Synthetic double inversion recovery imaging in brain MRI: quantitative evaluation and feasibility of synthetic MRI and a comparison with conventional double inversion recovery and fluid-attenuated inversion recovery sequences

Abstract

Background and purpose

Synthetic MR imaging (SyMRI) allows the reconstruction of various contrast images, including double inversion recovery (DIR), from a single scan. This study aimed to investigate the advantages of SyMRI by comparing synthetic DIR images with synthetic T2-weighted fluid-attenuated inversion recovery (T2W-FLAIR) and conventional DIR images.

Materials and methods

We retrospectively reviewed the imaging data of 100 consecutive patients who underwent brain MRI between December 2018 and March 2019. Synthetic DIR, T2W-FLAIR, T1-weighted, and phase-sensitive inversion recovery (PSIR) images were generated from SyMRI data. For synthetic DIR, the two inversion times required to suppress white matter and cerebrospinal fluid (CSF) were manually determined by two radiologists. Quantitative analysis was performed by manually tracing the region of interest (ROI) at the sites of the lesion, white matter, and CSF. Synthetic DIR, synthetic T2W-FLAIR, and conventional DIR images were compared on the basis of using the gray matter-to-white matter, lesion-to-white matter, and lesion-to-CSF contrast-to-noise ratios.

Results

The two radiologists showed no differences in setting inversion time (TI) values, and their evaluations showed excellent interobserver agreement. The mean signal intensities obtained with synthetic DIR were significantly higher than those obtained with synthetic T2W-FLAIR and conventional DIR.

Conclusion

Synthetic DIR images showed a higher contrast than synthetic T2WFLAIR and conventional DIR images.

Peer Review reports

Introduction

MRI is widely used to evaluate intracranial pathology because of its excellent soft-tissue contrast. To clearly depict lesions in the brain, inversion recovery sequences, such as fluid-attenuated inversion recovery (FLAIR) and double inversion recovery (DIR), are often employed in clinical practice. Redpath et al. reported that the DIR sequence improved the contrast between the lesion and background areas by suppressing signals from the cerebrospinal fluid (CSF) and normal white matter (WM) [1]. DIR has been used to evaluate various lesions in multiple sclerosis, ischemic diseases, epilepsy, tumors, and degenerative diseases [2,3,4,5,6].

Nevertheless, the addition of DIR sequences to routine clinical practice remains challenging because of the increased examination time. Moreover, adjustment of the dual inversion time (TI1 and TI2) parameters to suit various patient conditions to selectively suppress WM and CSF may be challenging in routine clinical practice.

Synthetic magnetic resonance imaging (SyMRI) techniques can solve the problems of increased examination times and selection of appropriate TI values by allowing manual changes to the appropriate TI values after image acquisition. The image data from SyMRI examinations can be used to generate multiple contrasts from a single scan based on a quantitative approach that includes absolute physical properties, such as the longitudinal T1-relaxation time, transverse T2-relaxation time, and proton density [7]. Thus, acquisition parameters such as repetition time (TR), echo time (TE), and inversion time (TI) can be generated using mathematical reasoning rather than predetermined methods [8].

Synthetic DIR images have been generated from SyMRI image data to set null points in two different brain structures: the WM and CSF. Although appropriate TI values can be determined by drawing manual regions of interest (ROIs) in these structures, qualitative and quantitative evaluation of the generated images is essential. Therefore, this study aimed to evaluate the feasibility of SyMRI in clinical practice and investigate the advantages of SyMRI by comparing it with synthetic T2-weighted (T2W)-FLAIR and conventional DIR images.

Materials and methods

Patients

The Institutional Review Committee of the Fujioka General Hospital approved the study as well as the informed consent waiver (FJ-152, see Additional file 1). All methods were carried out in accordance with relevant guidelines and regulations.

One hundred consecutive patients who underwent MRI between December 2018 and March 2019 were considered for evaluation. Those ninety-nine patients’ data was used in the quantitative evaluation of synthetic MRI. One patient was excluded because of a diffuse WM lesion. Eleven patients underwent both synthetic and conventional DIR imaging. Therefore, comparison between synthetic DIR and conventional DIR were based only on cases in which both were imaged (Fig. 1).

Fig. 1
figure 1

Flowchart for patient selection

Image acquisition

MRI was performed using a GE Healthcare 1.5 T system with a 32-channel sensitivity encoding head coil. The routine clinical MRI protocol consisted of axial DIR and FLAIR sections. SyMRI was performed using a magnetic resonance image compilation sequence (MAGiC; GE Healthcare Japan, Tokyo, Japan), and to generate TW2-FLAIR, Phase-sensitive inversion recovery (PSIR), and T1WI, parameters such as TR, TE, and TI were determined as shown in Table 1. Imaging parameters were as follows: field of view (FOV), 220 × 220 mm; matrix, 512 × 512; receiver bandwidth, 495 kHz; slice thickness/gap, 5.0 mm/2.4 mm; and slices, 22.

Table 1 The parameters used for synthetic MR sequences

Reconstructed synthetic DIR image

Circular ROIs were set to 5–10 mm by two radiologists (K. Y and J. A. with 18 and 15 years of experience, respectively). The MAGiC workstation is installed in the MR operation console. We use this software to draw ROIs and get values in each ROI. TI1 and TI2 were chosen to suppress both tissues concurrently. For TI1, ROIs were placed in the frontal WM, and for TI2, ROIs were placed in the CSF within the frontal horn of the lateral ventricle as shown in Fig. 2. ROIs were drawn on two slices at the corona radiata and corpus callosum levels.

Fig. 2
figure 2

An example of tissues suppressed with ROIs placement. Circular ROIs were nulled in the CSF and WM

Quantitative image analysis

For quantitative image analysis of synMRI, ROIs with a diameter of 5 mm were placed in three constructions in the same places on different contrast images on synMRI: normal-appearing WM (genu), normal-appearing gray matter (frontal cortex), and lesions. Mean signal intensities in the ROIs were calculated using the following formula:

  • SIGM/SIWM

  • SIlesion/SIWM

  • (SIGM/SIWM)/SIGM

  • (SIGM − SIWM)/(SIGM + SIWM)

where SIGM, SIWM, and SIlesion denote the signal intensities of GM, WM, and lesions, respectively.

In comparisons between synthetic DIR images and conventional DIR images, ROIs of normal-appearing WM, CSF, and lesions were set manually in the same slice and the same area. Mean signal intensities in the ROIs were calculated using the following formula:

  • SIlesion/SIWM

  • SIlesion/SICSF

where SICSF denotes the signal intensity of CSF.

Statistical analysis

All statistical analyses were performed using SPSS 26 for Windows (IBM Corp., Armonk, NY, USA) and GraphPad 9.0 software (GraphPad, California, USA). The differences between the TI values of the null points set by the two readers were analyzed using a nonparametric Mann–Whitney U test. Interobserver agreement between the two radiologists was determined using intraclass correlation coefficient (ICC) values. Spearman’s correlation coefficients were computed to assess the correlation between the measured TI values and age. Additionally, we determined whether statistically significant differences were present among the synthetic DIR, T2W-FLAIR, and conventional DIR images using a nonparametric Mann–Whitney U test. The mean and standard deviation were used to assess the entire population for all variables.

Results

Patient characteristics

A total of 99 patients (38 females and 61 males; mean age, 66 years; range, 15–95 years) were included in this study. Of the 44 patients, 18 showed periventricular hyperintensity, 9 showed chronic infractions, 6 showed acute infractions, 7 showed acute hemorrhages, 2 showed chronic hemorrhages, and the others showed malignant lymphomas and epilepsy. Seven patients with lesions were included in the comparison between synthetic and conventional DIR.

Reliability of TI values/comparison of TI values derived from MAGiC by the two readers

A comparison of the TI values is presented in Figs. 3 and 4. No significant differences were observed in the TI1 values for the corpus callosum and corona radiata in comparisons between the two readers (P = 0.359; P = 0.174, Fig. 2). The average ICC of the TI1 value for the corpus callosum and corona radiata was 0.95 (95% CI 0.93–0.97) and 0.90 (95% CI 0.85–0.93), respectively, indicating excellent agreement between readers. For the corpus callosum and corona radiata, TI2 values also showed excellent agreement between the two readers, with an ICC of 0.95 (95% CI 0.93–0.97) and 0.91 (95% CI 0.86–0.93), respectively, and no significant differences (P = 0.338; P = 0.198, Fig. 3).

Fig. 3
figure 3

The box plot represents the comparison of TI1 for each slice in assessments by the two readers with WM set to null point (A, B). The TI1 values were placed on the two slices at the position of the corona radiata and corpus callosum. No significant difference was observed in TI1 between the readers. The term “ns” indicates a non-significant difference. Note: TI inversion time, WM white matter

Fig. 4
figure 4

Comparison of TI2 for the two slices in the assessments performed by the two readers with CSF set to null point (A, B). No significant difference was observed in TI2 between the readers. The term “ns” indicates non-significant differences. Note: TI inversion time, CSF cerebrospinal fluid

Age effect

Figure 5 shows the correlation between the TI values and age. TI1 values showed a significant correlation with age in corpus callosum slices (r = 0.66; P < 0.0001). The corona radiata slices also showed a significant correlation between TI1 values and age (r = 0.61; P < 0.0001). No significant correlation was observed between the TI2 values and age in both imaging slices.

Fig. 5
figure 5

Correlation between age and TI1 values obtained in corona radiata image slices (A) and corpus callosum image slices (B). Correlation between age and TI2 values obtained in corona radiata image slices (C) and corpus callosum image slices (D). The TI1 values of the corona radiata image slices were moderately correlated with age (r = 0.61; P < 0.0001). Those of the corpus callosum image slices were also correlated (r = 0.66; P < 0.0001). The TI2 values are not significantly correlated with age in both image slice positions

Quantitative image assessments

Table 2 shows a comparison between the synthetic DIR and T2W-FLAIR images. For all formulas, the values obtained with synthetic DIR images were significantly higher than those obtained with synthetic T2W-FLAIR images. The mean SIGM/SIWM for synthetic DIR images and T2W-FLAIR images was 8.89 ± 4.93 and 1.66 ± 0.30 respectively, with significant differences (P < 0.001). The mean SIlesion/SIWM for synthetic DIR images and synthetic T2W-FLAIR images was 10.78 ± 4.81 and 2.32 ± 0.43, respectively, with significant differences (P < 0.001). The mean (SIGM/SIWM)/SIGM for synthetic DIR images and synthetic T2W-FLAIR images was 0.83 ± 0.13 and 0.38 ± 0.11, respectively, with significant differences (P < 0.001). The mean (SIGM − SIWM)/(SIGM + SIWM) for synthetic DIR images and synthetic T2W-FLAIR images was 0.74 ± 0.15 and 0.24 ± 0.08, respectively, with significant differences (P < 0.001).

Table 2 Comparison between synthetic DIR and synthetic T2W-FLAIR images

A comparison between synthetic and conventional DIR is shown in Table 3. Synthetic DIR images showed significantly higher values than conventional DIR images. The mean SIlesion/SIWM for synthetic DIR images and conventional DIR images was 9.81 ± 2.81 and 5.10 ± 1.36, respectively, with significant differences (P < 0.01). The mean SIlesion/SICSF for synthetic and conventional DIR images was 23.75 ± 4.36 and 5.37 ± 1.91, respectively, with significant differences (P < 0.001).

Table 3 Comparison between synthetic DIR and conventional DIR images

Discussion

We assessed the appropriate TI values for synthetic DIR images reconstructed from conventional MRI by using MAGiC software. Several studies have reported that synthetic DIR images may be helpful in the diagnosis of brain pathologies in both adults and children [9,10,11]. In patients with multiple plaques, the contrast and contrast-to-noise ratios of synthetic DIR images were significantly higher than those of conventional DIR images [9]. This finding is in accordance with our results because the contrast values of synthetic DIR were higher than those of conventional DIR in our study as well.

Our quantitative image analysis demonstrated that synthetic DIR images have higher contrast than synthetic TW2-FLAIR and conventional DIR sequences. A number of publications have demonstrated the superiority of synthetic DIR in disease assessment [9,10,11,12]. Hagiwara et al. reported synthetic DIR enabled for detection of multiple sclerosis plagues. Moreover, the quality of synthetic DIR using SyMRI software is superior to conventional DIR. To our knowledge this is first image quality evaluation of synthetic DIR images reconstructed by MAGiC software. In this study, the synthetic DIR, T2W-FLAIR, T1W, and PSIR images were reconstructed from conventional MRI data obtained with a scan time of approximately 4 min and 30 s. Consequently, synthetic MRI can be evaluated by merely adjusting the parameters offline, which is desirable for time-saving examinations. Furthermore, artifacts of synthetic images are seen in the DICOM file before processing. Therefore, it has no relation to the issue after image reconstruction by the SyMRI software. [13] In the diagnosis of lesion detection, synthetic DIR images can be used to suppress certain tissues for higher lesion detection. This could be another advantage of SyMRI imaging.

In this study, the average interclass correlation coefficient of TI values between the two radiologists showed excellent agreement with synthetic DIR images. Furthermore, the two radiologists showed no statistically significant difference in setting the TI values for both slice positions. Thus, in synthetic DIR, stable and appropriate TIs can be manually changed, regardless of the assessor who determines the ROI.

Interestingly, the TI1 value, which is a null point for WM, significantly increased with age in this study. This feature may cause difficulty in adequately setting the TI1 value of the WM in conventional DIR. Leukoaraiosis may also affect age-related white matter hyperintensity (WMH) [14]. Second, the genesis of WMH may reflect a failure of the glymphatic system. In glymphatic systems, perivascular space dilation with aging has been recently shown to reflect impaired failure to eliminate interstitial fluid from the WM, leading to excessive metabolic waste and inflammation [15, 16].

Our study had several limitations. The first limitation was the small number of patients who underwent both conventional DIR and SyMRI. Second, we did not measure TI1 and TI2 in conventional DIR images. Third, there has been no investigation into the disease because this study assessed the image quality. We did not evaluate the visual appearance of the lesion for diagnosis.

In conclusion, this study shows that synthetic DIR images can be helpful not only to reduce the examination time in clinical practice but also to provide high-contrast images than conventional DIR images.

Availability of data and materials

The datasets used and/or analyzed during the current study are available from the corresponding authors on reasonable request.

Abbreviations

FLAIR:

Fluid-attenuated inversion recovery

DIR:

Double inversion recovery

CSF:

Cerebrospinal fluid

WM:

White matter

SyMRI:

Synthetic magnetic resonance imaging

TR:

Repetition time

ET:

Echo time

TI:

Inversion time

ROI:

Region of interest

T2W:

T2-weighted

FOV:

Field of view

MAGiC:

Magnetic resonance image compilation sequence

PSIR:

Phase-sensitive inversion recovery

ICC:

Intraclass correlation coefficient

References

  1. Redpath TW, Smith FW. Imaging gray brain matter with a double-inversion pulse sequence to suppress CSF and white matter signals. Magn Reson Mater Phys, Biol Med. 1994;2(3):451–5.

    Article  Google Scholar 

  2. Wattjes MP, Lutterbey GG, Gieseke J, Traber F, Klotz L, Schmidt S, Schild HH. Double inversion¨ recovery brain imaging at 3T: diagnostic value in the detection of multiple sclerosis lesions. Am J Neuroradiol. 2007;28(1):54–9.

    CAS  PubMed  PubMed Central  Google Scholar 

  3. Choi NY, Park S, Lee CM, Ryu C-W, Jahng G-H. The role of double inversion recovery imaging in acute ischemic stroke. Investig Magn Reson Imaging. 2019;23(3):210–9.

    Article  Google Scholar 

  4. Rugg-Gunn FJ, Boulby PA, Symms MR, Barker GJ, Duncan JS. Imaging the neocortex in epilepsy with double inversion recovery imaging. Neuroimage. 2006;31(1):39–50. https://doi.org/10.1016/j.neuroimage.2005.11.034.

    Article  PubMed  Google Scholar 

  5. Umino M, Maeda M, Ii Y, Tomimoto H, Sakuma H. 3D double inversion recovery MR imaging: clinical applications and usefulness in a wide spectrum of central nervous system diseases. J Neuroradiol. 2018;46(2):107–16. https://doi.org/10.1016/j.neurad.2018.06.002.

    Article  PubMed  Google Scholar 

  6. Jahng G-H, Lee DK, Lee J-M, Rhee HY, Ryu C-W. Double inversion recovery imaging improves the evaluation of gray matter volume losses in patients with Alzheimer’s disease and mild cognitive impairment. Brain Imaging Behav. 2016;10(4):1015–28. https://doi.org/10.1007/s11682-015-9469-2.

    Article  PubMed  Google Scholar 

  7. Tanenbaum LN, Tsiouris AJ, Johnson AN, Naidich TP, DeLano MC, Melhem ER, Quarterman P, Parameswaran SX, Shankaranarayanan A, Goyen M, Field AS. Synthetic MRI for clinical neuroimaging: results of the magnetic resonance image compilation (MAGiC) prospective. Multicenter Multireader Trial. 2017;38(6):1103–10. https://doi.org/10.3174/ajnr.A5227.

    Article  CAS  Google Scholar 

  8. Hagiwara A, Warntjes M, Hori M, Andica C, Nakazawa M, Kumamaru KK, Abe O, Aoki S. SyMRI of the brain: rapid quantification of relaxation rates and proton density, with synthetic MRI, automatic brain segmentation, and myelin measurement. Investig Radiol. 2017;52(10):647–57. https://doi.org/10.1097/RLI.0000000000000365.

    Article  Google Scholar 

  9. Hagiwara A, Hori M, Yokoyama K, Takemura MY, Andica C, Tabata T, Kamagata K, Suzuki M, Kumamaru KK, Nakazawa M, Takano N, Kawasaki H, Hamasaki N, Kunimatsu A, Aoki S. Synthetic MRI in the detection of multiple sclerosis plaques. Am J Neuroradiol. 2017;38(2):257–63. https://doi.org/10.3174/ajnr.a5012.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  10. Andica C, Hagiwara A, Nakazawa M, Tsuruta K, Takano N, Hori M, Suzuki H, Sugano H, Arai H, Aoki S. The advantage of synthetic MRI for the visualization of early white matter change in an infant with Sturge–Weber syndrome. Magn Reson Med Sci. 2016;15(4):347–8. https://doi.org/10.2463/mrms.ci.2015-0164.

    Article  PubMed  PubMed Central  Google Scholar 

  11. Hagiwara A, Nakazawa M, Andica C, Tsuruta K, Takano N, Hori M, Suzuki H, Sugano H, Arai H, Aoki S. Dural enhancement in a patient with Sturge–Weber syndrome revealed by double inversion recovery contrast using synthetic MRI. Magn Reson Med Sci. 2015;15(2):151–2. https://doi.org/10.2463/mrms.ci.2015-0066.

    Article  PubMed  PubMed Central  Google Scholar 

  12. Andica C, Hagiwara A, Nakazawa M, Kumamaru KK, Hori M, Ikeno M, Shimizu T, Aoki S. Synthetic MR imaging in the diagnosis of bacterial meningitis. Magn Reson Med Sci. 2017;16(2):91–2. https://doi.org/10.2463/mrms.ci.2016-0082.

    Article  PubMed  Google Scholar 

  13. Granberg T, Uppman M, Hashim F, Cananau C, Nordin LE, Shams S, Berglund J, Forslin Y, Aspelin P, Fredrikson S, Kristoffersen-Wiberg M. Clinical feasibility of synthetic MRI in multiple sclerosis. Am J Neuroradiol. 2016;37(6):1023–9. https://doi.org/10.3174/ajnr.A4665.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  14. Grueter BE, Schulz UG. Age-related cerebral white matter disease (leukoaraiosis): a review. Postgrad Med J. 2012;88(1036):79. https://doi.org/10.1136/postgradmedj-2011-130307.

    Article  PubMed  Google Scholar 

  15. Huang P, Zhang R, Jiaerken Y, Wang S, Yu W, Hong H, Lian C, Li K, Zeng Q, Luo X, Yu X, Xu X, Wu X, Zhang M. Deep white matter hyperintensity is associated with the dilation of perivascular space. J Cereb Blood Flow Metab. 2021;41(9):2370–80. https://doi.org/10.1177/0271678x211002279.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  16. Aribisala BS, Wiseman S, Morris Z, Valdes-Hernandez MC, Royle NA, Maniega SM, Gow AJ, Corley J, Bastin ME, Starr J, Deary IJ, Wardlaw JM. Circulating inflammatory markers are associated with magnetic resonance imaging-visible perivascular spaces but not directly with white matter hyperintensities. Stroke. 2018;45(2):605–7. https://doi.org/10.1161/strokeaha.113.004059.

    Article  Google Scholar 

Download references

Acknowledgements

Thank you for colleagues, alumni, teachers and supporting staffs in the ”Asian Nuclear Medicine Graduation Program”. The authors declare no conflict of disclosure.

Funding

This research did not receive external funding.

Author information

Authors and Affiliations

Authors

Contributions

Z.O. wrote the main manuscript and collected and analyzed data. T.N. revised the manuscript and organized the research. A.J. and Y.K. evaluated images and also contributed to analyzing data. Y.T. gave Z.O. some suggestions about the research. All authors read and approved the final manuscript.

Corresponding author

Correspondence to Takahito Nakajima.

Ethics declarations

Ethics approval and consent to participate

The Institutional Review Committee of the Fujioka General Hospital approved the study as well as the informed consent waiver (FJ-152). The authors confirmed that all methods were carried out in accordance with relevant guidelines and regulations.

Consent for publication

Not applicable.

Competing interests

The authors declare that they have no competing interests.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Supplementary Information

Additional file 1.

Supplemental file S1. Ethical approval of this study.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Zorigt, O., Nakajima, T., Kumasaka, Y. et al. Synthetic double inversion recovery imaging in brain MRI: quantitative evaluation and feasibility of synthetic MRI and a comparison with conventional double inversion recovery and fluid-attenuated inversion recovery sequences. BMC Med Imaging 22, 183 (2022). https://doi.org/10.1186/s12880-022-00877-4

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s12880-022-00877-4

Keywords