maximum numerical of iterations, is provided to allow easy editing of these variables. This text file is read by ROCKETSHIP when AIF and model fitting sub-modules are run.

During testing of ROCKETSHIP, it was found that Ktrans fitting often converged to local minima instead of the desired global minimum solution. To address this, Ktrans was fitted using multiple starting values with the fit value converging with the lowest residual used as the final value. Other variables were less sensitive to the starting position and thus a single initial value was used to fit each of those variables.

Voxel-wide fitting is performed in parallel using functions provided by MATLAB’s Parallel Computing Toolbox for increased performance.

Inputs and outputs

To enable processing of both preclinical and clinical data, the software can read and write both NIFTI/Analyze and DICOM formats. ROIs can be input to ROCKETSHIP as image files drawn in other image visualization programs such as MRIcro (www.mricro.com) or as. roi files generated from ImageJ (imagej.nih.gov/ij).

Datasets can be processed through the pipeline individually. A batch mode is also available to generate multiple parametric maps in the fitting and DCE-MRI model fitting modules.

Text files logging outputs from each module are stored and emailed to the user upon task completion. Data from each module in the processing pipeline are also stored in files to facilitate debugging, transfer between users and allows processing pauses between modules.

Results

Validation of ROCKETSHIP software with simulation

Precision and accuracy of individual kinetic model fitting

The robustness of ROCKETSHIP was evaluated with simulations. Simulated datasets containing two (Tofts, Patlak models), three (Extended Tofts) and four parameters (two-compartment exchange model, 2CXM) were generated using MATLAB. The ability of ROCKETSHIP to recover the appropriate DCE-MRI parameter values from simulated data at different SNR and time resolutions was evaluated. To achieve this, curves were generated using specific parameter values defined in Table 2. For each parameter permutation Rician noise at different SNRs was added to the signal intensity versus time curves and fitted. This process was repeated with new noise 100 times in a Monte-Carlo manner to estimate the accuracy and precision of the fits. Resultant curves were fitted with ROCKETSHIP using the same model that was used to generate the curves. The population AIF published by Parker et al. was used in all simulations [15]. Default fitting options were used. The accuracy of the fitted parameters were evaluated with the concordance correlation coefficient (CCC) [45].
Table 2

Fitting parameters for simulation studies

Generating model

Fitting model

Acquistion duration (min)

Time resolution (s)

SNR

Ktrans (1/min)

ve

vp

Fp (1/min)

τi (s)

Patlak

Patlak

10

0.5, 6

5, 100

0.01, 0.02, 0.05, 0.1, 0.2, 0.35

N/A

0.001, 0.005 0.01 0.02 0.05 0.1

N/A

N/A

Tofts

Tofts

10

0.5, 6

5, 100

0.01, 0.02, 0.05, 0.1, 0.2, 0.35

0.01, 0.02, 0.05, 0.1, 0.2, 0.5

N/A

N/A

N/A

Ex-Tofts

Ex-Tofts

10

0.5, 6

5, 100

0.01, 0.02, 0.05, 0.1, 0.2, 0.35

0.01, 0.02, 0.05, 0.1, 0.2, 0.5

0.001, 0.005 0.01 0.02 0.05 0.1

N/A

N/A

2CXM

2CXM

10

0.5, 6

5, 100

0.01, 0.02, 0.05, 0.1, 0.2, 0.35

0.01, 0.02, 0.05, 0.1, 0.2, 0.5

0.001, 0.005 0.01 0.02 0.05 0.1

0.5, 1, 5

N/A

Tissue uptake

Tissue uptake

10

0.5, 6

5, 100

0.01, 0.02, 0.05, 0.1, 0.2, 0.35

N/A

0.001, 0.005 0.01 0.02 0.05 0.1

0.5, 1, 5

N/A

FXR

FXR

10

0.5, 6

5, 100

0.01, 0.02, 0.05, 0.1, 0.2, 0.35

0.01, 0.02, 0.05, 0.1, 0.2, 0.5

N/A

N/A

0.1, 0.5, 2

Steady-state

Nested

10

0.5

5, 100

N/A

N/A

0.005, 0.1

N/A

N/A

Patlak

Nested

10

0.5

5, 100

0.01, 0.35

N/A

0.005, 0.1

N/A

N/A

Ex-Tofts

Nested

10

0.5

5, 100

0.01, 0.35

0.01, 0.1

0.005, 0.1

N/A

N/A

Figure 3 shows plots comparing fitted Ktrans values to the values used to generate the simulated curves for different models. At high SNR and short time resolution, there is good concordance between simulated and fitted values. This is reflected in the CCC comparisons (Tables 3, 4 and 5). Fitting using ROCKETSHIP was generally able to recover parameter values with high accuracy and precision, as demonstrated by the small error bars seen in Fig. 3 and the high CCC values for the fitted to actual value comparisons across multiple models and with different parameters. Results shown in Fig. 3 and Tables 3, 4 and 5 demonstrate that ROCKETSHIP perform on par or better compared to QIBA simulation fittings using the Tofts model with DCE@UrLAB and DCEMRI.jl both qualitatively (compared to figures generated by DCE@UrLAB [33]) and quantitatively (CCC ≥ 0.9 in general for Ktrans and ve recovery using Tofts model, similar to or better than the reported CCCs generated by DCEMRI.jl [28]). As with all curve fitting algorithms, the accuracy of the fit will be dependent on a number of factors [18], including SNR, sampling resolution and the number of parameters being fitted (Fig. 4, Tables 3, 4 and 5).
Fig. 3

Ktrans fitting of simulated data. Simulated data with time resolution of 0.5 s and SNR = 100 were fitted using the same model used to generate the simulation with ROCKETSHIP using default settings for the Patlak method (a), Tofts (b) and Extended Tofts models (c). Ktrans simulated vs. fitted were plotted as a function of ve and vp. Dashed line is unity. Error bars denote standard deviation. Given the similar fits, points for different ve and vp may overlap. Concordance correlation coefficients for these (and other model fits) are shown in Tables 3, 4 and 5

Table 3

Concordance correlation coefficients (CCC) comparing fitted and simulated Ktrans using different models and dependent parameters

Generating model

Fitting model

Time resolution (s)

SNR

Dependent parameter

vp = 0.001

0.005

0.01

0.02

0.05

0.1

Patlak

Patlak

0.5

5

vp

1.00

1.00

1.00

1.00

1.00

1.00

  

0.5

100

vp

1.00

1.00

1.00

1.00

1.00

1.00

  

6

5

vp

0.98

0.98

0.98

0.98

0.98

0.98

  

6

100

vp

1.00

1.00

1.00

1.00

1.00

1.00

     

ve = 0.01

0.02

0.05

0.1

0.2

0.5

Tofts

Tofts

0.5

5

ve

0.38

0.85

0.98

0.99

1.00

1.00

  

0.5

100

ve

1.00

1.00

1.00

1.00

1.00

1.00

  

6

5

ve

0.08

0.32

0.76

0.92

0.96

0.97

  

6

100

ve

0.67

0.92

1.00

1.00

1.00

1.00

     

ve = 0.01

0.02

0.05

0.1

0.2

0.5

Ex-Tofts

Ex-Tofts

0.5

5

ve

0.01

0.33

0.95

0.99

1.00

1.00

  

0.5

100

ve

0.92

1.00

1.00

1.00

1.00

1.00

  

6

5

ve

−0.02

0.05

0.41

0.84

0.96

0.98

  

6

100

ve

0.22

0.48

0.98

1.00

1.00

1.00

     

vp = 0.001

0.005

0.01

0.02

0.05

0.1

  

0.5

5

vp

0.74

0.73

0.71

0.68

0.61

0.51

  

0.5

100

vp

0.99

0.98

0.98

0.99

0.99

0.98

  

6

5

vp

0.57

0.58

0.57

0.56

0.45

0.41

  

6

100

vp

0.89

0.90

0.88

0.81

0.54

0.35

     

ve = 0.01

0.02

0.05

0.1

0.2

0.5

2CXM

2CXM

0.5

5

ve

−0.02

−0.02

0.11

0.22

0.65

0.90

  

0.5

100

ve

0.07

0.42

0.65

0.76

0.98

0.99

  

6

5

ve

−0.01

−0.01

−0.01

0.10

0.52

0.84

  

6

100

ve

−0.03

−0.06

0.04

0.38

0.84

0.98

     

vp = 0.001

0.005

0.01

0.02

0.05

0.1

  

0.5

5

vp

0.18

0.20

0.23

0.31

0.43

0.47

  

0.5

100

vp

0.32

0.63

0.72

0.76

0.76

0.74

  

6

5

vp

0.20

0.19

0.21

0.21

0.28

0.29

  

6

100

vp

0.18

0.26

0.34

0.41

0.45

0.45

     

Fp = 0.5

1

5

   
  

0.5

5

Fp

0.21

0.30

0.40

   
  

0.5

100

Fp

0.46

0.69

0.81

   
  

6

5

Fp

0.20

0.22

0.27

   
  

6

100

Fp

0.26

0.35

0.43

   
     

vp = 0.001

0.005

0.01

0.02

0.05

0.1

     

vp = 0.001

0.005

0.01

0.02

0.05

0.1

  

0.5

5

vp

0.98

1.00

0.98

1.00

1.00

1.00

  

0.5

100

vp

1.00

1.00

1.00

1.00

1.00

1.00

  

6

5

vp

0.88

0.88

0.88

0.91

0.98

0.99

  

6

100

vp

1.00

1.00

1.00

1.00

1.00

1.00

Tissue uptake

Tissue uptake

0.5

5

vp

0.98

1.00

0.98

1.00

1.00

1.00

  

0.5

100

vp

1.00

1.00

1.00

1.00

1.00

1.00

  

6

5

vp

0.88

0.88

0.88

0.91

0.98

0.99

  

6

100

vp

1.00

1.00

1.00

1.00

1.00

1.00

     

Fp = 0.5

1

5

   
  

0.5

5

Fp

1.00

0.98

1.00

   
  

0.5

100

Fp

1.00

1.00

1.00

   
  

6

5

Fp

0.82

0.97

0.98

   
  

6

100

Fp

1.00

1.00

    
     

ve = 0.01

0.02

0.05

0.1

0.2

0.5

     

ve = 0.01

0.02

0.05

0.1

0.2

0.5

  

0.5

5

ve

−0.01

0.02

0.08

0.14

0.39

0.95

  

0.5

100

ve

0.07

0.31

0.97

1.00

1.00

1.00

  

6

5

ve

0.01

−0.01

0.03

0.06

0.12

0.38

  

6

100

ve

0.05

0.14

0.71

0.96

0.99

1.00

     

τi = 0.1

0.5

2

   
  

0.5

5

τi

0.11

0.14

0.14

   
  

0.5

100

τi

0.56

0.59

0.56

   
  

6

5

τi

0.06

0.07

0.07

   
  

6

100

τi

0.54

0.61

0.52

   

100 curves for each model and fixed dependent parameter were generated as described in the text and Table 2. Ktrans values simulated are defined in Table 2. The CCC was calculated from the Ktrans (simulated) vs. Ktrans (fitted), such as depicted in Figure 5. Ktrans values from which CCCs were calculated were segregated according to the dependent parameter (vp, ve or Fp). A value of 1 shows near-perfect concordance, while 0 represents a low concordance relationship

Table 4

Concordance correlation coefficients (CCC) comparing fitted and simulated vp using different models and dependent parameters

Generating model

Fitting model

Time resolution (s)

SNR

Dependent parameter

Ktrans = 0.01

0.02

0.05

0.1

0.2

0.35

Patlak

Patlak

0.5

5

Ktrans

0.99

0.99

0.99

0.99

1.00

0.99

  

0.5

100

Ktrans

1.00

1.00

1.00

1.00

1.00

1.00

  

6

5

Ktrans

0.95

0.95

0.95

0.95

0.94

0.95

  

6

100

Ktrans

1.00

1.00

1.00

1.00

1.00

1.00

     

Ktrans = 0.01

0.02

0.05

0.1

0.2

0.35

Ex-Tofts

Ex-Tofts

0.5

5

Ktrans

0.99

0.99

0.99

0.99

0.98

0.98

  

0.5

100

Ktrans

1.00

1.00

1.00

1.00

1.00

1.00

  

6

5

Ktrans

0.95

0.95

0.94

0.93

0.91

0.88

  

6

100

Ktrans

1.00

1.00

1.00

1.00

0.95

0.89

     

ve = 0.01

0.02

0.05

0.1

0.2

0.5

  

0.5

5

ve

0.98

0.98

0.99

0.99

0.99

0.99

  

0.5

100

ve

1.00

1.00

1.00

1.00

1.00

1.00

  

6

5

ve

0.92

0.90

0.89

0.93

0.95

0.94

  

6

100

ve

0.92

0.92

1.00

1.00

1.00

1.00

     

Ktrans = 0.01

0.02

0.05

0.1

0.2

0.35

2CXM

2CXM

0.5

5

Ktrans

0.79

0.85

0.51

0.26

0.10

0.06

  

0.5

100

Ktrans

1.00

1.00

0.95

0.96

0.81

0.41

  

6

5

Ktrans

0.02

0.07

0.37

0.16

0.07

0.05

  

6

100

Ktrans

0.96

0.93

0.90

0.73

0.38

0.09

     

ve = 0.01

0.02

0.05

0.1

0.2

0.5

  

0.5

5

ve

0.95

0.90

0.65

0.31

0.15

0.02

  

0.5

100

ve

0.99

0.97

0.89

0.75

0.82

0.57

  

6

5

ve

0.44

0.56

0.35

0.11

0.02

−0.03

  

6

100

ve

0.97

0.93

0.75

0.51

0.34

0.13

     

Fp = 0.5

1

5

   
  

0.5

5

Fp

0.22

0.28

0.31

   
  

0.5

100

Fp

0.64

0.88

0.95

   
  

6

5

Fp

0.09

0.09

0.03

   
  

6

100

Fp

0.39

0.56

0.54

   
     

Ktrans = 0.01

0.02

0.05

0.1

0.2

0.35

Tissue uptake

Tissue uptake

0.5

5

Ktrans

0.95

0.78

0.18

0.03

−0.02

−0.03

  

0.5

100

Ktrans

1.00

1.00

1.00

1.00

1.00

0.48

  

6

5

Ktrans

0.33

0.39

0.31

0.17

0.02

0.01

  

6

100

Ktrans

0.74

0.83

0.79

0.80

0.77

0.69

     

Fp = 0.5

1

5

   
  

0.5

5

Fp

-0.01

-0.01

0.90

   
  

0.5

100

Fp

0.67

1.00

0.99

   
  

6

5

Fp

0.03

0.20

0.07

   
  

6

100

Fp

0.95

0.97

0.44

   

100 curves for each model and fixed dependent parameter were generated as described in the text and Table 2. vp values simulated are defined in Table 2. The CCC was calculated from the vp (simulated) vs. vp (fitted) . vp values from which CCCs were calculated were segregated according to the dependent parameter (Ktrans, ve or Fp). A value of 1 shows near-perfect concordance, while 0 represents a low concordance relationship

Table 5

Concordance correlation coefficients (CCC) comparing fitted and simulated ve using different models and dependent parameters

      

CCC for ve (simulated) vs. ve (fitted)

 

Generating model

Fitting model

Time resolution (s)

SNR

Dependent parameter

Ktrans = 0.01

0.02

0.05

0.1

0.2

0.35

Tofts

Tofts

0.5

5

Ktrans

0.71

0.96

1.00

1.00

0.94

0.97

  

0.5

100

Ktrans

1.00

1.00

1.00

1.00

1.00

1.00

  

6

5

Ktrans

0.33

0.52

0.68

0.63

0.65

0.68

  

6

100

Ktrans

0.95

1.00

1.00

1.00

1.00

1.00

     

Ktrans = 0.01

0.02

0.05

0.1

0.2

0.35

Ex-Tofts

Ex-Tofts

0.5

5

Ktrans

0.67

0.86

0.87

0.80

0.66

0.54

  

0.5

100

Ktrans

1.00

1.00

1.00

1.00

1.00

1.00

  

6

5

Ktrans

0.22

0.40

0.48

0.47

0.35

0.35

  

6

100

Ktrans

0.95

1.00

1.00

1.00

0.93

0.74

     

vp = 0.001

0.02

0.05

0.1

0.2

0.5

  

0.5

5

vp

0.74

0.73

0.69

0.71

0.73

0.70

  

0.5

100

vp

1.00

1.00

1.00

1.00

1.00

1.00

  

6

5

vp

0.38

0.38

0.36

0.36

0.32

0.32

  

6

100

vp

0.92

0.94

0.93

0.93

0.93

0.94

     

Ktrans = 0.01

0.02

0.05

0.1

0.2

0.35

2CXM

2CXM

0.5

5

Ktrans

0.51

0.52

0.29

0.19

0.12

0.06

  

0.5

100

Ktrans

0.97

0.97

0.93

0.84

0.70

0.49

  

6

5

Ktrans

0.17

0.27

0.24

0.15

0.06

0.03

  

6

100

Ktrans

0.81

0.88

0.73

0.46

0.23

0.10

     

vp = 0.001

0.02

0.05

0.1

0.2

0.5

  

0.5

5

vp

0.09

0.15

0.24

0.36

0.41

0.37

  

0.5

100

vp

0.50

0.83

0.88

0.91

0.87

0.86

  

6

5

vp

0.11

0.12

0.14

0.16

0.21

0.19

  

6

100

vp

0.23

0.41

0.49

0.55

0.59

0.63

     

Fp = 0.5

1

5

   
  

0.5

5

Fp

0.19

0.24

0.31

   
  

0.5

100

Fp

0.68

0.81

0.90

   
  

6

5

Fp

0.14

0.15

0.17

   
  

6

100

Fp

0.40

0.48

0.54

   

FXR

FXR

0.5

5

Ktrans

0.37

0.41

0.43

0.48

0.52

0.46

  

0.5

100

Ktrans

0.90

0.92

0.96

0.93

0.93

0.94

  

6

5

Ktrans

0.20

0.23

0.25

0.24

0.23

0.23

  

6

100

Ktrans

0.63

0.76

0.80

0.81

0.74

0.82

     

τi = 0.1

0.5

2

   
  

0.5

5

τi

0.41

0.45

0.44

   
  

0.5

100

τi

0.94

0.93

0.92

   
  

6

5

τi

0.24

0.23

0.22

   
  

6

100

τi

0.76

0.75

0.73

   

100 curves for each model and fixed dependent parameter were generated as described in the text and Table 2. ve values simulated are defined in Table 2. The CCC was calculated from the ve (simulated) vs. ve (fitted) . ve values from which CCCs were calculated were segregated according to the dependent parameter (Ktrans, vp or Fp). A value of 1 shows near-perfect concordance, while 0 represents a low concordance relationship

Fig. 4

ve fitting at different time resolutions. Simulated data using the Tofts model were generated at SNR = 5 and at time resolutions of 0.5 s (a) and 6 s (b). Simulated vs. fitted v e were plotted as a function of Ktrans. Dashed line is unity. Error bars represent standard deviation. As expected, lower time resolution results in a high standard deviation of the curve fits. Given the similar fits, points for different Ktrans may overlap. Concordance correlation coefficients for these (and other model fits) are shown in Tables 3, 4 and 5

Accuracy of model selection using nested model analysis

The data-driven, nested model analysis function in ROCKETSHIP was also tested. Simulation curves reflective of each level of nesting (steady-state, Patlak and extended Tofts models) were generated as above. The resultant curves were fitted using the nested method.

Simulation results using the nested model fitting are shown in Fig. 5 and Table 6. The nested model fitting was able to recover the appropriate model for the majority of the fitting curves. Accuracy of the nested model fitting was shown to be dependent on SNR. Lower SNR led to more model misclassification of voxels.
Fig. 5

Nested model selection from simulated data. a and b show fitting for steady-state model simulated data. c and d show the fitting for Patlak simulated data. All the generated curves at SNR = 100 converged to the correct model. At lower SNR, some of the curves incorrectly converged to Model 3 (extended Tofts). e and f show fitting on extended Tofts simulated data. Again, the majority of the curves converged to the correct model. The percentage of voxels attributed to each model by the nest model algorithm is shown in Table 6

Table 6

Model selection of simulated data using the nested model method

  

Percentage of voxels selected (%)

Generating model

SNR

Model 0

Model 1

Model 2

Model 3

Steady-State (Model 1)

5

2.75

41

43

13.25

Steady-State (Model 1)

100

0

75

20.75

4.25

Patlak (Model 2)

5

0

0

94.75

5.25

Patlak (Model 2)

100

0

0

100

0

Extended Tofts (Model 3)

5

0

11.25

19

69.5

Extended Tofts (Model 3)

100

0

8.5

0

91.5

In vivo examples

Preclinical data set

DCE-MRI was performed on an athymic BLABL/c nude mouse bearing a HER2-expressing BT474 human breast cancer tumor. A 17β-estradiol pellet (Innovative Research of America) was implanted subcutaneously into the back of the mouse one day prior to orthotopic inoculation (6th and 9th mammary fat pad) of 5 × 106 cells in matrigel (BD Biosciences). The tumor volume was ≈ 200 mm3 at the time of imaging. Mouse care and experimental procedures were carried out in accordance with protocols approved by the Institutional Animal Care and Use Committee at Caltech.

Imaging was performed using a Biospec (Bruker-Biospin Inc. Billerica, MA) 7 T MRI scanner and a custom-built birdcage coil. The mouse was anesthetized during the session with 1.3-1.5 % isoflurane/air mixture. Body temperature was maintained at 36-37 °C with warmed air through the bore. A variable flip angle method was used to generate T1 maps. Gradient echo images (FLASH, FA = 12°, 24°, 36°, 48°, 60°, matrix size = 140 × 80, voxel size = 0.25 × 0.25 mm2, slice thickness = 1 mm, TR/TE = 200/2 ms) centered on the tumor (3 slices) and the left ventricle (1 slice) were acquired. Next, DCE-MRI was acquired (FLASH, FA = 35°, TR/TE = 25/2 ms, geometry the same as the T1 maps, time resolution = 2 s, duration = 22 min). After a baseline of 2.5 min, Gd-DTPA (0.1 mmol/kg, Magnevist, Bayer) was injected intravenously via a tail vein catheter using a powered-injector (New Era Inc.) at 0.5 mL/min. Imaging data were processed using ROCKETSHIP and analyzed with the nested model method.

Parametric maps and associated concentration vs. time curves are shown in Fig. 6. For this tumor, the extended Tofts model was deemed the most appropriate for the majority of the tumor. Generated Ktrans values were similar to those observed in prior studies using the same tumor cell line [46]. Consistent with other studies, there is a heterogeneous distribution of Ktrans, ve and vp highlighting a leaky and vascularized rim with a necrotic core.
Fig. 6

Nested model fitting of DCE-MRI data on a murine breast cancer tumor model. Parameters for Ktrans (a), ve (b), and vp (c) are shown. As shown in d, the majority of the voxels fitted best to the extended Tofts model, with some edge voxels fitting to the Patlak method. e shows the AIF used for the fit (taken from the left ventricle). f shows a sample time curve from the edge of the tumor (denoted by arrow) with corresponding fit (blue denotes the fit, red lines denote the 95 % prediction bounds for the fitted curve). Rod phantoms on either side of the mouse were present to allow for signal drift correction (not used in this case)

Clinical data set

The clinical data set was obtained from a wider study set being accrued by the University of Southern California (USC) Alzheimer’s disease Research Center. The study was approved by the USC Institutional Review Board. The imaging dataset used here was obtained from one of the participants from the no cognitive impairment cohort, as determined from medical examination and neuropsychological evaluation. All imaging was performed at the Keck Medical Center of USC. Participants underwent a medical examination, neuropsychological evaluations and blood draw to ensure appropriate kidney function for CA administration prior to imaging.

The imaging protocol performed was developed to study BBB changes in patients with cognitive impairment and is detailed in [6]. Briefly, all images were obtained on a GE 3 T HDXT MR scanner with a standard eight-channel array head coil. Anatomical coronal spin echo T2-weighted scans were first obtained through the hippocampi (TR/TE 1550/97.15 ms, NEX = 1, slice thickness 5 mm with no gap, FOV = 188 x 180 mm, matrix size = 384 x 384). Baseline coronal T1-weighted maps were then acquired using a T1-weighted 3D spoiled gradient echo (SPGR) pulse sequence and variable flip angle method using flip angles of 2°, 5° and 10°. (TR/TE = 8.29/3.08 ms, NEX = 1, slice thickness 5 mm with no gap, FOV 188 x 180 mm, matrix size 160 x 160). Coronal DCE MRI covering the hippocampi and temporal lobes were acquired using a T1-weighted 3D SPGR pulse sequence (FA = 15°, TR/TE = 8.29/3.09 ms, NEX = 1, slice thickness 5 mm with no gap, FOV 188 x 180 mm, matrix size 160 x 160, voxel size was 0.625 x 0.625 x 5 mm3). This sequence was repeated for a total of 16 min with an approximate time resolution of 15.4 s. Gadobenate dimegulumine (MultiHance®, Braaco, Milan, Italy) (0.05 mmol/kg), a gadolinium-based CA, was administered intravenously into the antecubital vein using a power injector, at a rate of 3 mL/s followed by a 25 mL saline flush, 30 s into the DCE scan. Imaging data were processed by ROCKETSHIP. The AIF, which was extracted from an ROI positioned at the internal carotid artery, was fitted with a bi-exponential function prior to fitting analysis with 2CXM.

Parametric maps and associated concentration vs. time curves are shown in Fig. 7. Compared to the murine tumor that contains leaky vasculature [41], Ktrans values for the human brain are lower, as expected given the intact blood brain barrier in normal human subjects [6]. As shown in Fig. 7e, the bi-exponential function estimated the AIF well and allowed for a good fit of concentration time curves within the brain parenchyma (Fig. 7f).
Fig. 7

2CXM fitting of a normal human brain. Parameters for Ktrans (a), ve (b), vp (c) and Fp (d) are shown. e shows the AIF used (taken from internal carotid artery). The AIF was fitted with a bi-exponential curve (blue) prior to tissue fitting. f shows a sample time curve from the brain parenchyma (denoted by arrow) with corresponding fit (blue denotes the fit, red lines denote the 95 % prediction bounds for the fitted curve). Fold over artifact is seen on the lateral brain edges due to truncation by the field of view bounding box

Discussion

Development and adoption of DCE-MRI by clinicians and researchers requires the availability of analysis software that not only has sufficient functionality for the novice end-user, but also the flexibility and extensibility capability for more complicated analyses. Using its default settings, users can follow ROCKETSHIP’s pipeline to generate parametric maps of common kinetic models. Its modular design allows other models to be incorporated in a straightforward manner. Recently, several groups have recognized that data-driven approaches in selecting the appropriate model may be important for proper physiological interpretation of the DCE-MRI data [23]. To facilitate this, ROCKETSHIP includes a nested model fitting option as well as statistical tools to analyze and compare the goodness-of-fit between different kinetic models.

We selected the MATLAB environment for a number of reasons. Firstly, MATLAB is widely used for image analysis in the academic community; programs written in MATLAB can generally be run on Mac, Windows or Unix-based operating systems with minimal porting issues. Secondly, MATLAB provides several toolboxes (such as the Curve Fitting Toolbox and Parallel Computing Toolbox) that facilitated development of the current software and allows future extensions to be easily implemented. MATLAB also has a strong user-contributed library of imaging processing and statistical analysis functions that aid this endeavor. Building and testing of the software are helped by the interpretive nature of the MATLAB language. Modifications to the existing code or testing of new modules can be performed in real time. These capabilities enable users of wide-ranging expertise (from experienced programmers to biologists/clinicians with less programming experience) to use ROCKETSHIP and develop new modules and extend the software.

Since it is run within the MATLAB environment, the execution time of ROCKETSHIP may be slower than other DCE-MRI software written in compiled languages (C, C++, IDL). The use of the Parallel Computing Toolboxto do the voxel-by-voxel processing in parallel mostly offsets this limitation. While this may be a disadvantage in a daily clinical workflow setup, most DCE-MRI data are analyzed offline and often in a research context. As discussed above, ROCKETSHIP will be advantageous in this scenario. Furthermore, MATLAB code can be ported to C/C++ and Julia [28] in a relatively straightforward way. Thus, once an optimal setup is developed using ROCKETSHIP, one can translate this to a lower level language and ultimately as a standalone program.

The precision and accuracy of DCE-MRI parameter estimation in vivo is dependent on several factors, ranging from the type of tissue being probed, the type of image acquisition protocol/system used to the post-acquisition processing and analysis methodologies. DCE-MRI software packages address the post-acquisition portion of this pipeline. To evaluate ROCKETSHIP’s ability in this regard, simulation datasets with a range of models and parameters expected in typical studies were generated and fitted. Results demonstrate that ROCKETSHIP was able to recover DCE-MRI parameters accurately using its default configuration, on par with (or better than) similar studies using the same pharmacokinetic models implemented within currently available software packages. Furthermore, the nested model functionality, unavailable in prior software suites, was generally able to select the appropriate kinetic model for a given simulation dataset. ROCKETSHIP was also able to fit both preclinical and clinical in vivo DCE-MRI data well, demonstrating its applicability for in vivo studies.

While the simulations presented cover a broad parameter range, it is likely that further modifications/tuning of ROCKETSHIP’s settings or functions will be required for specific in vivo applications in the future. Furthermore, extrinsic factors affecting the data fitting, including SNR, time resolution/duration, motion and most importantly the physiological question being explored, need to be considered when evaluating the appropriateness of the output values and model selection [23, 37, 47, 48].

Conclusion

We have implemented a modular, flexible and easily extensible software suite for dynamic MRI (in particular DCE-MRI) analysis. This software allows for DCE-MRI data analysis using several pharmacokinetic models used currently in the literature as well as data-driven analysis methods. It is compatible with both clinical and preclinical imaging data and is currently being used in DCE-MRI studies [6]. We envision that the flexibility and open source nature of our software will be useful for researchers and clinicians at varying levels of DCE-MRI expertise.

The source code repository and documentation for ROCKETSHIP is located at https://github.com/petmri/ROCKETSHIP. Future updates and support for ROCKETSHIP will be coordinated and maintained via this repository.

Availability and requirements

Project name: ROCKETSHIP v.1.1

Project homepage: https://github.com/petmri/ROCKETSHIP

Operating system(s): Windows/Mac OS X/Linux

Programming language: MATLAB

Other requirements: None, but image processing programs such as ImageJ or MRIcro are useful to pre-process inputs.

License: GPL-2.0

Notes

Declarations

Acknowledgments

The authors thank Drs. David Colcher and Andrew Raubitschek for providing the mouse tumor model for this study, Dr. Meng Law for providing the clinical data, Dr. Andrey Demyanenko, Hargun Sohi, Bita Alaghebandan and Sharon Lin for their technical assistance. Drs. Scott Fraser and Kazuki Sugahara provided helpful advice and support. This work was funded in part by NIBIB R01 EB000993, NIH R01 EB00194, NRSA T32GM07616, City of Hope Lymphoma SPORE Grant (P50 CA107399), the Beckman Institute, the Caltech/City of Hope Biomedical Initiative, and R37NS34467, R37AG23084, and R01AG039452.

Authors’ Affiliations

(1)
Division of Biology and Biological Engineering, California Institute of Technology
(2)
Department of Medicine, University of California, Irvine Medical Center
(3)
Zilkha Neurogenetic Institute and Department of Physiology and Biophysics, Keck School of Medicine, University of Southern California

References

  1. Parker GJ, Padhani AR. T1‐W DCE‐MRI: T1‐Weighted Dynamic Contrast‐Enhanced MRI, Quantitative MRI of the Brain: Measuring Changes Caused by Disease. 2004. p. 341–64.Google Scholar
  2. Cuenod C, Fournier L, Balvay D, Guinebretiere J-M. Tumor angiogenesis: pathophysiology and implications for contrast-enhanced MRI and CT assessment. Abdom Imaging. 2006;31(2):188–93.PubMedView ArticleGoogle Scholar
  3. Hylton N. Dynamic contrast-enhanced magnetic resonance imaging as an imaging biomarker. J Clin Oncol. 2006;24(20):3293–8.PubMedView ArticleGoogle Scholar
  4. O’Connor JP, Jackson A, Parker GJ, Jayson GC. DCE-MRI biomarkers in the clinical evaluation of antiangiogenic and vascular disrupting agents. Br J Cancer. 2007;96(2):189–95.PubMedPubMed CentralView ArticleGoogle Scholar
  5. Frias AE, Schabel MC, Roberts VHJ, Tudorica A, Grigsby PL, Oh KY, et al. Using dynamic contrast-enhanced MRI to quantitatively characterize maternal vascular organization in the primate placenta. Magn Reson Med. 2014;1570–1578.Google Scholar
  6. Montagne A, Barnes SR, Sweeney MD, Halliday MR, Sagare AP, Zhao Z, et al. Blood-Brain Barrier Breakdown in the Aging Human Hippocampus. Neuron. 2015;85(2):296–302.PubMedPubMed CentralView ArticleGoogle Scholar
  7. Sourbron S, Ingrisch M, Siefert A, Reiser M, Herrmann K. Quantification of cerebral blood flow, cerebral blood volume, and blood–brain-barrier leakage with DCE-MRI. Magn Reson Med. 2009;62(1):205–17.PubMedView ArticleGoogle Scholar
  8. Strijkers GJ, Mulder WJ, van Tilborg GA, Nicolay K. MRI Contrast Agents: Current Status and Future Perspectives. Anticancer Agents Med Chem. 2007;7(3):291–305.PubMedView ArticleGoogle Scholar
  9. Di Giovanni P, Azlan CA, Ahearn TS, Semple SI, Gilbert FJ, Redpath TW. The accuracy of pharmacokinetic parameter measurement in DCE-MRI of the breast at 3 T. Phys Med Biol. 2010;55(1):121.PubMedView ArticleGoogle Scholar
  10. Chassidim Y, Veksler R, Lublinsky S, Pell GS, Friedman A, Shelef I. Quantitative imaging assessment of blood-brain barrier permeability in humans. Fluids Barriers CNS. 2013;10(1):9.PubMedPubMed CentralView ArticleGoogle Scholar
  11. Chih‐Feng C, Ling‐Wei H, Chun‐Chung L, Chen‐Chang L, Hsu‐Huei W, Yuan‐Hsiung T, et al. In vivo correlation between semi‐quantitative hemodynamic parameters and Ktrans derived from DCE‐MRI of brain tumors. Int J Imag Syst Tech. 2012;22(2):132–6.View ArticleGoogle Scholar
  12. Sourbron SP, Buckley DL. Classic models for dynamic contrast-enhanced MRI. NMR Biomed. 2013;26(8):1004–27.PubMedView ArticleGoogle Scholar
  13. Jackson A, Li K-L, Zhu X. Semi-Quantitative Parameter Analysis of DCE-MRI Revisited: Monte-Carlo Simulation, Clinical Comparisons, and Clinical Validation of Measurement Errors in Patients with Type 2 Neurofibromatosis. PLoS One. 2014;9(3), e90300.PubMedPubMed CentralView ArticleGoogle Scholar
  14. Koh TS, Bisdas S, Koh DM, Thng CH. Fundamentals of tracer kinetics for dynamic contrast‐enhanced MRI. J Magn Reson Imaging. 2011;34(6):1262–76.PubMedView ArticleGoogle Scholar
  15. Parker GJM, Roberts C, Macdonald A, Buonaccorsi GA, Cheung S, Buckley DL, et al. Experimentally-derived functional form for a population-averaged high-temporal-resolution arterial input function for dynamic contrast-enhanced MRI. Magn Reson Med. 2006;56(5):993–1000.PubMedView ArticleGoogle Scholar
  16. Heye T, Boll DT, Reiner CS, Bashir MR, Dale BM, Merkle EM. Impact of precontrast T10 relaxation times on dynamic contrast‐enhanced MRI pharmacokinetic parameters: T10 mapping versus a fixed T10 reference value. J Magn Reson Imaging. 2013;39(5):1136–45.View ArticleGoogle Scholar
  17. Srikanchana R, Thomasson D, Choyke P, Dwyer A. A comparison of Pharmacokinetic Models of Dynamic Contrast Enhanced MRI. In: Computer-Based Medical Systems, 2004 CBMS 2004 Proceedings 17th IEEE Symposium on: 24-25 June 2004 2004. 2004. p. 361–6.Google Scholar
  18. Cramer SP, Larsson HB. Accurate determination of blood-brain barrier permeability using dynamic contrast-enhanced T1-weighted MRI: a simulation and in vivo study on healthy subjects and multiple sclerosis patients. J Cereb Blood Flow Metab. 2014;34(10):1655–65.PubMedPubMed CentralView ArticleGoogle Scholar
  19. Zloković BV, Lipovac MN, Begley DJ, Davson H, Rakić L. Transport of Leucine-Enkephalin Across the Blood-Brain Barrier in the Perfused Guinea Pig Brain. J Neurochem. 1987;49(1):310–5.PubMedView ArticleGoogle Scholar
  20. Zlokovic BV, Begley DJ, Chain-Eliash DG. Blood-brain barrier permeability to leucine-enkephalin, d-Alanine2-d-leucine5-enkephalin and their N-terminal amino acid (tyrosine). Brain Res. 1985;336(1):125–32.PubMedView ArticleGoogle Scholar
  21. Zlokovic B. Cerebrovascular Permeability to Peptides: Manipulations of Transport Systems at the Blood-Brain Barrier. Pharm Res. 1995;12(10):1395–406.PubMedView ArticleGoogle Scholar
  22. Larsson HB, Courivaud F, Rostrup E, Hansen AE. Measurement of brain perfusion, blood volume, and blood-brain barrier permeability, using dynamic contrast-enhanced T(1)-weighted MRI at 3 tesla. Magn Reson Med. 2009;62(5):1270–81.PubMedView ArticleGoogle Scholar
  23. Ewing JR, Bagher-Ebadian H. Model selection in measures of vascular parameters using dynamic contrast-enhanced MRI: experimental and clinical applications. NMR Biomed. 2013;26(8):1028–41.PubMedPubMed CentralView ArticleGoogle Scholar
  24. Ewing JR, Brown SL, Lu M, Panda S, Ding G, Knight RA, et al. Model selection in magnetic resonance imaging measurements of vascular permeability: Gadomer in a 9L model of rat cerebral tumor. J Cereb Blood Flow Metab. 2006;26(3):310–20.PubMedView ArticleGoogle Scholar
  25. Stefanovski D, Moate PJ, Boston RC. WinSAAM: a windows-based compartmental modeling system. Metabolism. 2003;52(9):1153–66.PubMedView ArticleGoogle Scholar
  26. Barrett PHR, Bell BM, Cobelli C, Golde H, Schumitzky A, Vicini P, et al. SAAM II: Simulation, analysis, and modeling software for tracer and pharmacokinetic studies. Metabolism. 1998;47(4):484–92.PubMedView ArticleGoogle Scholar
  27. Heye T, Davenport MS, Horvath JJ, Feuerlein S, Breault SR, Bashir MR, et al. Reproducibility of dynamic contrast-enhanced MR imaging. Part I. Perfusion characteristics in the female pelvis by using multiple computer-aided diagnosis perfusion analysis solutions. Radiology. 2013;266(3):801–11.PubMedView ArticleGoogle Scholar
  28. Smith DS, Li X, Arlinghaus LR, Yankeelov TE, Welch EB. DCEMRI.jl: A fast, validated, open source toolkit for dynamic contrast enhanced MRI analysis. PeerJ Pre Prints. 2014;2:e670v671.Google Scholar
  29. Cetin O. An analysis tool to calculate permeability based on the Patlak method. J Med Syst. 2012;36(3):1317–26.PubMedView ArticleGoogle Scholar
  30. Barboriak DP, MacFall JR, Padua AO, York GE, Viglianti BL, Dewhirst MW. Standardized software for calculation of Ktrans and vp from dynamic T1-weighted MR images. In: International Society for Magnetic Resonance in Medicine Workshop on MR in Drug Development: From Discovery to Clinical Therapeutic Trials: 2004; McLean, VA; 2004.Google Scholar
  31. Whitcher B, Schmid VJ. Quantitative analysis of dynamic contrast-enhanced and diffusion-weighted magnetic resonance imaging for oncology in R. J Stat Softw. 2011;44(5):1–29.View ArticleGoogle Scholar
  32. Ferl GZ. DATforDCEMRI: an R package for deconvolution analysis and visualization of DCE-MRI data. J Stat Softw. 2011;44(3):1–18.View ArticleGoogle Scholar
  33. Ortuno JE, Ledesma-Carbayo MJ, Simoes RV, Candiota AP, Arus C, Santos A. DCE@urLAB: a dynamic contrast-enhanced MRI pharmacokinetic analysis tool for preclinical data. BMC Bioinformatics. 2013;14:316.PubMedPubMed CentralView ArticleGoogle Scholar
  34. Zollner FG, Weisser G, Reich M, Kaiser S, Schoenberg SO, Sourbron SP, et al. UMMPerfusion: an open source software tool towards quantitative MRI perfusion analysis in clinical routine. J Digit Imaging. 2013;26(2):344–52.PubMedView ArticleGoogle Scholar
  35. Balvay D, Frouin F, Calmon G, Bessoud B, Kahn E, Siauve N, et al. New criteria for assessing fit quality in dynamic contrast-enhanced T1-weighted MRI for perfusion and permeability imaging. Magn Reson Med. 2005;54(4):868–77.PubMedView ArticleGoogle Scholar
  36. Khalifa F, Soliman A, El-Baz A, El-Ghar MA, El-Diasty T, Gimel’farb G, et al. Models and methods for analyzing DCE-MRI: A review. Med Phys. 2014;41(12):124301.PubMedView ArticleGoogle Scholar
  37. Bagher-Ebadian H, Jain R, Nejad-Davarani SP, Mikkelsen T, Lu M, Jiang Q, et al. Model Selection for DCE-T1 Studies in Glioblastoma. Magn Reson Med. 2012;68(1):241–51.PubMedView ArticleGoogle Scholar
  38. Huang W, Li X, Chen Y, Li X, Chang MC, Oborski MJ, et al. Variations of dynamic contrast-enhanced magnetic resonance imaging in evaluation of breast cancer therapy response: a multicenter data analysis challenge. Transl Oncol. 2014;7(1):153–66.PubMedPubMed CentralView ArticleGoogle Scholar
  39. Kim H, Folks KD, Guo L, Stockard CR, Fineberg NS, Grizzle WE, et al. DCE-MRI detects early vascular response in breast tumor xenografts following anti-DR5 therapy. Mol Imaging Biol. 2011;13(1):94–103.PubMedPubMed CentralView ArticleGoogle Scholar
  40. Yankeelov TE, DeBusk LM, Billheimer DD, Luci JJ, Lin PC, Price RR, et al. Repeatability of a reference region model for analysis of murine DCE-MRI data at 7T. J Magn Reson Imaging. 2006;24(5):1140–7.PubMedView ArticleGoogle Scholar
  41. Loveless ME, Lawson D, Collins M, Nadella MV, Reimer C, Huszar D, et al. Comparisons of the efficacy of a Jak1/2 inhibitor (AZD1480) with a VEGF signaling inhibitor (cediranib) and sham treatments in mouse tumors using DCE-MRI, DW-MRI, and histology. Neoplasia. 2012;14(1):54–64.PubMedPubMed CentralView ArticleGoogle Scholar
  42. Anderson KB, Conder JA. Discussion of Multicyclic Hubbert Modeling as a Method for Forecasting Future Petroleum Production. Energy Fuels. 2011;25(4):1578–84.View ArticleGoogle Scholar
  43. Glatting G, Kletting P, Reske SN, Hohl K, Ring C. Choosing the optimal fit function: comparison of the Akaike information criterion and the F-test. Med Phys. 2007;34(11):4285–92.PubMedView ArticleGoogle Scholar
  44. Motulsky H, Christopoulos A. Fitting models to biological data using linear and nonlinear regression: a practical guide to curve fitting: Oxford University Press; 2004.Google Scholar
  45. Lin LI. A concordance correlation coefficient to evaluate reproducibility. Biometrics. 1989;45(1):255–68.PubMedView ArticleGoogle Scholar
  46. Li K-L, Wilmes LJ, Henry RG, Pallavicini MG, Park JW, Hu-Lowe DD, et al. Heterogeneity in the angiogenic response of a BT474 human breast cancer to a novel vascular endothelial growth factor-receptor tyrosine kinase inhibitor: Assessment by voxel analysis of dynamic contrast-enhanced MRI. J Magn Reson Imaging. 2005;22(4):511–9.PubMedView ArticleGoogle Scholar
  47. Henderson E, Rutt BK, Lee TY. Temporal sampling requirements for the tracer kinetics modeling of breast disease. Magn Reson Imaging. 1998;16(9):1057–73.PubMedView ArticleGoogle Scholar
  48. Larsson C, Kleppesto M, Rasmussen Jr I, Salo R, Vardal J, Brandal P, et al. Sampling requirements in DCE-MRI based analysis of high grade gliomas: simulations and clinical results. J Magn Reson Imaging. 2013;37(4):818–29.PubMedView ArticleGoogle Scholar
  49. Ng TS, Wert D, Sohi H, Procissi D, Colcher D, Raubitschek AA, et al. Serial diffusion MRI to monitor and model treatment response of the targeted nanotherapy CRLX101. Clin Cancer Res. 2013;19(9):2518–27.PubMedPubMed CentralView ArticleGoogle Scholar
  50. McGrath DM, Bradley DP, Tessier JL, Lacey T, Taylor CJ, Parker GJ. Comparison of model‐based arterial input functions for dynamic contrast‐enhanced MRI in tumor bearing rats. Magn Reson Med. 2009;61(5):1173–84.PubMedView ArticleGoogle Scholar
  51. Huang W, Li X, Morris EA, Tudorica LA, Seshan VE, Rooney WD, et al. The magnetic resonance shutter speed discriminates vascular properties of malignant and benign breast tumors in vivo. Proc Natl Acad Sci U S A. 2008;105(46):17943–8.PubMedPubMed CentralView ArticleGoogle Scholar
  52. Paudyal R, Poptani H, Cai K, Zhou R, Glickson JD. Impact of transvascular and cellular–interstitial water exchange on dynamic contrast-enhanced magnetic resonance imaging estimates of blood to tissue transfer constant and blood plasma volume. J Magn Reson Imaging. 2013;37(2):435–44.PubMedView ArticleGoogle Scholar
  53. Li X, Rooney WD, Springer Jr CS. A unified magnetic resonance imaging pharmacokinetic theory: intravascular and extracellular contrast reagents. Magn Reson Med. 2005;54(6):1351–9.PubMedView ArticleGoogle Scholar
  54. Zhou R, Pickup S, Yankeelov TE, Springer CS, Glickson JD. Simultaneous measurement of arterial input function and tumor pharmacokinetics in mice by dynamic contrast enhanced imaging: Effects of transcytolemmal water exchange. Magn Reson Med. 2004;52(2):248–57.PubMedView ArticleGoogle Scholar

Copyright

© Barnes et al. 2015

This article is published under license to BioMed Central Ltd. This is an Open Access article distributed under under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/4.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly credited. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.