Usefulness of grayscale inverted images in addition to standard images in digital mammography

Abstract

Background

Mammography is essential for early diagnosis of breast cancer, which is the most common type of cancer in females that is associated with a high mortality rate. We investigated whether evaluation of the grayscale inverted images of mammograms would aid in increasing the diagnostic sensitivity of the mammographic imaging technique.

Methods

Our study included 636 mammograms of 159 women who had undergone digital mammography. Standard, grayscale inverted, and standard plus grayscale inverted images were sequentially examined three times, at 15-day intervals, for the presence or assessment of pathological changes in the skin, calcification, asymmetric density, mass lesions, structural distortions, and intramammary and axillary lymph nodes. To determine whether grayscale inverted image assessment improved detection rates, the results of the three assessment modes were compared using Cochran’s Q test and the McNemar test (p < 0.05 was considered statistically significant).

Results

The average age of 159 patients was 50.4 years (range, 35–80 years). There were significant differences among the three assessment modes with respect to calcification and intramammary lymph nodes (p < 0.05); however, no significant differences were observed for the detection of other parameters.

Conclusions

Assessment of grayscale inverted images in addition to standard images facilitates the detection of microcalcification.

Peer Review reports

Background

Breast cancer is the most frequent women’s cancer, with high mortality rates in Turkey and other parts of the world. In Turkey, its incidence is 40.6 per 100,000 women (http://www.kanser.gov.tr/Dosya/ca_istatistik/2009kanseraporu). Early diagnosis can decrease the likelihood of mortality and requires scanning. It has been reported that scanning of women over 50 years of age reduces breast cancer-related mortality rates by 3–36% [1, 2].

Mammography is the most effective screening method for breast cancer [3]. However, mammographic assessment is difficult and requires experience owing to the complex structure of mammary tissue. The false negative range in mammographic detection of breast cancer is 10–25%. Dense mammary structure, inferior mammogram quality, overlooked details of early breast cancers, and fatigue or carelessness of the radiologist may account for the high percentage of false negatives [4]. To lower this percentage, a double-reading system is recommended [5] but is costly and requires two professional readers and is thus impractical. Computer-aided detection (CAD) is an alternative method, especially in busy crowded centers that focus on mammography [6].

Unfortunately, not all mammography centers facilitate double reading or CAD system. In our center, one radiologist assesses the data obtained via digital mammography without CAD. In view of the fact that digital mammography is a digitalized system, digitalization allows different representations of the images. Therefore, in this paper we studied if the diagnostic assessment of grayscale inverted mammograms improved compared to a standard representation of mammographic images when double-reading or CAD is impossible.

Methods

Our study included 636 mammograms of 159 women who had undergone mammography at Dr. Faruk Sükan Obstetrics and Pediatrics Hospital between January 5 and February 4, 2015. Scanning was performed for diagnostic purposes, in the mediolateral oblique and cranio-caudal views, and using a digital mammography device, which has ring shaped gantry, 24X30 cm Amorphous Selenium detector, Tungsten anode tube (Giotto Image 3D, IMS, Bologna, Italy). This study was approved by the ethics committee of Necmettin Erbakan University. Mammograms were evaluated by an experienced radiologist. Mammography was performed according to the Breast Imaging Reporting and Data System (BIRADS) as recommended by the American College of Radiology [7].

Initially, the parenchymal patterns on mammograms were evaluated. and categorized as follows: type A, the whole breast consists predominantly of fat tissue; type B, the whole breast contains fibroglandular densities as well as fat tissue; type C, the density of the breast parenchyma is heterogeneous; and type D, the breast parenchyma is highly dense [7]. Pathological changes in the skin (increased thickness, mass, or shrinkage), calcification, asymmetric density, mass lesion, structural distortion, and intramammary and axillary lymph nodes were then assessed and scored as present or absent. Calcifications were classified as macro or micro (<0.5 mm) regardless of their morphological properties and distribution patterns. Asymmetric density included both focal and regional asymmetric density detected in either or both views. Densities with a three-dimensional structure and observed in both views were defined as mass. Structural distortion was defined as shrinkage withdrawal without mass. Centrically radiolucent lymph nodes with regular margins and oval-shaped nodular opacities were classified as intramammary (present in the breast parenchyma) or axillary (present in the axillary region).

Initial assessments were performed on standard images (the std method). Results were reported as present or absent or scored by using the BIRADS. In the final BIRADS assessment, BIRADS scores were defined as follows: 0, more examination needed; 1, normal; 2, benign; 3, probably benign; 4, suspicious; 5, highly suspicious; and 6, malignant [7]. At least 15 days after the first assessment, the same mammograms were evaluated as above using only the grayscale inverted images (the gsi method). At least 15 days later, a third assessment was performed using both standard and grayscale inverted images (the std + gsi method) (Fig. 1a-c). Cochran’s Q test was used to compare the results obtained with the three types of image evaluation. The McNemar multiple comparison test was used to compare the different methods (p < 0.05 was considered statistically significant). The intraobserver concordance among three assessment methods was measured by using the Fleiss’ Kappa test.

Fig. 1
figure 1

a-c Standard and grayscale inverted image. Arrows indicate the early vascular macrocalcification, intramammary lymph node and nodular density with smooth contour respectively

Results

The average age of 159 patients in our study was 50.4 years (range, 35–80 years; standard deviation, 7.81 years). The parenchymal patterns on standard images were as follows: type A, 6 (3.8%) women; type B, 68 (42.8%) women; type C, 54 (34%) women; and type D, 31 (19.5%) women. According to the BIRADS, 45 (28.3%) mammograms were categorized as 0, 39 (24.5%) as 1, 74 (46.5%) as 2, and 1 (0.6%) as 3 (Table 1). There were no cases with BIRADS 4,5 and 6 score.

Table 1 Parenchyma pattern of mammograms and BIRADS classification

The intraobserver concordance for the results of the three assessment methods (standard, grayscale inverted, and std + gsi) is shown in Table 2. There were statistically significant differences in presence/absence of macrocalcification, microcalcification, and intramammary lymph nodes among the methods (p < 0.05) (Table 3), but not in the other parameters (Table 4).

Table 2 Intraobserver concordance assessment
Table 3 Significant differences between the image types
Table 4 Insignificant differences between the image types

Among the three assessment methods, a significant difference was observed in the detection of macrocalcification. In std + gsi images more macrocalcification (93 images) were detected than in standard images (84 images) (Cochran’s Q = 8.375, p = 0.015). There was also a significant difference in microcalcification detection among the methods. Multiple comparisons between the methods showed that microcalcification was more often observed in grayscale inverted (53 images) and std + gsi images (81 images) than in standard images (31 images). Moreover, microcalcification was the most often observed in std + gsi images (81 images) (Cochran’s Q = 68.509, p < 0.001). Lastly, there was a significant difference in the detection of intramammary lymph nodes among the 3 assessment methods. In std + gsi images more intramammary lymph nodes (59 images) were detected than in standard images (49 images) (Cochran’s Q =6.461, p = 0.039).

Discussion

In Turkey, yearly mammographic scanning beginning at 40 years of age is recommended [8]. The double-reading system is preferred as it decreases the false-negative rate in scanning mammograms. In previous studies, this system increased breast cancer detection rates by 4–15% [5, 9] and better detected small breast cancers owing to its relatively high sensitivity [10]. However, the double-reading system is impractical because it increases the cost of the mammogram and requires two readers. Moreover, it has been shown to decrease the positive estimation value and increase the recall rate and the patient’s anxiety level [11].

The CAD system is an alternative to the double-reading system, especially in busy centers [6]. The CAD system directs the radiologist’s attention to suspicious regions, thereby aiding detection of the lesion [12] Although the CAD system has existed for 10 years, its additive contribution to routine scanning is controversial [13]. In the study by Fiona et al., the double-reading system and CAD system had similar detection rates [6]. However, another study suggested that the CAD system improved diagnostic performance and reduced assessment time [14]. An additional benefit of the CAD system is its capacity to detect microcalcification even when used by radiologists with little experience [15, 16].

Mammographic sensitivity is lowest during the premenopausal period, at which time the fibroglandular tissue is especially dense. In this situation, digital mammography more efficiently detects lesions than does conventional mammography [17]. Digital mammography can also process the data in various formats because its image forming and display phases are different.

We assessed the usefulness of grayscale inverted images for detecting mammary abnormalities using digital mammography facilities when double reading or CAD system is not feasible. To our knowledge, there are previous studies in the medical literature that has performed such an assessment [1820]. But we have not found a study on mammograms. Including grayscale inverted images in mammogram assessments is not routine; however, it is not a difficult application (standard images can be converted to grayscale inverted images by the touch of a button in digital systems). However, evaluation of the negative image, as well as the standard image, somewhat prolongs the assessment time.

Our study did not examine the distribution pattern, number, or morphological properties of microcalcifications except for their size. Therefore, were unable to categorize them as benign, suspicious, or malignant. An advantage of grayscale inverted images for detection of microcalcification is that black spots indicative of microcalcification are easily seen and taken attention on white backgrounds and graphs. By altering the perception of the reader, even a single white spot not apparent on a standard image becomes readily recognizable as a black spot on a negative image.

Additionally, intramammary lymph nodes were also easier to identify on grayscale inverted images than on standard images, which increases diagnostic and detection rates. We could readily detect the characteristic small fatty hilum of intramammary lymph nodes on negative images. However, grayscale inverted images provided no advantage in terms of detecting axillary lymph nodes, even though they also have a fatty hilum. In our opinion, this may be explained by the fact that the axillary lymph node hilum is large and therefore apparent on both negative and standard images. Evaluation of grayscale inverted images in addition to standard images aids diagnosis owing to the ease of detecting the small white fatty hila of intramammary lymph nodes, which appear as bean-shaped, darkly colored masses.

The limitations of our study are the small number of cases and the performance of the assessments by a single radiologist.

In our study, assessment of grayscale inverted images, compared with standard images, improved detection of microcalcifications and intramammary lymph nodes. We believe that this finding is important because microcalcification is a strong indicator of benign or malignant lesions. Hence, we recommend the use of grayscale inverted images for evaluating microcalcification.

Conclusions

Assessment of grayscale inverted images in addition to standard images improves mammographic evaluation and facilitates the detection of small but crucial signs of breast cancer such as microcalcification. Therefore, assessment of grayscale inverted images can be practicable especially in busy centers.

Abbreviations

ACR:

American College of Radiology

BIRADS:

Breast Imaging Reporting and Data System

CAD:

Computer-aided detection

Gis:

Grayscale inverted

Std + Gsi:

Standard plus grayscale inverted image

Std:

Standard

References

  1. Miller AB, et al. Canadian National Breast Screening Study: I and II. Can Med Assoc J. 1992;147:1459–88.

    CAS  Google Scholar 

  2. Frisell J, et al. Randomized study of mammography screening-preliminary report on mortality in the Stockholm trial. Breast Cancer Res Treat. 1991;18:49–56.

    Article  CAS  PubMed  Google Scholar 

  3. Fletcher SW, et al. Mammographic screening for breast cancer. N Engl J Med. 2003;348:1672–80.

    Article  PubMed  PubMed Central  Google Scholar 

  4. Birdwell RL, et al. Computer-aided detection with screening mammography in a university hospital setting. Radiology. 2005;236:451–7.

    Article  PubMed  Google Scholar 

  5. Thurfjell EL, et al. Benefit of independent double reading in a population-based mammography screening program. Radiology. 1994;191:241–4.

    Article  CAS  PubMed  Google Scholar 

  6. Gilbert FJ, et al. Single reading with computer-aided detection for screening mammography. N Engl J Med. 2008;359(16):1675–84.

    Article  CAS  PubMed  Google Scholar 

  7. D’Orsi CJ, et al. ACR BI-RADS® Atlas, Breast Imaging Reporting and Data System. 5th ed. Reston: American College of Radiology; 2013.

    Google Scholar 

  8. Turkish Radiology Assosciation. Guide for Breast Cancer Scanning in Turkey. TRD Sufficiency Board; 2011.

  9. Hofvind S, et al. Screening-detected breast cancers: discordant independent double reading in a population-based screening program. Radiology. 2009;253:652–60.

    Article  PubMed  Google Scholar 

  10. Blanks RG, et al. Observer variability in cancer detection during routine repeat (incident) mammographic screening in a study of two versus one view mammography. J Med Screen. 1999;6:152–8.

    Article  CAS  PubMed  Google Scholar 

  11. Anderson ED, et al. The efficacy of double reading mammograms in breast screening. Clin Radiol. 1994;49(4):248–51.

    Article  CAS  PubMed  Google Scholar 

  12. Azavedo E, et al. Is single reading with computer-aided detection (CAD) as good as double reading in mammography screening? A systematic review. BMC Med Imaging. 2012;12:22. doi:10.1186/1471-2342-12-22.

    Article  PubMed  PubMed Central  Google Scholar 

  13. Guerriero C, et al. Is computer aided detection (CAD) cost effective in screening mammography? A model based on the CADET II study. BMC Health Serv Res. 2011;11(1):11.

    Article  PubMed  PubMed Central  Google Scholar 

  14. Jung NY, et al. Who could benefit the most from using a computer-aided detection system in full-field digital mammography? World J Surg Oncol. 2014;12:168. doi:10.1186/1477-7819-12-168.

    Article  PubMed  PubMed Central  Google Scholar 

  15. Bolivar AV, et al. Computer-aided detection system applied to full-field digital mammograms. Acta Radiol. 2010;51:1086–92.

    Article  PubMed  Google Scholar 

  16. Dromain C, et al. Computed-aided diagnosis (CAD) in the detection of breast cancer. Eur J Radiol. 2013;82(3):417–23. doi:10.1016/j.ejrad.2012.03.005.

    Article  CAS  PubMed  Google Scholar 

  17. Pisano ED, et al. American College of Radiology Imaging Network digital mammographic imaging screening trial: objectives and methodology. Radiology. 2005;236:404–12.

    Article  PubMed  Google Scholar 

  18. Park JB, et al. Diagnostic accuracy of the inverted grayscale rib series for detection of rib fracture in minor chest trauma. Am J Emerg Med. 2015;33(4):548–52.

    Article  PubMed  Google Scholar 

  19. Thompson JD, et al. The impact of greyscale inversion for nodule detection in an anthropomorphic chest phantom: a free-response observer study. Br J Radiol. 2016;89:(1064):20160249.

  20. Reese DJ, et al. Intra- and interobserver variability of board-certified veterinary radiologists and veterinary general practitioners for pulmonary nodule detection in standard and inverted display mode images of digital thoracic radiographs of dogs. J Am Vet Med Assoc. 2016;238(8):998–1003.

    Article  Google Scholar 

Download references

Acknowledgements

None

Funding

This study was not funded by any organization.

Availability of data and materials

The datasets used and/or analysed during the current study available from the corresponding author on reasonable request.

Authors’ contributions

AA planned the study, and participated in its design and coordination and collected of data and drafted the manuscript. KK performed the statistical analysis and tables and interpreted the data. All the authors read and approved the final version of the manuscript.

Competing interests

The authors declare that they have no competing interests.

Consent for publication

Not applicable

Ethics approval and consent to participate

The study was approved by the Ethics Committee of the Faculty of Medicine, Necmettin Erbakan University. All procedures performed in studies involving human participants were in accordance with the ethical standards of the institutional committee and with the 1964 Helsinki declaration and its later amendments or comparable ethical standards. The Ethics committee waived the need for consent because it is a retrospective study.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Ayşegül Altunkeser.

Rights and permissions

Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Altunkeser, A., Körez, M.K. Usefulness of grayscale inverted images in addition to standard images in digital mammography. BMC Med Imaging 17, 26 (2017). https://doi.org/10.1186/s12880-017-0196-6

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s12880-017-0196-6

Keywords