Standardized quality metric system for structural brain magnetic resonance images in multi-center neuroimaging study

Abstract

Background

Multi-site neuroimaging offer several benefits and poses tough challenges in the drug development process. Although MRI protocol and clinical guidelines developed to address these challenges recommend the use of good quality images, reliable assessment of image quality is hampered by the several shortcomings of existing techniques.

Methods

Given a test image two feature images are extracted. They are grayscale and contrast feature images. Four binary images are generated by setting four different global thresholds on the feature images. Image quality is predicted by measuring the structural similarity between appropriate pairs of binary images. The lower and upper limits of the quality index are 0 and 1. Quality prediction is based on four quality attributes; luminance contrast, texture, texture contrast and lightness.

Results

Performance evaluation on test data from three multi-site clinical trials show good objective quality evaluation across MRI sequences, levels of distortion and quality attributes. Correlation with subjective evaluation by human observers is ≥ 0.6.

Conclusion

The results are promising for the evaluation of MRI protocols, specifically the standardization of quality index, designed to overcome the challenges encountered in multi-site clinical trials.

Peer Review reports

Background

Brain imaging studies using magnetic resonance imaging (MRI) system is one of the strongest biomarker candidates for neurological diseases [18]. MRI system is highly flexible. A single MRI system examination can be configured to generate several image sequences which can potentially provide high contrast structural information and the connectivity between brain structures [911]. The first listing of MRI as criteria for the diagnosis of multiple sclerosis and Alzheimer’s diseases was in 2001 and 2010, respectively [12, 13]. Over the years, advances in technology, strong collaboration among researchers coupled with availability of clinical data encouraged modifications [1417] to the initial diagnostic criteria. Despite several modifications to the initial diagnostic criteria, MRI criteria maintains a strong position to demonstrate the dissemination of multiple sclerosis disease in space and time and the exclusion of other disorders that can mimic its clinical and laboratory profile [18].

Pharmaceutical companies incorporate multi-center neuroimaging in the design of clinical trials for the treatment and the monitoring of neurological diseases. Multi-site clinical trials has several advantages. Main benefit is the ability to obtain more neuroimaging data per unit time across a wide variety of the patient population [19]. Other benefits include monitoring the progression of disease across a geographically, culturally and environmentally diverse population [20].

Several challenges encountered in multi-center neuroimaging studies result from the differences in MRI system parameters and image reconstruction routines. Two challenges are the variation of scanner technologies across the clinical trials sites and the different magnetic field inhomogeneities produced by a specific MRI model from the same manufacturer [21, 22]. These challenges renders acquired data including image quality evaluation unreliable with high risk of inaccurate diagnosis [20].

There have been proposals to mitigate the effects of these challenges. They include careful coordination, the use of standardized phantom studies and strict quality assurance of recommended imaging protocols across the clinical trial sites [18, 23, 24]. Another mitigation measure is to ensure that end points derived from brain MRI images are subjective and heavily dependent on radiologist interpretation [25]. This approach is grossly inefficient in large scale clinical trials where large volume of data are processed. Conflicting results can result from the variability and lack of reproducibility in the interpretation of data by radiologists [2, 26, 27].

Acquisition and reconstruction of MRI image from k-space data is time consuming relative to other imaging modalities such as X-ray, ultrasound and computed tomography. The consequences which include patient discomfort and motion-related artifacts significantly limits the potential to acquire high quality images [28, 29]. At the post-acquisition stage such as multi-center clinical trials, the challenge is centered on standardizing MRI data from different MRI scanners and different MRI sequences. Quality evaluation beyond the acquisition stage is necessary to evaluate the performance of MRI protocols developed to mitigate the challenges in multi-site neuroimaging.

Most proposed quality evaluation methods focus on the acquisition stage. There are few contributions on post-acquisition quality evaluation of brain MRI images. They include [30] which apply analysis of variance (ANOVA) algorithm to assess the variation of several quality measures with different levels of distortions. The authors in [31] combine the detection of artifacts and estimation of noise level to measure image quality. In [32] null space analysis and just noticeable difference scanning method was proposed as a better quality metric compared to root-mean-square error (RMSE). The popular signal-to-noise ratio (SNR) is the quality metric adopted in [33]. Recently the report in [34] propose a new method which predict brain MRI quality based on five quality attributes. The attributes are lightness, contrast, sharpness, texture details and noise. This report provide brief review of image quality evaluation. Detailed review of generic approaches to image quality evaluation can be found in [35, 36]. Review focussed on medical images is available in [31, 37, 38].

We share the same opinion in [39, 40] which suggest that the design of application-specific, no-reference quality evaluation system is more realistic and practicable than the design of generic image quality evaluation systems. It will be futile to design a generic image quality metric because images possess unique characteristics that distinguish them within and across different classes. Furthermore current state-of-the-art generic quality evaluation algorithms will require significant modifications before application to medical images [37].

Shortcomings of current state-of-the-art quality evaluation algorithm justify the need for new approach to quality evaluation in multi-center clinical trials. The contribution by [33] which adopt SNR is a full reference method based on the assumption that there exists a perfect image. In the real world a perfect image does not exist [37]. There are many definitions of SNR with no clearly defined range of quality index. These characteristics makes it difficult to compare quality measures from different imaging system, modalities and researchers [41]. Quality indices derived from SNR does not always correlate with the performance of observers using the imaging system on the task for which they are intended [41]. The use of SNR for quality evaluation can be said to be diagnostically misleading [42] because it cannot discriminate the quality of two images that are perceptually dissimilar [32]. The adoption of only artifacts and noise in [31] are too few attributes to evaluate the quality of an image. There are two setbacks for ANOVA based technique proposed in [30]. They are the risk of ambiguity in quality measures and the inability to transform the different levels of distortion into a quality index [43].

Image quality evaluation is difficult and complicated. Perceived image quality is influenced by several types of quality attributes and the different attributes influence each other [44]. The proposed method follow the three-step framework reported by Bartleson in 1982 [45]. The first step identify the most significant quality attributes of brain MRI images. They are luminance contrast, texture, texture contrast and lightness. The second step adopt image moments as global threshold to binarize and capture structural information contained in grayscale and contrast feature images derived from brain MRI images. Image moments have been successfully applied in many areas of image analysis such as image denoising [46], speech recognition [47], image normalization [48, 49], reconstruction [50], feature selection [51, 52], content-based image retrieval systems [53] and segmentation [54], [55]. In the third step image quality is evaluated based on structural similarity between the pair of images derived by using different moments to binarize each feature image.

This paper is organized as follows. The next section is the methods section. It include the setup of the experiment, problem formulation and implementation of the algorithm. The results of performance evaluation are presented in “Results” section and discussed in “Discussion” section. “Discussion” section highlight the limitations of our proposed method and the future work to overcome these limitations. “Conclusions” section concludes this report.

Methods

Sources of data

Data used for the performance evaluation of the proposed method were obtained from NeuroRx research Inc. (https://www.neurorx.com), BrainCare Oy. (http://braincare.fi/) and the Alzheimer’s disease neuroimaging initiative (ADNI) database (www.adni.loni.usc.edu).

NeuroRx research Inc. is an international clinical research organization dedicated to working with the pharmaceutical industry to facilitate clinical trials of new drugs for multiple sclerosis (MS) and other neurological diseases. BrainCare Oy (http://braincare.fi/) is a Tampere University of Technology spin-off company founded in 2013 to deliver personalized solutions to improve the quality of life of epilepsy patients. The organization recently concluded clinical trials for a novel mobile application and supporting solutions for long-term monitoring for epileptic patients. The ADNI was launched in 2003 as a public-private partnership, led by Principal Investigator Michael W. Weiner, MD. The primary goal of ADNI has been to test whether serial MRI, positron emission tomography (PET), other biological markers, and clinical and neuropsychological assessment can be combined to measure the progression of mild cognitive impairment and early Alzheimer’s disease.

Test data

The test data consist of three hundred MRI slices extracted from 15 T2 and 10 T1 MRI volume data. In the results section we demonstrate the efficacy of our proposed method with six MRI volume data; three from NeuroRx, two from BrainCare and one from ADNI. The dataset from NeuroRx are one T2 weighted and two conventional T1 weighted. Each volume data contain 60 slices with dimension 256×256 pixels and 2.4 mm thickness. The T2 weighted and one of the T1 weighted MRI data are without any perceived distortion. The slices in the other T1 weighted data were originally acquired with various configurations of bias fields. There are two 50-slice, T2 weighted data without perceived distortion from BrainCare Oy. Each slice has dimension 448×408 pixels and 2.6 mm thickness. The remaining data is T1 magnetization-prepared rapid gradient echo (MPRAGE) pulse sequence from ADNI. It contains 180 slices. Each slice has dimension 190×160 pixels and 1.2 mm thickness.

Performance evaluation was carried out in six categories. The first three categories evaluate the data in their original form of acquisition. They are T2 volume data without perceived distortion, T1 volume data without perceived distortion and T1 volume data degraded by bias fields. In the remaining three categories, three different types of degradation; circular blur, motion blur and Rician noise at different levels were artificially induced on the test data.

An image was degraded with circular blur by convolution with space-varying pillbox function. The parameter of circular blur is determined from the radial distance, in pixels, of the pillbox function. Motion blur degradation was done by a filter which approximates the linear motion of a camera. Motion blur is defined by linear and angular distances in pixels and degrees, respectively. Rician noise level was computed from the magnitude of Gaussian noise added to the real and imaginary components of the image. The real and imaginary components of the image are generated by making two duplicte copies of the image. Rician noise parameter is defined by the percentage of the maximum pixel intensity level in the image [56]. The different levels of circular blur, motion blur and Rician noise degradation were defined by scaling the respective parameters from 1 to 15.

Subjective Evaluation

In order to determine how our proposed method correlate with the human visual system we conducted subjective experiments with human observers. The experiment was aided by QuickEval [57], a web-based tool for psychometric image evaluation provided by the Norwegian Colour and Visual Computing Laboratory (www.colourlab.no/quickeval) at the Norwegian University of Science and Technology, Gjøvik, Norway. The observers are one radiologist and one MRI reader. MRI reader is a trained professional with experience working on MRI images that are affected by pathology [33]. Mean opinion score (MOS) subjective experiment was chosen for the validation study because it is simple and popular. Mean opinion score is the average of the quality scores assigned to an image by multiple viewers [58]. The six categories of the objective experiment was used for the subjective experiment. The observer assigns a score between 0 and 100, in steps of 1 to each slice. In the category of MRI volume data with artificially induced distortion, each observer was first presented with an undistorted version of an MRI slice, followed by increasing distortion levels of the original slice. The distorted levels are 5, 10 and 15. The relationship between our objective results and the score assigned by human observers was determined using the spearman’s rank correlation coefficient ρ [59]:

$$ \rho=1 - \frac{6\sum d^{2}}{n^{3} - n} $$
(1)

where n, the number of observations is the total number of slices contained in all the volume data in each category of the experiment, d is the difference between the two ranks of each observation.

Notational definitions

Here we explain and define the four binary images derived from the grayscale and contrast feature images. Each binary version is identified by a four-word name. The first three words describe the image moment of the feature image used as the global threshold to derive the binary image. The last word describe the original feature image before it was transformed to the binary domain. Each binary version is denoted by \(J_{I}^{M}\). The superscript denote the image moment M that was set as the global threshold to derive the binary image. The subscript is the original feature image I before transformation to the binary domain. Table 1 is a summary of the notational definitions.

Table 1 Notational definitions for the four binary feature images used for quality evaluation of brain MRI images

First grayscale moment grayscale binary image

First Grayscale Moment Grayscale (FGMG) binary image \(J_{I_{d}}^{\mu _{d}}\) is defined as the threshold version of the grayscale feature image Id at global threshold μd:

$$ J_{I_{d}}^{\mu_{d}}= \left\{\begin{array}{ll} 1 & \textbf{if} \quad I_{d} > \mu_{d}\\ 0 & \textbf{otherwise} \end{array}\right. $$
(2)

where μd is the first moment of the grayscale feature image. This image can be regarded as the brightness quality descriptor for the observed image because the number of bright pixels determines the luminosity of the image.

First contrast moment grayscale binary image

First Contrast Moment Grayscale (FCMG) binary image \(J^{\mu _{c}}_{I_{c}}\) is defined as the binary version of the grayscale feature image Id at global threshold μc:

$$ J^{\mu_{c}}_{I_{d}}= \left\{\begin{array}{ll} 1 & \textbf{if} \quad I_{d} > \mu_{c}\\ 0 & \textbf{otherwise} \end{array}\right. $$
(3)

The bright pixels measures the influence of the brightness quality attribute on the contrast quality attribute. It determines the number of grayscale pixels that contribute to the contrast quality attribute

First contrast moment contrast binary image

First Contrast Moment Contrast (FCMC) binary image \(J^{\mu _{c}}_{I_{c}}\) is defined as the threshold version of the local contrast feature image Ic at global threshold μc :

$$ J^{\mu_{c}}_{I_{c}}= \left\{\begin{array}{ll} 1 & \textbf{if} \quad I_{c} > \mu_{c}\\ 0 & \textbf{otherwise} \end{array}\right. $$
(4)

where μc is the first moment of the contrast feature image. This image can be regarded as the texture quality descriptor for the observed image.

First grayscale moment contrast binary image

First Grayscale Moment Contrast (FGMC) binary Image \(J^{\mu _{d}}_{I_{c}}\) is defined as the binary version of the contrast feature image Ic at global threshold μd:

$$ J^{\mu_{d}}_{I_{c}}= \left\{\begin{array}{ll} 1 & \textbf{if} \quad I_{c} > \mu_{d}\\ 0 & \textbf{otherwise} \end{array}\right. $$
(5)

The number of bright pixels in this image describe the interaction between the brightness and contrast quality attributes in the observed image.

Problem formulation

Classical quality attributes are generally adopted for simple images. The report in [60] recognize the need for better description of quality attributes in complex images. Contributions in the literature such as [61, 62] adopt terms such as luminance contrast, texture and texture contrast to describe quality attributes in specific complex images. Texture features has been widely applied to distinguish normal and abnormal structures in MRI images [6365].

We regard MRI image as a two-tissue class complex image. With reference to a T2 weighted MRI slice the bright pixels describe the high density of edges that characterize the cortical gray matter and the boundaries between the different anatomical structures. The white matter and other anatomical structures other than the cortical gray matter are described by the dark pixels.

We assume that all the possible distortions in an image can be condensed into either space-invariant point spread function or multiplicative spatially varying factor \(\mathcal {H}\) and random noise n according to the mathematical model of a 2D image acquisition process [66, 67] expressed by:

$$ I_{d}=\mathcal{H}I_{f} + n $$
(6)

where Id is the observed grayscale image and If is the underlying ideal image. In the absence of distortion and following the two-tissue class model, the observed image and its local contrast feature image are replica of the underlying ideal image:

$$ I_{d}=I_{c}=I_{f} $$
(7)

Based on Eq. 7 the local contrast feature image and the observed image will have same pixel intensity level:

$$ \mu_{d}=\mu_{c} $$
(8)

We hereby propose four quality scores. They are luminance contrast, texture, texture contrast and lightness. Each quality score is derived from the structural matching of relevant pair of binary feature images. The structural matching are described as pixel-wise structural matching and as edge-pixel structural matching. Pixel-wise and edge-pixel structural matching compares all the corresponding pixels and only edge pixels, respectively in the foreground of both images.

Luminance contrast quality score

The structural matching between edge pixels in FGMG and corresponding edge pixels in FCMG expressed as:

$$ q_{11}=J_{I_{d}}^{\mu_{d}} \cap J_{I_{d}}^{\mu_{c}} $$
(9)

measures how well the brightness quality attribute can be used to gauge the contrast quality attribute. This gives the luminance contrast quality score:

$$ q1= \frac{n_{q_{11}}}{\max \left(n_{J_{I_{d}}^{\mu_{d}}},n_{J_{I_{d}}^{\mu_{c}}}\right)} $$
(10)

where \(n_{q_{11}}\) is the number of bright pixels common to both FGMG and FCMG and the denominator is the highest number computed from the number \(n_{J_{I_{d}}^{\mu _{d}}}, n_{J_{I_{d}}^{\mu _{c}}}\) of bright pixels in both feature images. The number of bright pixels in both images are compared and used as denominator to ensure that the quality index q1≤1.

Texture quality score

The structural matching between edge pixels in FGMC and corresponding edge pixels in FCMC expressed as

$$ q_{22}=J_{I_{c}}^{\mu_{d}}\cap J_{I_{c}}^{\mu_{c}} $$
(11)

measures how well the brightness quality attribute can be used to gauge the texture quality attribute. This gives the texture quality score:

$$ q2= \frac{n_{q_{22}}}{\max \left(n_{J_{I_{c}}^{\mu_{d}}},n_{J_{I_{c}}^{\mu_{c}}}\right)} $$
(12)

where \(n_{q_{22}}\) is the number of bright pixels common to both FGMC and FCMC and the denominator is the highest number computed from the number \(n_{J_{I_{c}}^{\mu _{d}}}, n_{J_{I_{c}}^{\mu _{c}}}\) of bright pixels in both feature images.

Texture contrast quality score

The pixel-wise structural similarity matching between FGMC and FCMC expressed as

$$ q_{33}=J_{I_{c}}^{\mu_{d}}\cap J_{I_{c}}^{\mu_{c}} $$
(13)

gives the texture contrast quality score expressed as:

$$ q3= \frac{n_{q_{33}}}{n_{t}} $$
(14)

where \(n_{q_{33}}\) is the number of dark and bright pixels common to both FGMC and FCMC, and nt is the number of foreground pixels. In an ideal image where quality distortions are absent, Eq. 8 holds. Therefore the texture contrast quality score expressed by Eq. 14 is 1. For real images, texture contrast quality score is determined by the disparity between μd and μc.

Lightness quality score

The pixel-wise structural similarity matching between FGMG and FCMG expressed as

$$ q_{44}=J_{I_{d}}^{\mu_{d}} \cap J_{I_{d}}^{\mu_{c}} $$
(15)

gives the lightness quality score expressed as:

$$ q4 = \frac{n_{q_{44}}}{n_{t}} $$
(16)

where \(n_{q_{44}}\) is the number of dark and bright pixels common to both FGMG and FCMG. The disparity between μd and μc determines the lightness quality score.

Total quality score

The total quality score Q is the weighted sum of the four quality scores:

$$ Q= w_{q_{1}}q_{1} + w_{q_{2}}q_{2} + w_{q_{3}}q_{3} + w_{q_{4}}q_{4} $$
(17)

where \(w_{q_{1}}, w_{q_{2}}, w_{q_{3}}, w_{q_{4}}\) are the perceptual weights assigned to luminance contrast, texture, texture contrast and lightness quality attributes, respectively.

Implementation

The algorithm was implemented in the MatLab computing environment. The flow chart of Fig. 1 and the images displayed in Fig. 2 describe the six steps to implement the algorithm. The first two steps, foreground extraction and pixel intensity rescaling are meant to normalize data from the three different sources.

Fig. 1
figure 1

The flow chart for post-acquisition quality evaluation of a brain MRI slice. Foreground FRG is extracted FRX from the test image TIM. The test image is rescaled REX so that its pixel intensity levels is between 0 and 1. Two feature images, local contrast feature image CIM and grayscale image GIM are extracted from the rescaled image RIM. Global thresholding transforms the feature images into four binary feature images (only two BCM and BGM of the four binary feature images are shown). Determination of quality attributes QAX gives luminance contrast, texture contrast, texture and lightness quality attributes (only two, FQA and SQA of the four quality attributes are shown). The quality attributes are determined by matching relevant combinations of the binary feature images. Computation of quality score QSX for each quality attribute gives luminance contrast, texture, texture contrast and lightness quality scores (only two, FQS and SQS, of the four quality scores are shown). The total quality score QA is the weighted sum of the scores assigned to each quality attribute

Fig. 2
figure 2

The different stages of the algorithm for post-acquisition quality evaluation of a brain MRI slice. a The test image has its (b) foreground extracted. c The test image in (a) has its pixel intensity levels rescaled to lie between 0 and 1. d Grayscale and contrast feature images are extracted from the test image. Duplicating the rescaled image in (c) extracts the grayscale image at no computational cost. e, f, g and h are the four binary feature images generated by using the first moments of the feature images in (c) and (d) as global thresholds. i Luminance contrast, texture, texture contrast, lightness and total quality scores are computed by matching relevant pairs of the feature images in (e), (f), (g) and (h)

Step 1 - Foreground extraction

The first task in the implementation of the algorithm is to extract FRX the foreground FRG shown in Fig. 2b. The foreground pixels describe the actual anatomical structures of the test image TIM shown in Fig. 2a. Foreground extraction is a very critical step because the number of foreground pixels are required as input at step 4 where the binary feature images are generated and at step 6 for the computation of the quality scores.

Step 2 - Intensity rescaling

The intensity level of the test image TIM is rescaled REX to lie between 0 and 1 so that the rescaled test image RIM in Fig. 2c can be regarded as a blurred version of an ideal binary image according to the novel work in [68].

Step 3 - Feature image extraction

The rescaled test image RIM is convolved with a local range filter to extract FEX local contrast feature image CIM shown in Fig. 2d. The algorithm is sensitive to the size of filter. Larger filter size causes loss of fine details while smaller filter size will result in loss of spatial coherence in the filtered image [69]. For the aforementioned reasons and based on our experience during performance evaluation of the proposed algorithm we recommended filter width of 3×3 pixels, 5×5 pixels and 7×7 pixels for images with dimension comparable to 250×250 pixels, 350×350 pixels and 450×450 pixels, respectively. The mean of the local contrast image is computed with reference to the foreground pixels. The gray level feature image GIM is extracted at no computational cost by making a duplicate copy of the rescaled test image shown in Fig. 2c.

Step 4 - Binary feature images

According to Eqs. 2 - 5 four binary feature images (FGMG, FCMG, FCMC and FGMC) shown in Fig. 2e, f, g and h, respectively are generated from the two feature images; CIM and GIM. In the flow chart the four binary feature images are represented with two symbols; BCM and BGM.

Step 5 - Quality attribute

Quality attributes QAX of the test image are determined from similarity matching of relevant pairs of binary feature images generated in step 4 according to Eqs. 9, 11, 13 and 15. In the flow chart two symbols FQA, SQA are used to represent the four quality attributes.

Step 6 - Quality score

Quality score QSX for each quality attribute displayed in Fig. 2i is computed according to Eqs. 10, 12, 14 and 16. Two symbols FQS, SQS in the flow chart represent the quality scores. The total quality score QA is the weighted sum of each quality attribute. The perceptual weight assigned to each quality attribute is arbitrary but is based on [70] which reports that texture contrast quality attribute contributes approximately 10 times more to the generation of saliency in natural scenes than luminance contrast. Throughout the study the weight assigned to each quality attribute was fixed as follows; \(w_{q_{1}}=0.1\), \(w_{q_{2}}=0.1\), \(w_{q_{3}}=0.7\) and \(w_{q_{4}}=0.1\).

Results

The images and the tables in this section demonstrate some results in the six different categories of the objective and subjective performance evaluation of our proposed quality evaluation method.

Good Quality T2 MRI Volume Data

Six slices in T2 weighted MRI volume data from NeuroRx and BrainCare are shown in Figs. 3a – f and 4a – f, respectively. The plots in Figs. 3g and 4g are the luminous contrast, texture, texture contrast, lightness and total quality scores for 14 and 18 successive slices in the MRI volume data.

Fig. 3
figure 3

Six slices with indices (a) 1, (b) 4, (c) 6, (d) 9, (e) 11 and (f) 14 in a T2 weighted MRI volume data from NeuroRx Research Inc, (g) luminance contrast, texture, texture contrast, lightness and total quality scores of 14 successive slices in the MRI volume data

Fig. 4
figure 4

Six slices with indices (a) 1, (b) 4, (c) 8, (d) 11, (e) 15 and (f) 18 in a T2 weighted MRI volume data from BrainCare Oy, (g) luminance contrast, texture, texture contrast, lightness and total quality scores of 18 successive slices in the MRI volume data

Good Quality T1 MRI Volume Data

The images in Figs. 5a – f and 6a – f are slices in T1 weighted MRI volume data from NeuroRx and ADNI, respectively. Their respective luminous contrast, texture, texture contrast, lightness and total quality scores are shown in Figs. 5h and 6h, respectively.

Fig. 5
figure 5

Six slices with indices (a) 1, (b) 4, (c) 6, (d) 9, (e) 11 and (f) 14 in a T1 weighted MRI volume data from NeuroRx Research Inc., (g) luminance contrast, texture, texture contrast, lightness and total quality scores of 14 successive slices in the MRI volume data

Fig. 6
figure 6

Six slices with indices (a) 1, (b) 4, (c) 8, (d) 11, (e) 13 and (f) 17 in a T1 MPRAGE pulse sequence MRI volume data from ADNI, (g) luminance contrast, texture, texture contrast, lightness and total quality scores of 17 successive slices in the MRI volume data

Circular Blur

The image in Fig. 7a is a slice from a T2 weighted MRI volume data from BrainCare. The images shown in Fig. 7b – f are the same image in Fig. 7a but were blurred with circular averaging filter of radius 5 pixels, 8 pixels, 10 pixels, 13 pixels and 15 pixels, respectively. The luminous contrast, texture, texture contrast, lightness and total quality scores for blur levels from 1 pixels to 15 pixels are shown in Fig. 7g.

Fig. 7
figure 7

a A T2 weighted slice degraded by circular averaging filter of radius (b) 5, (c) 8, (d) 10, (e) 13 and (f) 15 pixels. g variation of the luminance contrast, texture, texture contrast, lightness and total quality scores with blur levels increasing from 1 to 15

Motion blur

A slice in a T2 weighted MRI volume data from BrainCare is shown in Fig. 8a. Its motion blurred versions are shown in Fig. 8b – f for motion blur levels of 4 pixels, 7 pixels, 9 pixels, 12 pixels and 15 pixels, respectively. The plot of the motion blur levels from 1 pixels to 15 pixels versus luminous contrast, texture, texture contrast, lightness and total quality scores are displayed in Fig. 8g.

Fig. 8
figure 8

a A T2 weighted slice and its degraded versions at motion blur levels (b) 5, (c) 8, (d) 10, (e) 13 and (f) 15 pixels, (g) variation of the luminance contrast, texture, texture contrast, lightness and total quality scores with blur levels increasing from 1 to 15

Noise

The image in Fig. 9a is a slice in the T2 weighted volume data from BrainCare. Its Rician noise degraded versions are shown in Fig. 9b – f for noise levels of 5, 8, 10, 13 and 15%, respectively. The plot of luminous contrast, texture, texture contrast, lightness and total quality scores for noise levels from 1 to 15 are displayed in Fig. 9g.

Fig. 9
figure 9

a A T2 weighted slice and its degraded versions at Rician noise levels (b) 5, (c) 8, (d) 10, (e) 13 and (f) 15 percent, (g) variation of the luminance contrast, texture, texture contrast, lightness and total quality scores with noise levels increasing from 1 to 15

Bias Fields

Six images shown in Fig. 10a – f are slices in a T1 weighted MRI volume data from NeuroRx. They were originally degraded with different configurations of bias fields during acquisition stage. The luminous contrast, texture, texture contrast, lightness and total quality scores for 14 successive slices in the volume data is shown in Fig. 10g.

Fig. 10
figure 10

Six slices with indices (a) 1, (b) 4, (c) 6, (d) 9, (e) 11, (f) 14 in a T1 weighted MRI volume data degraded by different configurations of bias fields, (g) luminance contrast, texture, texture contrast, lightness and total quality scores for each slice in the volume data

Validation of Results

Results from the subjective evaluation of our proposed method are tabulated in Tables 2, 3, 4 and 5. In Table 2 are the results for T2 and T1 MRI volume data without perceived distortion as well as T1 volume data that were originally acquired with bias fields. Tables 3, 4 and 5 are the results for blurring with circular averaging filter, blurring with motion blur and degradation with Rician noise, respectively.

Table 2 Validation results for T2 and T1 MRI images without perceived distortion and T1 MRI images degraded by bias fields
Table 3 Validation results for T2 MRI images degraded by blurring with circular averaging filter
Table 4 Validation results for T2 MRI images degraded by blurring with circular motion blur
Table 5 Validation results for T2 MRI sequence images degraded by Rician noise

Discussion

Evaluation across good quality T2 MRI images

The plots in Figs. 3 and 4 show variation in the quality scores across slices in the T2 weighted MRI volume data from NeuroRx and BrainCare, respectively. The luminance contrast, texture. texture contrast and lightness quality scores for the images from NeuroRx vary from 0.85 to 0.90, 0.65 to 0.70, 0.40 to 0.50 and 0.50 to 0.60, respectively. Corresponding quality scores for the images from BrainCare are 0.90 to 0.95, 0.80 to 0.85, 0.65 to 0.75 and 0.90 to 0.95. The results show that our proposed method objectively evaluate the variations in image quality in the slices contained in perceived good quality T2 weighted MRI volume data.

Evaluation across good quality T1 MRI images

The plot in Fig. 5 show that the average luminance contrast, texture. texture contrast, lightness and total quality scores of the slices in the NeuroRx T1 MRI volume data are 0.80, 0.45, 0.65, 0.70 and 0.50, respectively. Corresponding quality scores for the same type of MRI volume data from ADNI are 0.82, 0.65, 0.70, 0.80 and 0.70, respectively. Since both data were tagged as good quality images before the experiment, it can be said that our proposed method demonstrated good objective evaluation across slices in perceived good quality T1 weighted MRI data.

Evaluation across poor quality T1 MRI images

The average luminance contrast, texture. texture contrast, lightness and total quality scores for the images degraded by different configurations of bias fields are 0.75, 0.30, 0.7, 0.75 and 0.42, respectively (See Fig. 10). A cursory visualization of the six slices in Fig. 10 shows that there is contrast between the different anatomical structures but there is evidently loss of edges which is a measure of details that describe each anatomical structure within the image. The texture quality score of 0.30 recorded by our proposed method can be said to be a very good objective evaluation of the presence of bias field in the image. The presence of bias field can cause an automated image analysis system to erroneously misclassify the structures in the image. The texture quality index of 0.3 recorded by our proposed method can draw the attention of image analysis experts to carry out restoration on the degraded images before further processing.

Evaluation across different levels of distortion

The plot in Fig. 7 shows that the total quality score successively decrease from 0.95 to 0.15 for circular blur levels that vary from 1 to 15 in a T2 weighted slice. For the same slice in Fig. 8 there is an initial increase of total quality score for motion blur that increase from 1 to 3. This initial increase in quality score can be attributed to extraneous features which mimic image details during the initial introduction of motion blur. Thereafter there is successive decrease in quality score from 0.98 to 0.40 for blur levels that vary from 3 to 15. The slice in Fig. 9a is the same slice shown in Figs. 7a and 8a but degraded by Rician noise level that vary from 1 percent to 15 percent. The plot in Fig. 9g show successive decrease in total quality score from 0.85 to 0.42. These results show that our proposed method demonstrate promising performance in the evaluation of images with different perceptual quality.

Evaluation across different quality attributes

The slice in Fig. 9a is one of the slices in the T2 weighted volume data shown in Fig. 4. The plot in Fig. 4g indicates that, in the absence of distortion, the average total quality scores vary from from 0.8 to 0.85. In the presence of degradation by circular motion blur, motion blur and noise the average total quality scores varies from 0.95 to 0.15, 0.98 to 0.42 and 0.85 to 0.42. This is an indication that our proposed method can objectively evaluate different quality attributes for different levels of distortion in an image.

Correlation with subjective evaluation by human observers

The validation results in Tables 2, 3, 4 and 5 show that our proposed objective method has good correlation with human visual perception. Correlation between the scores assigned by the observers varies with the image sequence and the level of distortion in the images. The validation results shows that there is a trend of better agreement between the human observers at lower levels of distortion than at higher levels of distortion. An example is T2 MRI images degraded by blurring with circular averaging filter shown in Table 3. In the absence of distortion, the inter-observer correlation coefficient is 0.85. The corresponding inter-observer coefficients for 5, 10 and 15 distortion levels are 0.8, 0.75 and 0.65, respectively. In Table 2 the correlation coefficient of ρ≥0.80 for MRI volume data without perceived distortion show that there is very good correlation between quality scores recorded by our proposed method and the quality scores assigned by human observers. There is high correlation of 0.75 for validation result of T1 weighted volume data that were originally acquired with bias fields. The validation results table show that for different levels of circular blur, motion blur and Rician noise, the inter-observer correlation coefficient ρ≥0.65, and the objective-subjective correlation coefficient is ρ≥0.60.

Standardization of quality metric

There is a clearly defined lower and upper limit of quality index. The lower limit is 0 for an extremely degraded image and upper limit of 1 for an ideal image. Quality index for a real MRI image lies between 0 and 1. This quality evaluation index is applied across images derived from different clinical trial sites, different scanners and different acquisition protocols. Thus a remarkable characteristic of our proposed method is the standardization of quality metric.

Cut-off quality index

Our proposed quality metric system predict different quality scores for different MRI sequences. There are three reasons to suggest that the cut-off quality metrics to determine images of acceptable quality should be flexibly applied across different MRI sequences. First, different MRI sequences such as T2 and T1 that are without perceived distortion reveal different levels of structural information. Second, different MRI sequence images are acquired for different tasks. This leads to the philosophy of task-based quality evaluation [71, 72]. Third, it is more realistic to compare the quality of similar MRI sequence images acquired for the same task.

The quality measure is applied slice-wise across the MRI volume data. Generally, quality scores of slices contained in a MRI volume data lies within a narrow band of quality index. Based on our experience during the performance evaluation of our proposed algorithm and with specific importance to the opinion of human observers we recommend a cut-off quality index of 0.40 and 0.45 for T1 and T2 weighted MRI images, respectively. Although our methodology allows the computation of the total quality score for individual slices we hereby emphasize that the cut-off quality metric does not suggest rejection of individual slices. The recommended cut-off quality metric is the total quality score computed from the average quality scores of the slices contained in the MRI volume data.

Absence of comparative performance evaluation

Three characteristics of existing methods makes it difficult to include comparative performance evaluation in this report. First, existing methods adopt different distortion models. Second, there are no clearly defined lower and upper limits of quality indices. Third, there are many definitions of the popular quality models such as SNR.

Limitations of proposed method and future work

Three characteristics of our proposed method limit its efficacy in real-life scenarios. First, is the two-tissue class model we adopt for MRI images. It excludes the ventricular system. The model assumes that the brain consists of only white matter and cortical gray matter structures. The fixed perceptual weights assigned to the four quality attributes throughout the study is a theoretical approach that is yet to be validated. The third limitation is that our proposed method predict image quality based on the assumption that distortion process have same effect on the different structures of the brain.

Future work will adopt a new model which accounts for all the three major anatomical structures of the brain. Segmentation algorithm will be incorporated to delineate the boundaries of the three major anatomical structures so that quality score prediction will be assigned to region-of-interest. The perceptual weights assigned to each quality attributes will be refined based on the subjective scores from human observers. We will also explore the use of higher moments as a basis to describe image quality attributes.

Conclusions

There is increasing clinical interest in the use of brain MRI images for the study of human anatomy, treatment and diagnosis of diseases as well as the clinical trials of drugs for the treatment of neurological diseases. Post-acquisition image quality evaluation is necessary to re-evaluate and standardize the quality of brain MRI images acquired from different clinical trial sites across the globe. No-reference objective image quality assessment is highly desired in environment where large volumes of MRI data are processed. We propose a new method to evaluate the quality of brain MRI images. The proposed method will be suitable for fully automated environments because processing of the quality metrics is on binary images. Experimental results demonstrates that our proposed method had good correlation with human visual judgement and gives fairly accurate quality evaluation within and across good quality images and different levels of degradation.

Abbreviations

ADNI:

Alzheimer’s disease neuroimaging initiative

ANOVA:

Analysis of variance

BCM:

Binary feature image extracted from the local contrast feature image

BGM:

Binary feature image extracted from the grayscale feature image

CIM:

Contrast feature image

CNR:

Contrast-to-noise ratio

FEX:

Feature extraction

FRG:

Foreground extraction

FRX:

Foreground extraction

FQA:

Luminance contrast and lightness quality attributes

FQS:

Luminance contrast and lightness quality attributes

GIM:

Grayscale feature image

MPRAGE:

Magnetization-prepared rapid gradient echo

MRI:

Magnetic resonance imaging

QA:

Total quality score

QAX:

Quality attribute extraction

QSX:

Quality score extraction

REX:

Intensity rescaling operation

RIM:

Intensity rescaled image

RMSE:

Root means square error

SNR:

Signal-to-noise ratio

SQA:

Texture and texture contrast quality attributes

SQS:

Texture and texture contrast quality attributes

TIM:

Test image

T2:

Transverse relaxation

T1:

Longitudinal relaxation

References

  1. Arnold D, Stone L. The clinical usefulness of routine mri assessment. Int MS J. 2010; 17(2):58–63.

    Google Scholar 

  2. Lillemark L, Sørensen L, Pai A, Dam EB, Nielsen M. Brain regions relative proximity as marker for alzheimers disease based on structural mri. BMC Med Imaging. 2014; 14(1):21.

    Article  Google Scholar 

  3. Oreja-Guevara C. Overview of magnetic resonance imaging for management of relapsing- remitting multiple sclerosis in everyday practice. Eur J Neurol. 2015; 22(S2):22–27.

    Article  Google Scholar 

  4. Zhang Y, Wang S, Phillips P, Dong Z, Ji G, Yang J. Detection of alzheimer’s disease and mild cognitive impairment based on structural volumetric mr images using 3d-dwt and wta-ksvm trained by psotvac. Biomed Signal Proc Control. 2015; 21:58–73.

    Article  CAS  Google Scholar 

  5. Frisoni GB, Fox NC, Jack CR, Scheltens P, Thompson PM. The clinical use of structural mri in alzheimer disease. Nat Rev Neurol. 2010; 6(2):67–77.

    Article  Google Scholar 

  6. Collins D, Maranzano J, Li T, Arnold D. Robust automatic segmentation of hippocampus from multisite mri data. Alzheimers Dement: J Alzheimers Assoc. 2009; 5(4):103.

    Article  Google Scholar 

  7. Zhang Y, Wang S, Dong Z. Classification of alzheimer disease based on structural magnetic resonance imaging by kernel support vector machine decision tree. Prog Electromagn Res. 2014; 144:171–84.

    Article  Google Scholar 

  8. Tabelow K, Clayden JD, De Micheaux PL, Polzehl J, Schmid VJ, Whitcher B. Image analysis and statistical inference in neuroimaging with r. NeuroImage. 2011; 55(4):1686–93.

    Article  CAS  Google Scholar 

  9. Clayden JD. Imaging connectivity: Mri and the structural networks of the brain. Funct Neurol. 2013; 28(3):197.

    PubMed  PubMed Central  Google Scholar 

  10. Wang S, Phillips P, Yang J, Sun P, Zhang Y. Magnetic resonance brain classification by a novel binary particle swarm optimization with mutation and time-varying acceleration coefficients. Biomed Eng/Biomed Tech. 2016; 61(4):431–41.

    Article  CAS  Google Scholar 

  11. Zhang Y, Yang J, Yang J, Liu A, Sun P. A novel compressed sensing method for magnetic resonance imaging: exponential wavelet iterative shrinkage-thresholding algorithm with random shift. Inter J Biomed Imaging. 2016; 2016:10. Article ID 9416435.

    Article  Google Scholar 

  12. McDonald WI, Compston A, Edan G, Goodkin D, Hartung H-P, Lublin FD, McFarland HF, Paty DW, Polman C, Reingold SC, et al. Recommended diagnostic criteria for multiple sclerosis: guidelines from the international panel on the diagnosis of multiple sclerosis. Ann Neurol. 2001; 50(1):121–27.

    Article  CAS  Google Scholar 

  13. Jack CR, Albert MS, Knopman DS, McKhann GM, Sperling RA, Carrillo MC, Thies B, Phelps CH. Introduction to the recommendations from the national institute on aging-alzheimer’s association workgroups on diagnostic guidelines for alzheimer’s disease. Alzheimer’s Dementia. 2011; 7(3):257–62.

    Article  Google Scholar 

  14. Swanton J, Fernando K, Dalton CM, Miszkiel KA, Thompson AJ, Plant GT, Miller DH. Modification of mri criteria for multiple sclerosis in patients with clinically isolated syndromes. J Neurol, Neurosurg Psychiatry. 2006; 77(7):830–33.

    Article  CAS  Google Scholar 

  15. Montalban X, Tintoré M., Swanton J, Barkhof F, Fazekas F, Filippi M, Frederiksen J, Kappos L, Palace J, Polman C, et al. Mri criteria for ms in patients with clinically isolated syndromes. Neurol. 2010; 74(5):427–34.

    Article  CAS  Google Scholar 

  16. Polman C, Reingold SC, Edan G, Filippi M, Hartung H-P, Kappos L, Lublin FD, Metz LM, McFarland HF, O’Connor PW, et al. Diagnostic criteria for multiple sclerosis: 2005 revisions to the mcdonald criteria. Ann Neurol. 2005; 58(6):840–46.

    Article  Google Scholar 

  17. Polman C, Reingold SC, Banwell B, Clanet M, Cohen JA, Filippi M, Fujihara K, Havrdova E, Hutchinson M, Kappos L, et al. Diagnostic criteria for multiple sclerosis: 2010 revisions to the mcdonald criteria. Ann Neurol. 2011; 69(2):292–302.

    Article  Google Scholar 

  18. Filippi M, Rocca MA, Ciccarelli O, De Stefano N, Evangelou N, Kappos L, Rovira A, Sastre-Garriga J, Tintorè M., Frederiksen JL, et al. Mri criteria for the diagnosis of multiple sclerosis: Magnims consensus guidelines. Lancet Neurol. 2016; 15(3):292–303.

    Article  Google Scholar 

  19. Grech-Sollars M, Hales PW, Miyazaki K, Raschke F, Rodriguez D, Wilson M, Gill SK, Banks T, Saunders DE, Clayden JD, et al. Multi-centre reproducibility of diffusion mri parameters for clinical sequences in the brain. NMR Biomed. 2015; 28(4):468–85.

    Article  Google Scholar 

  20. Van Horn JD, Toga AW. Multi-site neuroimaging trials. Curr Opin Neurol. 2009; 22(4):370.

    Article  Google Scholar 

  21. Chalavi S, Simmons A, Dijkstra H, Barker GJ, Reinders AS. Quantitative and qualitative assessment of structural magnetic resonance imaging data in a two-center study. BMC Med imaging. 2012; 12(1):1.

    Article  Google Scholar 

  22. Friedman L, Glover GH, Krenz D, Magnotta V, BIRN TF. Reducing inter-scanner variability of activation in a multicenter fmri study: role of smoothness equalization. Neuroimage. 2006; 32(4):1656–68.

    Article  Google Scholar 

  23. Friedman L, Glover GH. Report on a multicenter fmri quality assurance protocol. J Magn Reson Imaging. 2006; 23(6):827–39.

    Article  Google Scholar 

  24. Traboulsee A, Simon J, Stone L, Fisher E, Jones D, Malhotra A, Newsome S, Oh J, Reich D, Richert N, et al. Revised recommendations of the consortium of ms centers task force for a standardized mri protocol and clinical guidelines for the diagnosis and follow-up of multiple sclerosis. Am J Neuroradiol. 2016; 37(3):394–401.

    Article  CAS  Google Scholar 

  25. Ashton E. Quantitative mr in multi-center clinical trials. J Magn Reson Imaging. 2010; 31(2):279–88.

    Article  Google Scholar 

  26. Krupinski EA. The importance of perception research in medical imaging. Radiat Med-Med Imaging Radiat Oncol. 2000; 18(6):329–34.

    CAS  Google Scholar 

  27. De Stefano N, Giorgio A, Battaglini M, Rovaris M, Sormani M, Barkhof F, Korteweg T, Enzinger C, Fazekas F, Calabrese M, et al. Assessing brain atrophy rates in a large population of untreated multiple sclerosis subtypes. Neurol. 2010; 74(23):1868–76.

    Article  CAS  Google Scholar 

  28. Saloner D, Liu J, Haraldsson H. Mr physics in practice: How to optimize acquisition quality and time for cardiac mri. Magn Reson Imaging Clin N Am. 2015; 23(1):1.

    Article  Google Scholar 

  29. Hollingsworth KG. Reducing acquisition time in clinical mri by data undersampling and compressed sensing reconstruction. Phys Med Biol. 2015; 60(21):297.

    Article  Google Scholar 

  30. Woodard J, Carley-Spencer M. No-reference image quality metrics for structural mri. Neuroinformatics. 2006; 4(3):243–62.

    Article  Google Scholar 

  31. Mortamet B, Bernstein MA, Jack CR, Gunter JL, Ward C, Britson PJ, Meuli R, Thiran J-P, Krueger G. Automatic quality assessment in structural brain magnetic resonance imaging. Magn Reson Med. 2009; 62(2):365–72.

    Article  Google Scholar 

  32. Prieto F, Guarini M, Tejos C, Irarrazaval P. Metrics for quantifying the quality of mr images. In: Proceedings of the 17th Annual Meeting of ISMRM: 2009. p. 4696.

  33. Gedamu E, Collins D, Arnold D. Automated quality control of brain mr images. J Magn Reson Imaging. 2008; 28(2):308–19.

    Article  Google Scholar 

  34. Osadebey M, Pedersen M, Arnold D, Wendel-Mitoraj K. No-reference quality measure in brain mri images using binary operations, texture and set analysis. IET Image Process. 2017; 11(9):672–84.

    Article  Google Scholar 

  35. Pedersen M, Hardeberg JY. Full-reference image quality metrics: Classification and evaluation. Found Trends Comput Graph Vis. 2012; 7(1):1–80.

    Article  Google Scholar 

  36. Lin W, Kuo C-CJ. Perceptual visual quality metrics: A survey. J Vis Commun Image Represent. 2011; 22(4):297–312.

    Article  Google Scholar 

  37. Chow LS, Paramesran R. Review of medical image quality assessment. Biomed Signal Process Control. 2016; 27:145–54.

    Article  Google Scholar 

  38. Moraru L, Moldovanu SS, Obreja CD. A survey over image quality analysis techniques for brain mr images. Int J Radiol. 2015; 2(1):24–28.

    Article  Google Scholar 

  39. Wang Z, Bovik AC. Modern image quality assessment. Synth Lect Image Video Multimedia Process. 2006; 2(1):1–156.

    Article  CAS  Google Scholar 

  40. Leclaire A, Moisan L. No-reference image quality assessment and blind deblurring with sharpness metrics exploiting fourier phase information. J Math Imaging Vis. 2015; 52(1):145–72.

    Article  Google Scholar 

  41. Kupinski MA, Clarkson E. Objective assessment of image quality In: Kupinski MA, Barrett HH, editors. Small-Animal Spect Imaging. Boston: Springer: 2005.

    Chapter  Google Scholar 

  42. Miao J, Huang F, Narayan S, Wilson DL. A new perceptual difference model for diagnostically relevant quantitative image quality evaluation: A preliminary study. Magn Reson Imaging. 2013; 31(4):596–603.

    Article  Google Scholar 

  43. McCarthy G, Wood CC. Scalp distributions of event-related potentials: an ambiguity associated with analysis of variance models. Electroencephalogr Clin Neurophysiol/Evoked Potentials Sect. 1985; 62(3):203–08.

    Article  CAS  Google Scholar 

  44. Pedersen M, Bonnier N, Hardeberg JY, Albregtsen F. Attributes of image quality for color prints. J Electron Imaging. 2010; 19(1):011016–01101613.

    Article  Google Scholar 

  45. Bartleson C. The combined influence of sharpness and graininess on the quality of colour prints. J Photogr Sci. 1982; 30(2):33–38.

    Article  Google Scholar 

  46. Singh C, Ranade SK, Singh K. Invariant moments and transform-based unbiased nonlocal means for denoising of {MR} images. Biomed Signal Process Control. 2016; 30:13–24.

    Article  Google Scholar 

  47. Sun Y, Wen G, Wang J. Weighted spectral features based on local hu moments for speech emotion recognition. Biomed signal Process Control. 2015; 18:80–90.

    Article  Google Scholar 

  48. Abu-Mostafa YS, Psaltis D. Image normalization by complex moments. IEEE Trans Pattern Anal Mach Intell. 1985; PAMI-7:46–55. 1.

    Article  Google Scholar 

  49. Bujack R, Hotz I, Scheuermann G, Hitzer E. Moment invariants for 2d flow fields via normalization in detail. IEEE Trans Vis Comput Graph. 2015; 21(8):916–29.

    Article  Google Scholar 

  50. Honarvar B, Paramesran R, Lim C-L. Image reconstruction from a complete set of geometric and complex moments. Signal Process. 2014; 98:224–32.

    Article  Google Scholar 

  51. Wang S, Zhang Y, Zhan T, Phillips P, Zhang Y, Liu G, Lu S, Wu X. Pathological brain detection by artificial intelligence in magnetic resonance imaging scanning (invited review). Prog Electromagn Res. 2016; 156:105–33.

    Article  Google Scholar 

  52. Bouguila N, Almakadmeh K, Boutemedjet S. A finite mixture model for simultaneous high-dimensional clustering, localized feature selection and outlier rejection. Expert Syst Appl. 2012; 39(7):6641–56.

    Article  Google Scholar 

  53. Karakasis E, Amanatiadis A, Gasteratos A, Chatzichristofis SA. Image moment invariants as local features for content based image retrieval using the bag-of-visual-words model. Pattern Recogn Lett. 2015; 55:22–27.

    Article  Google Scholar 

  54. Rapantzikos K, Zervakis M, Balas K. Detection and segmentation of drusen deposits on human retina: Potential in the diagnosis of age-related macular degeneration. Med Image Anal. 2003; 7(1):95–108.

    Article  CAS  Google Scholar 

  55. Christodoulou CI, Michaelides SC, Pattichis CS. Multifeature texture analysis for the classification of clouds in satellite imagery. IEEE Trans Geosci Remote Sens. 2003; 41(11):2662–68.

    Article  Google Scholar 

  56. Coupé P., Manjón J. V., Gedamu E, Arnold D, Robles M, Collins D. Robust rician noise estimation for mr images. Med Image Anal. 2010; 14(4):483–93.

    Article  Google Scholar 

  57. Van Ngo K, Storvik JJ, Dokkeberg CA, Farup I, Pedersen M. Quickeval: a web application for psychometric scaling experiments. In: SPIE/IS&T Electronic Imaging. Bellingham: SPIE, the international society for optics and photonics: 2015. p. 93960.

  58. Streijl RC, Winkler S, Hands DS. Mean opinion score (mos) revisited: methods and applications, limitations and alternatives. Multimedia Syst. 2016; 22(2):213–27.

    Article  Google Scholar 

  59. Corder GW, Foreman DI. Nonparametric Statistics: A Step-by-step Approach.New Jersey: Wiley; 2014.

    Google Scholar 

  60. Peli E. Contrast in complex images. JOSA A. 1990; 7(10):2032–40.

    Article  CAS  Google Scholar 

  61. Pu Y-F, Zhou J-L, Yuan X. Fractional differential mask: a fractional differential-based approach for multiscale texture enhancement. IEEE Trans Image Process. 2010; 19(2):491–511.

    Article  Google Scholar 

  62. Chen J, Wang B. High quality solid texture synthesis using position and index histogram matching. Vis Comput. 2010; 26(4):253–62.

    Article  Google Scholar 

  63. Khalvati F, Wong A, Haider MA. Automated prostate cancer detection via comprehensive multi-parametric magnetic resonance imaging texture feature models. BMC Med Imaging. 2015; 15(1):27.

    Article  Google Scholar 

  64. Holli KK, Harrison L, Dastidar P, Wäljas M, Liimatainen S, Luukkaala T, Öhman J, Soimakallio S, Eskola H. Texture analysis of mr images of patients with mild traumatic brain injury. BMC Med Imaging. 2010; 10(1):8.

    Article  Google Scholar 

  65. Skogen K, Schulz A, Dormagen JB, Ganeshan B, Helseth E, Server A. Diagnostic performance of texture analysis on mri in grading cerebral gliomas. Eur J Radiol. 2016; 85(4):824–29.

    Article  Google Scholar 

  66. Campisi P, Egiazarian K. Blind Image Deconvolution: Theory and Applications.Boca Raton: CRC press; 2007.

    Book  Google Scholar 

  67. Wilkinson MH, Schut F. Digital Image Analysis of Microbes: Imaging, Morphometry, Fluorometry and Motility Techniques and Applications.West Sussex: Wiley; 1998.

    Google Scholar 

  68. Tsai W-H. Moment-preserving thresolding: A new approach. Comput Vis Graph Image Process. 1985; 29(3):377–93.

    Article  Google Scholar 

  69. Lee JS. Digital image enhancement and noise filtering by use of local statistics. IEEE Trans Pattern Anal Mach Intel. 1980; PAMI-2(2):165–68.

    Article  Google Scholar 

  70. Parkhurst DJ, Niebur E. Texture contrast attracts overt visual attention in natural scenes. Eur J Neurosci. 2004; 19(3):783–89.

    Article  Google Scholar 

  71. Barrett HH, Yao J, Rolland JP, Myers KJ. Model observers for assessment of image quality. Proc Natl Acad Sci. 1993; 90(21):9758–65.

    Article  CAS  Google Scholar 

  72. Kupinski MA, Hoppin JW, Clarkson E, Barrett HH. Ideal-observer computation in medical imaging with use of markov-chain monte carlo techniques. JOSA A. 2003; 20(3):430–38.

    Article  Google Scholar 

Download references

Acknowledgements

Data used in the design of algorithm and the performance evaluation in this article were partly obtained from the Alzheimer’s Disease Neuroimaging Initiative (ADNI) database (adni.loni.usc.edu). As such, the investigators within the ADNI contributed to the design and implementation of ADNI and/or provided data but did not participate in analysis or writing of this report. A complete listing of ADNI investigators can be found at: http://adni.loni.usc.edu/wp-content/uploads/how_to_apply/ADNI_Acknowledgement_List.pdf

Funding

Marius Pedersen have been supported by the Research Council of Norway, project no. 247689 ’IQMED: Image Quality enhancement in MEDical diagnosis, monitoring and treatment’. Data collection and sharing for this project was funded by the Alzheimer’s Disease Neuroimaging Initiative(ADNI) (National Institutes of Health Grant U01 AG024904) and DOD ADNI (Department of Defense award number W81XWH-12-2-0012). ADNI is funded by the National Institute on Aging, the National Institute of Biomedical Imaging and Bioengineering, and through generous contributions from the following: AbbVie, Alzheimer’s Association; Alzheimer’s Drug Discovery Foundation; Araclon Biotech; BioClinica, Inc.; Biogen; Bristol-Myers Squibb Company; CereSpir, Inc.; Cogstate; Eisai Inc.; Elan Pharmaceuticals, Inc.; Eli Lilly and Company; EuroImmun; F. Hoffmann- La Roche Ltd and its affiliated company Genentech, Inc.; Fujirebio; GE Healthcare; IXICO Ltd.; Janssen Alzheimer Immunotherapy Research & Development, LLC.; Johnson & Johnson Pharmaceutical Research & Development LLC.; Lumosity; Lundbeck; Merck & Co., Inc.; Meso Scale Diagnostics, LLC.; NeuroRx Research; Neurotrack Technologies; Novartis Pharmaceuticals Corporation; Pfizer Inc.; Piramal Imaging; Servier; Takeda Pharmaceutical Company; and Transition Therapeutics. The Canadian Institutes of Health Research is providing funds to support ADNI clinical sites in Canada. Private sector contributions are facilitated by the Foundation for the National Institutes of Health(www.fnih.org). The grantee organization is the Northern California Institute for Research and Education, and the study is coordinated by the Alzheimer’s Therapeutic Research Institute at the University of Southern California. ADNI data are disseminated by the Laboratory for Neuro Imaging at the University of Southern California.

Availability of data and materials

The data that support the findings of this study are available from NeuroRx research Inc., BrainCare Oy and the ADNI but restrictions apply to the availability of these data, which were used under license for the current study, and so are not publicly available. Data are however available from the authors upon reasonable request and with permission of the aforementioned organizations.

Author information

Authors and Affiliations

Authors

Contributions

MEO carried out the design and implementation of the proposed metric system. MP contributed to the technical development, analysis and interpretation of the results. DLA and KEM were involved in data analysis as well as interpretation of the experimental results. All authors have been involved in drafting and revising the manuscript and approved the final version to be published. All authors read and approved the final manuscript.

Corresponding author

Correspondence to Marius Pedersen.

Ethics declarations

Ethics approval and consent to participate

This study which focus on quality evaluation was based on the analysis of anonymized and retrospectively acquired data. Administrative permissions to access the anonymized data were separately granted by the three organizations that provided the data. ADNI gave permission to access data on April 8, 2017. NeuroRx Research Inc. gave permission to access data on January 7, 2015. BrainCare Oy. gave permission to access data on December 27, 2015. Ethics approval was deemed unnecessary according to the regulations of the Norwegian Regional Committees for Medical and Health Research Ethics (REK) published on 12 January 2012 and available at (https://helseforskning.etikkom.no/reglerogrutiner/soknadsplikt/sokerikkerek?p_dim=34999&_ikbLanguageCode=us).

Consent for publication

Not Applicable.

Competing interests

The authors declare that they have no competing interests.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Additional information

Authors’ Information

Michael Osadebey obtained his master’s degree with distinction in biomedical engineering from Tampere University of Technology, Finland, in 2009. He was a Ragnar Granit Research grant recipient from October 2009 to December 2009. Michael obtained a PhD in engineering and computer science from Concordia University, Montreal, Canada, in 2015. His PhD study was focused on the processing of MRI images of the brain. He is a MRI Reader at NeuroRx Research Inc. a Montreal-based clinical research organization (CRO). His duties at NeuroRx include application of advanced image analysis software in the reading of MRI data of neurological diseases patients undergoing clinical trial drug treatment.

Marius Pedersen received his BSc degree in computer engineering and MiT degree in media technology both from Gj"ø"vik University College, Norway, in 2006 and 2007, respectively. He completed his PhD program in color imaging from the University of Oslo, Norway, sponsored by Oce in 2011. He is a professor in the faculty of computer science at the Norwegian University of Science and Technology (NTNU), Gj"ø"vik, Norway. He is also the director of the Norwegian Colour and Visual Computing Laboratory (Colourlab) at same university. His work is centered on subjective and objective image quality.

Douglas Arnold is the director of Magnetic Resonance Spectroscopy Lab, McGill University, Montreal, Canada, and the president/CEO NeuroRx Research Inc., a Montreal-based CRO. He is a neurologist with special expertise in MRI. His personal research interests are centered on the use of advanced neuroimaging techniques to assess the pathological evolution of multiple sclerosis and Alzheimer’s disease and to quantify the effects of therapy on these diseases.

Katrina Wendel-Mitoraj obtained her PhD in biomedical engineering from Tampere University of Technology in 2010. Her PhD study was focused on electroencephalography electrode sensitivity distributions. She is the CEO and founder of BrainCare Oy. BrainCare Oy is a Tampere University of Technology spin-off company founded in 2013 to deliver personalized solutions to improve the quality of life of epilepsy patients. The organization recently concluded clinical trials for an innovative mobile application and supporting solutions for long-term monitoring for epileptic patients.

Rights and permissions

Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Osadebey, M., Pedersen, M., Arnold, D. et al. Standardized quality metric system for structural brain magnetic resonance images in multi-center neuroimaging study. BMC Med Imaging 18, 31 (2018). https://doi.org/10.1186/s12880-018-0266-4

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s12880-018-0266-4

Keywords