Evaluation of the diagnostic utility on 1.5T and 3.0T 1H magnetic resonance spectroscopy for temporal lobe epilepsy

Abstract

Background

1H magnetic resonance spectroscopy (1H-MRS) can be used to study neurological disorders because it can be utilized to examine the concentrations of related metabolites. However, the diagnostic utility of different field strengths for temporal lobe epilepsy (TLE) remains unclear. The purpose of this study is to make quantitative comparisons of metabolites of TLE at 1.5T and 3.0T and evaluate their efficacy.

Methods

Our retrospective collections included the single-voxel 1H-MRS of 23 TLE patients and 17 healthy control volunteers (HCs) with a 1.5T scanner, as well as 29 TLE patients and 17 HCs with a 3.0T scanner. Particularly, HCs were involved both the scans with 1.5T and 3.0T scanners, respectively. The metabolites, including the N-acetylaspartate (NAA), creatine (Cr), and choline (Cho), were measured in the left or right temporal pole of brain. To analyze the ratio of brain metabolites, including NAA/Cr, NAA/Cho, NAA/(Cho + Cr) and Cho/Cr, four controlled experiments were designed to evaluate the diagnostic utility of TLE on 1.5T and 3.0T MRS, included: (1) 1.5T TLE group vs. 1.5T HCs by the Mann-Whitney U Test, (2) 3.0T TLE group vs. 3.0T HCs by the Mann-Whitney U Test, (3) the power analysis for the 1.5T and 3.0T scanner, and (4) 3.0T HCs vs. 1.5T HCs by Paired T-Test.

Results

Three metabolite ratios (NAA/Cr, NAA/Cho, and NAA/(Cho + Cr) showed the same statistical difference (p < 0.05) in distinguishing the TLE from HCs in the bilateral temporal poles when using 1.5T or 3.0T scanners. Similarly, the power analysis demonstrated that four metabolite ratios (NAA/Cr, NAA/Cho, NAA/(Cho + Cr), Cho/Cr) had similar distinction abilities between 1.5T and 3.0T scanner, denoting both 1.5T and 3.0T scanners were provided with similar sensitivities and reproducibilities for metabolites detection. Moreover, the metabolite ratios of the same healthy volunteers were not statistically different between 1.5T and 3.0T scanners, except for NAA/Cho (p < 0.05).

Conclusions

1.5T and 3.0T scanners may have comparable diagnostic potential when 1H-MRS was used to diagnose patients with TLE.

Peer Review reports

Background

Epilepsy is a kind of serious neurological disorder of the brain [1]. The seizures of epilepsy can be divided into the types of focal, generalized and unknown [2]. As the most common form of focal seizures [3], temporal lobe epilepsy (TLE) approximately accounts for 60% of adult epilepsy cases [2, 4]. During this disease attack, a loss of consciousness, disturbances of limb movements, laloplegia like temporary alogia and other symptoms will occur. Nevertheless, the diagnosis of TLE is still challenging since some resources, including the gold standard, complete clinical history and reliable patient testimony are not accessible [1, 5]. Video electroencephalogram (V-EEG) and magnetic resonance imaging (MRI) are typical of current diagnosis techniques of TLE [5]. V-EEG can judge the possible type of TLE (focal seizures or generalized seizures) and assess the risk of recurrence by detecting abnormal patterns, but it is time-consuming in monitoring as the complete procedure may take more than 3 days on average [6]. MRI can identify the location of epilepsy lesions but roughly 30% of TLE patients have normal brain MRI results [7].

Studies have shown that TLE is associated with extensive neuronal dysfunction, which may be caused by brain damage or genetic mutations [8]. 1H magnetic resonance spectroscopy (1H-MRS) is an important non-invasive diagnostic tool for TLE. Brain metabolite concentrations can be measured by 1H-MRS, mainly including N-acetylaspartate (NAA), creatine (Cr), and choline (Cho) [9, 10]. NAA is synthesized in mitochondria [11] and serves as a crucial marker for neuronal impairment. Anomalies in neuronal structures, such as diminished neuronal viability, result in decreased level of NAA [12]. Cr is thought to be involved in the process of neuronal damage and alter neural metabolite levels in recurrent patient with TLE [13]. Stored in the cell membrane, Cho is necessary for all cells to function normally, affecting nerve signaling, cell signaling and lipid transport/metabolism [14]. Indications of abnormalities in the temporal lobe of TLE patients can be provided by these metabolite spectra, even if no abnormality is found on patients’ MRI images [15, 16]. Comparing with the contralateral hippocampus, NAA was less on the side of the affected temporal lobe in TLE patients’ hippocampi [17], and the concentration of NAA in the epileptogenic foci was lower than that in the non-epileptogenic regions [18] as well. Significantly high Cho/Cr ratios were observed in the right thalamus in the focal impaired awareness seizures [9, 19]. The NAA/Cr ratio proved to be a useful biomarker to discriminate TLE seizures from organic non-epileptic seizures [10]. A decline in the NAA/Cho ratio is another significant indicator for identifying the region of epilepsy [18].

These observations are based on the MRS data acquired from a sole magnetic resonance specific field strength, e.g. 1.5T or 3.0T. For instance, the research of epilepsy was conducted on 1.5T scanners [20, 21] while others were performed on 3.0T scanners [19, 22]. High magnetic field strength scanners can achieve a higher signal-to-noise ratio (SNR) and improve spectral dispersion. However, it is unclear whether 1H-MRS for TLE diagnosis can be affected by MRI scanners at different field strengths.

The purpose of this research is to study the differences in the diagnostic efficacy of 1H-MRS for TLE between 1.5T and 3.0T.

Methods

Participants and experimental design

This study was approved by the ethics committee of the Second Affiliated Hospital of Xiamen Medical College. Figure 1 shows the data collection process, we retrospectively collected single-voxel 1H-MRS of bilateral temporal poles from 2017 to 2021. There were 23 TLE patients diagnosed with the 1.5T scanner (15 males and 8 females, age: 29.52 ± 13 [Mean ± Standard Deviation (SD)]), 29 TLE patients diagnosed with the 3.0T scanner (20 males and 9 females, age: 28.2 ± 9.0), and 17 healthy controls (HCs) (11 males and 6 females, age: 23.35 ± 4.11) were scanned with both 1.5T and 3.0T MR scanners. To minimize the impact of multi-factor experimental conditions on outcomes, data were collected using scanners with different field strengths on the same healthy volunteer. To reduce the effect of time, HCs was scanned in batches in July 2021, with each batch scanned at both 1.5T and 3.0T in the same evening.

Fig. 1
figure 1

1H-MRS data collection flowchart of the study. The left and right temporal pole were used as regions of interest (ROI) and represented by red boxes to obtain metabolite concentrations of Cho, Cr, and NAA of subjects. The outer volume suppression (OVS), which are not parallel to the voxel edges, are manually added to enhance the saturation effect

The diagnosis procedure of TLE seizures was as follows: First, the neurologist identifies whether the patient was having a seizure according to clinical symptoms, and then employed auxiliary monitoring, such as blood tests, neurological examinations, and electrocardiograms, to rule out causes of non-brain abnormalities; Second, TLE seizures were confirmed by recording epileptiform discharges through an electroencephalogram; Finally, by analyzing imaging (MRI and 1H-MRS) and genes, neuroradiologists ascertained the cause of temporal lobe epileptic seizures (TLES). The whole process was completed by a neuroradiologist and an experienced neurologist, and the diagnosis result was finally given by an authoritative neurologist.

The 1H-MRS data were acquired from the bilateral temporal poles of all subjects, and four control experiments were designed (Fig. 2) as followed: (1) 1.5T TLE group vs. 1.5T HCs; (2) 3.0T TLE group vs. 3.0T HCs; (3) the power analysis between the 3.0T scanner and 1.5T scanner based on the statistical test of the TLE and HCs; and (4) 3.0T HCs vs. 1.5T HCs. These comparisons were aimed at evaluating the differences in the diagnostic utility of TLE at 1.5T and 3.0T, both of which were the most widespread magnetic resonance fields.

Fig. 2
figure 2

Schematic diagram of four control experiments. The number on the left side of the box indicates the sample size in an experiment

MRI and 1H-MRS acquisition

MRI acquisition parameters

1.5T MR scanner was equipped with GE SIGNA HD medical system, and the specific imaging parameters were as follows: (1) T2 FLAIR sequence: TR = 8600ms, TE = 120ms, TI = 2100ms, FOV = 240 × 240mm2, matrix = 288 × 160, slice thickness = 5 mm, NEX = 1; (2) FRFSE T2WI sequence: TR = 4760ms, TE = 102ms, FOV = 240 × 240mm2, matrix = 320 × 256, slice thickness = 5 mm, number of slices = 19, NEX = 1; (3) DWI: b value = 1000s/mm2, TR = 6000ms, TE = 56ms, FOV = 240 × 240mm2, matrix = 128 × 128, slice thickness = 5 mm, number of slices = 19, NEX = 2.

3.0T MR scanner was equipped with GE Discovery Silent MR (750 W) medical system, and the specific imaging parameters were as follows: (1) T2 FLAIR sequence: TR = 8600ms, TE = 140ms, TI = 2100ms, FOV = 240 × 240mm2, matrix = 288 × 224, slice thickness = 5 mm, number of slices = 22, NEX = 1; (2) T2WI: TR = 4425ms, TE = 90ms, FOV = 240 × 240mm2, matrix = 384 × 384, slice thickness = 5 mm, number of slices = 22; NEX = 1, (3) DWI: b value = 1000s/mm2, TR = 4250ms, TE = 56ms, FOV = 240 × 240mm2, matrix = 128 × 128, slice thickness = 5 mm, number of slices = 22, NEX = 2.

1H-MRS acquisition parameters

As shown in Table 1, PRESS sequence was applied to 1.5T scanner with the following sequence parameters: voxel size = 20 × 20 × 20mm3, chemical shift imaging layer thickness = 20 mm, NEX = 128, TR = 2000ms, TE = 144ms. PRESS sequence was also applied to 3.0T scanner with the following sequence parameters: voxel size = 20 × 20 × 20mm3, chemical shift imaging layer thickness = 20 mm, NEX = 128, TR = 1500ms, TE = 144ms. High-resolution T2W was used to localize single voxels in regions of interest (ROI) in the temporal lobe that coincides with the seizure onset zone determined by electroencephalography (EEG).

Table 1 The settings of the PRESS sequence parameters at 1.5T and 3.0T

Data pre-processing

After the MRS data of bilateral temporal lobes are collected, technicians use the built-in MRS software package (SAGE7.1) of GE scanner to quantify the data with the default mode to obtain the metabolite concentrations, and then calculate the metabolite concentration ratios, including NAA/Cr, NAA/Cho, NAA/(Cho + Cr) and Cho/Cr.

Statistical analysis

The variables were the ratios (NAA/Cr, NAA/Cho, NAA/ (Cho + Cr), Cho/Cr) of the 1H-MRS metabolites of the bilateral temporal poles at 1.5T and 3.0T. The data were grouped according to different magnetic field strengths and subjects’ characteristics. The Mann-Whitney U Test model was used for the 1.5T TLE group vs. 1.5T HCs, and 3.0T TLE group vs. 3.0T HCs. Each healthy volunteer was scanned using both 1.5T and 3.0T scanners, providing a set of paired and correlated data. Therefore, the Pair T-Test was used to compare 3.0T HCs with 1.5T HCs. The difference is considered statistically significant if p < 0.05. The power analysis was used to compare diagnostic capability differences between 1.5T and 3.0T scanners, and the probability of error, whose value depends on the significance criterion (α), the sample size (N), and the population effect size (ES) [23] was assessed by the G*Power (version 3.1.9.7). As shown below [24], the power value can be calculated:

$$\begin{array}{c}power=\text{Pr}\left(\text{r}\text{e}\text{j}\text{e}\text{c}\text{t} {H}_{0}|{H}_{1}\text{i}\text{s} \text{t}\text{r}\text{u}\text{e}\right)=1-\beta ,\end{array}$$
(1)

where H0 means the null hypothesis, H1 means the alternative hypothesis, and β means the probability of the error. The smaller power value suggests that β value is larger, and thus, the probability of the misjudgment increases, i.e. it is misjudged, even though the metabolite ratio is statistically different between 1.5T and 3.0T.

Results

Four controlled experiments were designed as followed: (1) 1.5T TLE group vs. 1.5T HCs; (2) 3.0T TLE group vs. 3.0T HCs; (3) the power analysis between the 3.0T scanner and 1.5T scanner based on the statistical test of the TLE and HCs; and (4) 3.0T HCs vs. 1.5T HCs. The comparisons of the spectra obtained from the control experiments were shown in Fig. 3.

Fig. 3
figure 3

Comparison of temporal lobe MRS with different data. (A) Left and (B) right temporal poles of [TLE group vs. HCs] with 1.5T scanner. (C) Left and (D) right temporal poles of [TLE group vs. HCs] with 3.0T scanner. (E) Left and (F) right temporal poles of HCs with [1.5T vs. 3.0T] scanners. Note: L: left temporal pole, R: right temporal pole. MRS: magnetic resonance spectroscopy

1.5T TLE group vs. 1.5T HCs

Compared with the HCs (Table 2), the ratios of the three metabolites (NAA/Cr, NAA/Cho, NAA/ (Cho + Cr)) are lower in the TLE group in the bilateral temporal poles with the 1.5T scanner. As shown in Fig. 4(A) and Table 2, for the left temporal pole, NAA/Cr, NAA/Cho and NAA/ (Cho + Cr)) ratios of the TLE group, compared with those of HCs, were statistically significant (p < 0.05), while Cho/Cr was not statistically different. The results are the same for the right temporal pole (Fig. 4(B) and Table 2). Therefore, the NAA/Cr, NAA/Cho, and NAA/(Cho + Cr) ratios of the TLE group were statistically different compared with HCs in the bilateral temporal poles.

Table 2 Comparison between 1.5T TLE groups and 1.5T HCs

3.0T TLE group vs. 3.0T HCs

The results acquired from the 3.0T scanner demonstrated that the TLE group’s NAA/Cr, NAA/Cho, NAA/(Cho + Cr), and Cho/Cr ratios were lower than those of the HCs (Table 3). In the bilateral temporal poles, NAA/Cr, NAA/Cho, and NAA/(Cho + Cr) ratios of the TLE group are statistically different from those of the HCs (Fig. 4(C)(D) and Table 3), which is consistent with the findings of the 1.5T scanner above.

Table 3 Comparison between 3.0T TLE and 3.0T HCs
Fig. 4
figure 4

Box plots of the ratios of four metabolites in control experiments. (A) and (B) show the comparisons of the metabolite concentration ratios between the TLE group and HCs with a 1.5T scanner. (C) and (D) show the comparisons of the metabolite concentration ratios between the TLE group and HCs with a 3.0T scanner. (E) and (F) are comparisons of the ratios of metabolite concentrations of HCs at different field strengths. L: left temporal pole. R: right temporal pole

Power analysis between TLE groups and HCs

The power analysis was used to calculate the probability of the error of the Mann-Whitney U Test in Tables 2 and 3, and assess the utility of each metabolite ratio discriminating between TLEs and HCs on the 1.5T and 3.0T scanners. As shown in Fig. 5; Table 4, comparing 1.5T and 3.0T scanners, the power values of NAA/Cr, NAA/Cho, NAA/(Cho + Cr) and Cho/Cr were similar in the bilateral temporal poles, indicating that the four metabolite ratios had similar … abilities for TLEs and HCs on the 1.5T and 3.0 T scanners. The power values of NAA/Cr, NAA/Cho, NAA/(Cho + Cr) were greater than 0.85, namely the probability of the error was less than 15%, which means the high reliability of the results.

Table 4 Power analysis between TLE groups and HCs
Fig. 5
figure 5

The power values and beta error of the NAA/Cr ratios. (A) shows the comparison between the TLE and HCs in the right temporal pole with the 1.5T scanner. (B) shows the comparison between the TLE and HCs in the right temporal pole with the 3.0T scanner

3.0T HCs vs. 1.5T HCs

Each healthy volunteer was scanned with both 1.5T and 3.0T scanners. Compared with the left temporal pole of the 1.5T HCs (Fig. 4(E) (F) and Table 5), four metabolite ratios (NAA/Cr, NAA/Cho, NAA/(Cho + Cr), Cho/Cr) of the 3.0T HCs decreased by 7.63%, 2.08%,4.22%, and 6.00%, respectively. Compared with the right temporal pole of the 1.5T HCs, NAA/Cr, NAA/Cho, and NAA/(Cho + Cr) ratios of the 3.0T HCs decreased by 4.10%, 10.35%, and 7.27%, respectively, but Cho/Cr ratio was slightly elevated (5.42%). There was no significant difference in the four metabolites ratios compared 3.0T HCs with 1.5T HCs, except for NAA/Cho in the right temporal pole (p = 0.035).

Table 5 Comparison between 3.0T HCs and 1.5T HCs

Discussion

Recently, MRS has been extensively studied for epilepsy, including neuronal damage in mesial TLE [25], searching for biomarkers to detect epileptogenic zone in non-lesional focal impaired awareness epilepsy [26], change of GABA level in hippocampus and anterior cingulate cortex [13], and discuss the differences in brain metabolic between TLE and organic non-epileptics [10]. To our knowledge, no other work has involved the evaluation of diagnostic utility for TLE on different field strengths.

In this work, these three metabolites NAA, Cr, and Cho were calculated as concentration ratios to evaluate the diagnostic utility of 1.5T and 3.0T 1H MRS for TLE. According to our findings, the TLE group’s NAA/Cr, NAA/Cho, and NAA/(Cho + Cr) ratios were lower than those of HCs. This observation was consistent with the findings of earlier studies [27]. One potential explanation is the increased cell density resulting from neuron loss or metabolic damage [10, 17, 28]. Other studies have shown that NAA concentrations are reduced in TLE. Low NAA may reflect degree of discharge from temporal lobe seizures. More important, the same statistical characteristics were discovered in the bilateral temporal poles whether 1.5T or 3.0T scanner was used.

The Mann-Whitney U Test was used to compare the two experiment results shown in Tables 2 and 3, which demonstrated that three metabolic ratios (NAA/Cr, NAA/Cho, and NAA/(Cho + Cr)) had the same statistically significant differences at both 1.5T and 3.0T. Thus, these metabolite ratios might be used as potential biomarkers to identify TLEs from HCs. Therefore, these metabolite ratios can show significant differences no matter with the 1.5T or 3.0T scanner, which may indicate that 1.5T and 3.0T have similar diagnostic efficacy for TLE.

The power analysis was used to evaluate metabolites’ ability to distinguish the TLE from the HCs as well as both 1.5T and 3.0T scanners’ diagnostic performances, i.e. sensitivity and reproducibility. Experimental results show that, the power values of NAA/Cr, NAA/Cho, NAA/(Cho + Cr) and Cho/Cr are close whether on the 1.5T or the 3.0T scanner (Fig. 5; Table 4), and the four metabolite ratios have similar ability to distinguish TLE from HC, which denotes 1.5T and 3.0T scanners have similar sensitivities and reproducibility on metabolite detection with the validation of the results of the Mann-Whitney U Test.

To reduce the influence of different subjects on the experimental results, Paired T-test (Table 5) was used to analyse the metabolite ratios of each same volunteer at 1.5T and 3.0T scanners. The results show no significant difference in metabolite ratios between the 1.5T and 3.0T scanners, whether in the left or right temporal lobe, except for NAA/Cho in the right temporal pole (p < 0.05). This suggests that 1.5T and 3.0T scanners may have similar diagnostic efficacy for HCs. There was just one metabolite ratio, NAA/Cho, that differed statistically (p < 0.05) between HCs when different field strengths are compared. There might be several reasons for the abnormal value, such as insufficient healthy volunteers, differences in manual operations during scanning.

MRS studies mainly focused on the metabolites of NAA, Cr, and Cho [10] and gradually shifted to glutamate (Glu) and gamma-aminobutyric acid (GABA) recently [29, 30]. Studies showed that the mechanism for epileptogenesis involves an imbalance between excitatory and inhibitory processes [25, 30]. This imbalance is probably associated with a higher level of extracellular Glu in the brain and a decrease in GABA concentration, leading to excitotoxicity, seizures, and cell damage [29, 30].

This work has some limitations. First, only four metabolic ratios were included in the analysis, which may be not enough. Other metabolites, such as Glu and GABA, are also important potential biomarkers of epilepsy [29]. Second, the study of NAA/Cho ought to be paid more attention if more samples can be included. Finally, studies have shown that 7 T MRS can achieve higher spectral resolution and signal-to-noise ratio, which helps to distinguish epilepsy metabolites [29]. However, this paper does not offer a profound study about the influence of the signal-to-noise ratio and resolution at different field strengths [31, 32], instead focusing on the comparison of the abilities of 1.5T and 3.0T scanners to diagnose TLE using 1H-MRS metabolite concentrations.

Conclusions

To evaluation of the diagnostic utility on 1.5T and 3.0T 1H magnetic resonance spectroscopy for TLE, 4 controlled experiments were designed. The results showed that potential biomarkers (NAA/Cr, NAA/Cho, NAA/(Cho + Cr)) had the same efficacy using the 1.5T and 3.0T scanners for distinguishing the TLE from HCs in the bilateral temporal lobes. Thus, both 1.5T and 3.0T scanners may have comparable potential in distinguishing TLEs from HCs when 1H-MRS is used to identify patients with TLE.

Data Availability

The datasets generated and/or analysed during the current study are not publicly available due to [REASON WHY DATA ARE NOT PUBLIC] but are available from the corresponding author on reasonable request.

Abbreviations

1H-MRS:

1H magnetic resonance spectroscopy

TLE:

Temporal lobe epilepsy

HCs:

Healthy control volunteers

NAA:

N-acetylaspartate

Cr:

Creatine

Cho:

Choline

V-EEG:

Video electroencephalogram

MRI:

Magnetic resonance imaging

SD:

Standard Deviation

ES:

Effect size

References

  1. Brodie MJ, French JA. Management of Epilepsy in adolescents and adults. Lancet. 2000;356(9226):323–9.

    Article  CAS  PubMed  Google Scholar 

  2. Engel Jr. J. A proposed diagnostic scheme for people with epileptic seizures and with rpilepsy: report of the ILAE task force on classification and terminology. Epilepsia. 2001;42(6):796–803.

    Article  PubMed  Google Scholar 

  3. Wiebe S. Epidemiology of temporal lobe Epilepsy. Can J Neurol Sci. 2000;27(S1):6–S10.

    Article  Google Scholar 

  4. Téllez-Zenteno JF, Hernández-Ronquillo L. A review of the epidemiology of temporal lobe epilepsy. Epilepsy Res Treat 2012, 2012.

  5. Thijs RD, Surges R, O’Brien TJ, Sander JW. Epilepsy in adults. Lancet. 2019;393(10172):689–701.

    Article  PubMed  Google Scholar 

  6. Liu J, Meng F, Liu Z. Seizure-related adverse events during video-electroencephalography monitoring. Epileptic Disord. 2012;14(1):51–6.

    Article  CAS  PubMed  Google Scholar 

  7. Muhlhofer W, Tan YL, Mueller SG, Knowlton R. MRI-negative temporal lobe epilepsy-what do we know? Epilepsia. 2017;58(5):727–42.

    Article  PubMed  Google Scholar 

  8. Pitkänen A, Löscher W, Vezzani A, Becker AJ, Simonato M, Lukasiuk K, Gröhn O, Bankstahl JP, Friedman A, Aronica E, et al. Advances in the development of biomarkers for Epilepsy. Lancet Neurol. 2016;15(8):843–56.

    Article  PubMed  Google Scholar 

  9. Hammen T, Kuzniecky R. Magnetic resonance spectroscopy in Epilepsy. Handb Clin Neurol. 2012;107:399–408.

    Article  PubMed  Google Scholar 

  10. Liu D, Yang Y, Chen D, Wang Z, Guo D, Bao L, Dong J, Wang X, Qu X. Brain metabolic differences between temporal lobe epileptic seizures and organic non-epileptic seizures in postictal phase: a retrospective study with magnetic resonance spectroscopy. Quant Imaging Med Surg. 2021;11(8):3781–91.

    Article  PubMed  PubMed Central  Google Scholar 

  11. Sullivan EV, Adalsteinsson E, Spielman DM, Hurd RE, Pfefferbaum A. N-acetylaspartate–a marker of neuronal integrity. Ann Neurol. 2001;50(6):824–5.

    Article  Google Scholar 

  12. Basha MAA, Refaat R, Ahmed AF, Yousef HY, Alsowey AM, Metwally MI, Aly SA, Hussien HM, El-Saadany HF, AlGhobashy AA. Brain magnetic resonance spectroscopy (MRS) as a diagnostic tool for detecting early neurological changes in children with Wilson’s Disease. Eur J Radiol. 2019;111:41–6.

    Article  Google Scholar 

  13. He C, Liu P, Wu Y, Chen H, Song Y, Yin J. Gamma-aminobutyric acid (GABA) changes in the hippocampus and anterior cingulate cortex in patients with temporal lobe Epilepsy. Epilepsy Behav. 2021;115:107683.

    Article  PubMed  Google Scholar 

  14. Zeisel SH. Choline: needed for normal development of memory. J Am Coll Nutr. 2000;19(5):528S–31.

    Article  CAS  PubMed  Google Scholar 

  15. Todua F, Beraia M, Okujava M. Magnetic resonance spectroscopy (MRS) in patients with the temporal lobe epilepsy. European Congress of Radiology (ECR) 2017; 2017.

  16. Malmgren K, Thom M. Hippocampal sclerosis origins and imaging. Epilepsia. 2012;53(4):19–33.

    Article  CAS  PubMed  Google Scholar 

  17. Hugg JW, Laxer KD, Matson GB, Maudsley AA, Weiner MW. Neuron loss localizes human temporal lobe Epilepsy by in vivo proton magnetic resonance spectroscopic imaging. Ann Neurol. 1993;34(6):788–94.

    Article  CAS  PubMed  Google Scholar 

  18. Stanley JA, Cendes F, Dubeau F, Andermann F, Arnold DL. Proton magnetic resonance spectroscopic imaging in patients with extratemporal Epilepsy. Epilepsia. 1998;39(3):267–73.

    Article  CAS  PubMed  Google Scholar 

  19. Tan Z, Long X, Tian F, Huang L, Xie F, Li S. Alterations in brain metabolites in patients with Epilepsy with impaired consciousness: a case-control study of interictal multivoxel 1H-MRS findings. Am J Neuroradiol. 2019;40(2):245–52.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  20. Li S, Luo X, Zhang S, Tang Y, Sun J, Meng Q, Yu H, Sun C. Evaluation of multilevel surgeries in children with spastic cerebral palsy based on surface electromyography. Front Neurosci 2021, 15(728).

  21. Stefan H, Pauli E, Schäfer I, Huk W, Tomandl B. 1H-MR spectroscopy: a promising method in distinguishing subgroups in temporal lobe Epilepsy? J Neurol Sci. 2003;215(1–2):21–5.

    PubMed  Google Scholar 

  22. Riederer F, Bittsanský M, Schmidt C, Mlynárik V, Baumgartner C, Moser E. Serles W. 1H magnetic resonance spectroscopy at 3 T in cryptogenic and mesial temporal lobe Epilepsy. NMR Biomed. 2006;19(5):544–53.

    Article  CAS  PubMed  Google Scholar 

  23. Cohen J. Statistical power analysis. Curr Dir Psychol. 1992;1(3):98–101.

    Article  Google Scholar 

  24. Cohen J. The concepts of power snalysis. Statistical power analysis for the behavioral sciences. Hillsdale: Elrbaum; 1988.

    Google Scholar 

  25. Granussio E, Silva G, Pimentel-Silva L, Cendes F. Progression of neuronal damage in mesial temporal lobe Epilepsy measured by proton magnetic resonance spectroscopy. J Neurol Sci 2021;429.

  26. Aboelsafa AA, Dawoud RM. Arterial spin-labelling and magnetic resonance spectroscopy as imaging biomarkers for detection of epileptogenic zone in non-lesional focal impaired awareness Epilepsy. Egypt J Radiol Nucl Med. 2020;51(1):1–17.

    Google Scholar 

  27. Fadaie F, Mobarakeh NM, Fesharaki SS, Harirchian MH, Kharazi HH, Rad HS, Habibabadi JM. 1H-MRS metabolite’s ratios show temporal alternation in temporal lobe seizure: comparison between interictal and postictal phases. Epilepsy Res. 2016;128:158–62.

    Article  CAS  PubMed  Google Scholar 

  28. Ende GR, Laxer KD, Knowlton RC, Matson GB, Schuff N, Fein G, Weiner MW. Temporal lobe Epilepsy: bilateral hippocampal metabolite changes revealed at proton MR spectroscopic imaging. Radiology. 1997;202(3):809–17.

    Article  CAS  PubMed  Google Scholar 

  29. Gonen OM, Moffat BA, Desmond PM, Lui E, Kwan P, O’Brien TJ. Seven-tesla quantitative magnetic resonance spectroscopy of glutamate, γ‐aminobutyric acid, and glutathione in the posterior cingulate cortex/precuneus in patients with Epilepsy. Epilepsia. 2020;61:2785–94.

    Article  CAS  PubMed  Google Scholar 

  30. Sarlo GL, Holton KF. Brain concentrations of glutamate and GABA in human Epilepsy: a review. Seizure. 2021;91:213–27.

    Article  PubMed  Google Scholar 

  31. Ladd ME, Bachert P, Meyerspeer M, Moser E, Nagel AM, Norris DG, Schmitter S, Speck O, Straub S, Zaiss M. Pros and cons of ultra-high-field MRI/MRS for human application. Prog Nucl Magn Reson Spectrosc. 2018;109:1–50.

    Article  CAS  PubMed  Google Scholar 

  32. Kim J-h, Chang K-H, Na DG, Song IC, Kim SJ, Kwon BJ, Han MH. Comparison of 1.5T and 3T 1H MR spectroscopy for human brain tumors. Korean J Radiol. 2006;7(3):156–61.

    Article  PubMed  PubMed Central  Google Scholar 

Download references

Acknowledgements

We sincerely thanks to Meijin Lin for improved the language, all participants in this project and all our collaborators in the second affiliated hospital of Xiamen Medical College.

Funding

This study was financially supported by National Natural Science Foundation of China (62122064 and 61971361), Science and Technology Planning Project of Fujian Province (2020H6003), Fujian Provincial Health Young and Middle-aged Key Talents Training Project (2020GGB067), General project of Natural Science Foundation of Fujian Province (2022J011384), Nanqiang Outstanding Talent program of Xiamen University, Fundamental Research Funds for the Central Universities (0621ZK1035), President Fund of Xiamen University. The funding bodies played no role in the design of the study and collection, analysis, interpretation of data, and in writing the manuscript.

Author information

Authors and Affiliations

Authors

Contributions

BQ, YGY, GY, DBL, XBQ: conception and design. GFZ and XBQ: administrative support. YWZ, YGY and GY: provision of study materials or patients. BQ, DBL, SJW, HJT and MX: data analysis. BQ, HJT and MX: writing – drafting of the manuscript. GFZ, XBQ, HJT, MX and RHC: writing –review and editing. All authors contributed to the article and approved the submitted version.

Corresponding authors

Correspondence to Yonggui Yang, Gen Yan or Xiaobo Qu.

Ethics declarations

Ethics approval and consent to participate

This study was conducted in accordance with the principles of the Declaration of Helsinki and approved by the Medical Ethics Committee of the Second Affiliated Hospital of Xiamen Medical College (Approval No: 2020044). Written informed consent was obtained from all participants before the examination. All methods were carried out in accordance with the relevant guidelines and regulations.

Consent for publication

Not applicable.

Competing interests

Xiaobo Qu is a Senior Board Member of BMC Medical Imaging. The other authors have no competing interests.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Qu, B., Tan, H., Xiao, M. et al. Evaluation of the diagnostic utility on 1.5T and 3.0T 1H magnetic resonance spectroscopy for temporal lobe epilepsy. BMC Med Imaging 23, 185 (2023). https://doi.org/10.1186/s12880-023-01136-w

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s12880-023-01136-w

Keywords