Skip to main content
  • Research article
  • Open access
  • Published:

Non-local means based Rician noise filtering for diffusion tensor and kurtosis imaging in human brain and spinal cord

Abstract

Background

To investigate the effect of using a Rician nonlocal means (NLM) filter on quantification of diffusion tensor (DT)- and diffusion kurtosis (DK)-derived metrics in various anatomical regions of the human brain and the spinal cord, when combined with a constrained linear least squares (CLLS) approach.

Methods

Prospective brain data from 9 healthy subjects and retrospective spinal cord data from 5 healthy subjects from a 3 T MRI scanner were included in the study. Prior to tensor estimation, registered diffusion weighted images were denoised by an optimized blockwise NLM filter with CLLS. Mean kurtosis (MK), radial kurtosis (RK), axial kurtosis (AK), mean diffusivity (MD), radial diffusivity (RD), axial diffusivity (AD) and fractional anisotropy (FA), were determined in anatomical structures of the brain and the spinal cord. DTI and DKI metrics, signal-to-noise ratio (SNR) and Chi-square values were quantified in distinct anatomical regions for all subjects, with and without Rician denoising.

Results

The averaged SNR significantly increased with Rician denoising by a factor of 2 while the averaged Chi-square values significantly decreased up to 61% in the brain and up to 43% in the spinal cord after Rician NLM filtering. In the brain, the mean MK varied from 0.70 (putamen) to 1.27 (internal capsule) while AK and RK varied from 0.58 (corpus callosum) to 0.92 (cingulum) and from 0.70 (putamen) to 1.98 (corpus callosum), respectively. In the spinal cord, FA varied from 0.78 in lateral column to 0.81 in dorsal column while MD varied from 0.91 × 10−3 mm2/s (lateral) to 0.93 × 10−3 mm2/s (dorsal). RD varied from 0.34 × 10−3 mm2/s (dorsal) to 0.38 × 10−3 mm2/s (lateral) and AD varied from 1.96 × 10−3 mm2/s (lateral) to 2.11 × 10−3 mm2/s (dorsal).

Conclusions

Our results show a Rician denoising NLM filter incorporated with CLLS significantly increases SNR and reduces estimation errors of DT- and KT-derived metrics, providing the reliable metrics estimation with adequate SNR levels.

Peer Review reports

Background

Diffusion magnetic resonance imaging (dMRI) is widely applied imaging modality allowing to characterize the microstructural properties of brain tissues. Among various kinds of dMRI methods, diffusion tensor imaging (DTI) has been the most commonly used [1,2,3,4,5,6,7,8,9,10,11]. While DTI focuses on the study of white matter (WM) structure providing important information about the tissue anisotropy, diffusion kurtosis imaging (DKI), an extension of DTI provides complementary information especially for analysis of microstructure in gray matter (GM) by probing non-Gaussian diffusion properties [12,13,14,15,16]. In recent years, the interest for DKI has been continuously growing and mean kurtosis (MK), a principal metric of the diffusional non-Gaussianity has demonstrated improved sensitivity in measuring developmental [16,17,18] and pathological [15, 19] changes in neural tissues of both animal and human over conventional DTI. However, one of the challenges in DKI measurement which reduces its practical usage in clinical research is a use of multiple and higher diffusion-weighting (b-values) as compared to a conventional DTI. This results in a low signal-to-noise ratio (SNR) of which the effect plays a large role in yielding the erroneous tensor estimation due to the bias induced by Rician nature of random noise [20, 21]. The low SNR can also affect the accuracy of diffusion tensor properties such as trace and fractional anisotropy (FA) [11, 22, 23]. Especially, the application of DTI in the spinal cord suffers from the low SNR [24, 25] in addition to other challenges such as the small size of the cord, physiological motion, local field inhomogeneity and susceptibility artefacts [26,27,28,29].

To date, various denoising methods have been developed to improve the quality of DW images, such as the Gaussian filter [30, 31], anisotropic diffusion filter [32,33,34], linear minimum mean squared error filter [21], random matrix theory [35], multi-shell position-orientation-adaptive smoothing [36] and nonlocal means (NLM) filter [37, 38]. In particular, NLM filter has been suggested to significantly improve MR data quality by reducing Rician noise [20, 39,40,41] and implemented to DTI [42,43,44] and DKI [45] in the human brain. Moreover, NLM filter has shown to provide efficiency of noise removal while the fine structures and details of images are well preserved [40, 46,47,48,49,50,51]. Despite considerable research effort in identifying the best denoising algorithm among the existing algorithms for DKI [45, 46, 52] as well as DTI [21, 37, 45, 48, 53], the standard denoising methodology for DKI post-processing part is still to be established, which makes DKI more difficult to use clinically. In order to translate DKI as a clinical tool, it is required to evaluate the direct impact of correcting such bias on regional DKI values associated with SNR in the healthy brain. A few studies investigated inter-subject variability of DKI metrics in the brain of healthy subjects without including noise bias corrections [54, 55]. To the best of our knowledge, only one study [56] investigated the influence of noise correction by data processing related to inter-subject variability of DKI metrics, however the reported diffusion metrics are limited to mean diffusivity (MD), FA and MK. In DTI of spinal cord, previous studies suggested various correction methods to boost SNR [25, 57,58,59,60,61], however the impact of Rician noise filtering in DTI of the human spinal cord has not been reported yet.

In this study, we evaluate the effect of using the Rician NLM filtering in combination with constrained linear least squares (CLLS), one of the least-squares (LS) fitting algorithms commonly used in DTI and DKI studies [13, 62, 63] of the human brain and spinal cord. In particular, we present here diffusivity and kurtosis measures in different anatomical regions of the healthy human volunteers. Additionally, error measures are quantitatively compared between each DT- and KT-metric with and without implementing a Rician NLM filter (CLLS-R and CLLS, respectively).

Methods

Subjects

For brain scans, 9 healthy volunteers (5 male, 4 females; mean age = 24; ± standard deviation = 2 years) were recruited through local advertisements. The exclusion criteria for healthy controls were subjects who suffer from any neurological/psychological conditions and any physical disabilities. All participants were studied after signed, informed consent. The study was approved by the local Institutional Review Board. Additionally, the previously obtained spinal cord data including five healthy volunteers (1 male, 4 female; mean age = 32 ± 7 years) [64] were retrospectively used in the present study for Rician denoising filtering.

Data acquisition

All MRI scans were performed on a Philips 3 T MRI Achieva scanner (Philips Healthcare, Best, The Netherlands). Brain scans were carried out with a body coil excitation and an 8-channel SENSE head coil for reception. For DKI data, four averaged minimally weighted (b0) and 32 non-collinear diffusion-encoding directions with two b-values (1000 and 2000 s/mm2) were acquired using single-shot EPI sequence as similar to previous studies [65, 66]. The imaging parameters were: TR/TE = 2000/69 ms, FOV = 224 × 224 mm2, acquisition matrix = 88 × 88, reconstructed resolution = 2 × 2 mm2, fold-over direction = AP, EPI factor = 47, 44 axial slices with 3 mm thickness and no inter-slice gap to cover the whole brain, SENSE factor = 2, 3/4 partial Fourier encoding, 2 averages and total scan time = 19 min. 38 s. The number of averages for DWIs was optimized to sustain the reasonable SNR while compromising scan time and six sequential observations of the DWIs with 1 average were obtained in the additional scans for computing the SNR. For anatomical reference, three-dimensional T1-weighted images were also acquired using 3D-MPRAGE sequence with the following parameters: TR/TE = 7.0/3.2 ms, TI = 800–818 ms, FOV = 224 × 224 × 167 mm3, acquisition matrix = 224 × 224, reconstruction resolution = 1 × 1 mm2, 167 sagittal slices with 1 mm thickness and no inter-slice gap to cover the whole brain, 1 average and total scan time = 10 min. 41 s. Acquisition parameters for spinal cord DTI data can be found from the previous study [64] which included four averaged minimally weighted (b0) and 15 diffusion-weighted volumes (b-value = 500 s/mm2) using single-shot echo-planar-imaging (EPI) sequence. Two imaging stacks in axial plane covered upper (C1 to C3) and lower (C4 to C6) portions of the cervical spinal cord (slice thickness = 2.5 and 5 mm, respectively).

Data postprocessing

Data were analysed using a custom-written program in Matlab (Mathworks, Natick, MA) and the DKE software (https://www.nitrc.org/projects/dke/) [52]. Detailed data postprocessing procedures for DKI are illustrated in Fig. 1. All DWIs were co-registered to the initial b0 image using a 6-degree-of-freedom, rigid body transformation procedure supplied in automated image registration (AIR) [67] in order to reduce the impact of motion artefact and eddy current distortion. Each gradient direction was revised with respect to transformation matrix after registration. Prior to tensor estimation, registered DWIs were denoised by an optimized blockwise NLM filter consisting of a blockwise implementation with automatic tuning of the smoothing parameter and block selection as described elsewhere [39, 40]. The numerical value for SNR in regions of interest (ROIs) for baseline image (b = 0) was computed by the signal statistics in a difference image as described in Farrell et al. [68]. DT and KT were estimated using CLLS algorithm suggested by Tabesh et al. [52] based on the definition of all DT- and KT-metrics [10, 14, 52].

Fig. 1
figure 1

Data processing pipeline

ROI-based measurements

For brain images, 6 ROIs including putamen (PUT; averaged area size = 985 mm2), globus pallidus (GP; averaged area size = 525 mm2), corpus callosum (CC; averaged area size = 2497 mm2), internal capsule (IC; averaged area size = 1643 mm2), external capsule (EC; averaged area size = 1241 mm2) and cingulum (Cg; averaged area size = 678 mm2) were manually drawn in a FA map and a color-coded map using ImageJ (National Institutes of Health, Bethesda, MD) by a trained rater (Z.Z.) referencing to the standard human brain atlas (Fig. 2) [69]. All 6 ROIs were then transferred to the rest of the DT- and KT-derived maps in order to compute the average and error measures within the ROIs for each subject. Chi-square value (χ2) was calculated by computing the difference between the observed and estimated signals to estimate the goodness of fitting. Later, another trained rater (W.Q.) independently placed ROIs on all datasets to ensure the indices were reliable across raters.

Fig. 2
figure 2

Location of ROIs on the FA and color-coded maps of the brain. Abbreviations PUT, putamen; GP, globus pallidus; CC, corpus callosum; IC, internal capsule; EC, external capsule; Cg, cingulum

For spinal cord images, 6 slices matching C1 to C6 levels of the spinal cord in a FA map per each subject, were selected around mid-level of two intervertebral discs to minimize susceptibility artefact typically caused near the disc-tissue boundaries. In each chosen slice, 3 ROIs were manually placed in the lateral left and right, and dorsal columns (Fig. 4a), a couple of pixels away from the boundaries of GM/WM and WM/cerebrospinal fluid (CSF) to avoid partial volume effect using ImageJ (National Institutes of Health, Bethesda, MD, USA). The area size of 1.563 mm2 for each ROI was consistent across all subjects. These ROIs were then transferred to the rest of the DTI-derived maps including mean FA, MD, radial diffusivity (RD) and axial diffusivity (AD), in order to compute the average of DTI-indices within the ROIs for each subject. χ2 was calculated in a voxel-by-voxel basis to evaluate the goodness of denoising performance. SNR and DTI-derived metrics were recalculated without the prior-denoising process and using the same ROIs that were previously defined for comparison.

Statistical analysis

To evaluate the inter-rater reliability, the intraclass correlation coefficient (ICC) was computed to assess reliability of mean values measured by two raters for each ROI as described by Shrout and Fleiss [70]. Averaged SNR values of b0 images and ROI-based diffusion and kurtosis metrics were compared between CLLS and CLLS-R by paired Mann–Whitney U test. All statistical analyses were performed using commercially available software (SPSS, Chicago, IL, USA). P-values of 0.05 were considered to be significant.

Results

In a single b0 image of the brain, SNR ranged from 10.85 (GP) to 20.50 (EC) with CLLS (without Rician NLM filtering) and from 25.25 (CC) to 74.38 (EC) with CLLS-R (with Rician NLM filtering) (Table 1). SNR significantly increased in all 6 ROIs using CLLS-R (244%, 177%, 62%, 162%, 263% and 125% for PUT, GP, CC, IC, EC and Cg, respectively with p < 0.05 for all ROIs) as compared to CLLS. In upper spinal cord levels C1–C3 (2.5 mm slice thickness), SNR ranged from 9.77 (Lateral) to 10.66 (Dorsal) with CLLS and from 16.32 (Lateral) to 21.97 (Dorsal) with CLLS-R (Table 2). In lower spinal cord levels C4–C6 (5 mm slice thickness), SNR ranged from 23.02 (Dorsal) to 23.73 (Lateral) with CLLS and from 45.21 (Dorsal) to 49.23 (Lateral) with CLLS-R (Table 2). In C1–C3, mean SNR significantly increased by 67.09% and 106.11% for lateral and dorsal respectively (p < 0.001 for both columns). Similarly, significant increase of mean SNR by 107.50% and 96.38% was observed for lateral and dorsal respectively in C4–C6 (p < 0.001 for both columns).

Table 1 Comparison of averaged SNR ± standard deviation of b0 images in the brain of healthy volunteers (n = 9) measured using CLLS and CLLS-R. Paired Mann–Whitney U test was used in the comparison between CLLS and CLLS-R
Table 2 Mean ± standard deviation of averaged SNR of b0 images in 5 healthy volunteers in lateral & dorsal columns before and after denoising

The χ2 maps in the brain of a representative subject (25, male) (Fig. 3a) and group differences (Fig. 3b) demonstrate that χ2 values estimated by CLLS-R are decreased in deep GM, WM and overall cortical GM as compared to those by CLLS. Group comparison averaged across 9 subjects at each ROI shows that mean χ2 values significantly decreased in all 6 regions of brain using CLLS-R as compared to CLLS (32—61% with p < 0.05). Similarly, the averaged χ2 values of the spinal cord over 5 healthy subjects significantly decreased by maximum 43% with Rician denoising as compared those without Rician denoising in all 3 columns throughout C1–C6 (p < 0.001) (Fig. 4B). In the representative DT- and KT-derived brain maps, KT-derived metrics (MK, axial kurtosis (AK) and radial kurtosis (RK)) show visual difference between CLLS and CLLS-R (Fig. 5). Kurtosis maps (AK and RK) using CLLS-R illustrate significant decrease in a number of erroneous pixels as compared to those using CLLS while moderate improvement is noticeable in MK maps (Fig. 5).

Fig. 3
figure 3

a Representative 4 neighbouring axial slices of χ2 maps in a healthy subject (25 years old, male) using CLLS and CLLS-R. b Group differences of the χ2 measurement at each brain region between before- (CLLS) and after-Rician denoising (CLLS-R). Significant differences between groups are indicated by *p < 0.05. Abbreviations PUT, putamen; GP, globus pallidus; CC, corpus callosum; IC, internal capsule; EC, external capsule; Cg, cingulum

Fig. 4
figure 4

a Location of anatomically defined ROIs on the FA map of the spinal cord. ROIs were selected in the ventro-lateral (red) and dorsal (green) column of the cervical spinal cord, to include most descending and ascending tracts, respectively. b The before-and-after plots for mean χ2 values at each ROI from 6 spinal cord levels (C1-C6) of all 5 healthy volunteers (30 pairs). All pairs show significant differences between before- (CLLS) and after-Rician denoising (CLLS-R) (p < 0.001). Abbreviations LR, lateral right; LL, lateral left; Dor, dorsal

Fig. 5
figure 5

Representative 3 axial slices of mean kurtosis (MK), axial kurtosis (AK) and radial kurtosis (RK) maps calculated using CLLS and CLLS-R. Scale bars represent dimensionless range of KT-derived metrics

All the KT-derived metrics (MK, RK, and AK), together with the DT-derived metrics (FA, MD, AD, and RD) by CLLS-R averaged across all subjects at each ROI are listed in Table 3. In the brain, MK varied from 0.70 to 1.27 with the lowest MK value obtained in PUT and highest in IC while RK varied from 0.70 (PUT) to 1.98 (CC). AK varied from 0.58 in the CC to 0.92 in Cg. FA varied from 0.12 to 0.65, with the lowest FA value obtained in the PUT and highest in CC while MD varied from 0.75 × 10−3 mm2/s (PUT) to 1.04 × 10−3 mm2/s (CC). RD varied from 0.50 × 10−3 mm2/s in IC to 0.71 × 10−3 mm2/s in PUT and AD varied from 0.85 × 10−3 mm2/s in PUT to 1.95 × 10−3 mm2/s in CC. ICC of DT- and KT-indices measured in various ROIs by two raters ranged from 0.744 to 0.999 (Table 4). In the spinal cord, FA varied from 0.78 in lateral column to 0.81 in dorsal column while MD varied from 0.91 × 10−3 mm2/s (lateral) to 0.93 × 10−3 mm2/s (dorsal). RD varied from 0.34 × 10−3 mm2/s (dorsal) to 0.38 × 10−3 mm2/s (lateral) and AD varied from 1.96 × 10−3 mm2/s (lateral) to 2.11 × 10−3 mm2/s (dorsal). DT- and KT-derived metrics estimated using CLLS-R were significantly different from those by CLLS in most of examined anatomical regions (Additional File 1: Supporting Information Table S1). The mean FA values using CLLS-R were significantly decreased in all 6 regions (3–11% with p < 0.05) compared to CLLS. Similar results were found for AD with exception of the PUT, CC and Cg. Meanwhile MD, RD, RK and MK values were significantly increased with exception of the PUT, IC and EC for MD and PUT for RD, CC and Cg for RK and PUT for MK. Additionally, the ROI-based mean χ2 values showed that CLLS-R significantly reduced χ2 values over all ROIs compared to CLLS (32—61% with p < 0.05).

Table 3 ROI-based mean ± standard deviation of DT- and KT-indices averaged across all subjects at each ROI in the brain and the spinal cord using CLLS-R
Table 4 Intraclass correlation coefficient of DT- and KT-indices measured in various ROIs by two ratersa

DISCUSSION

Despite great deal of effort in comparing various denoising algorithms for DKI in previous studies [45, 46, 52] and inter-subject variability of DKI metrics in brain of healthy subjects [54, 55], the direct impact of Rician noise correction on DKI data has not been sufficiently studied, especially related to improvement on SNR and error measures in various anatomical regions. In this study, we evaluate the influence of the Rician NLM filter in combination with CLLS algorithm on the brain DKI and the spinal cord DTI data. Rician NLM filter in combination with CLLS algorithm has the potential to be useful in clinical research because it is easy and efficient to implement. To the best of our knowledge, this is the first study to assess the influence of combining the NLM filter with CLLS in DTI of the spinal cord and DKI in the brain and to report ROI-based changes in SNR and the associated error estimates. To compare diffusional measures using different processing methods (CLLS and CLLS-R), we obtained whole brain DKI data from 9 healthy volunteers and compared mean DT- and KT-derived indices using CLLS and CLLS-R. Rician NLM filter was also applied to the previously acquired spinal cord data retrospectively [64].

Measurement of SNR

SNR assessment is important for reliable quantification of diffusional metrics [71]. When SNR is low, Rician noise does not only cause random fluctuations but also a signal dependent bias to the data, which may lead to difficulty in postprocessing such as tensor calculation. However, SNR levels are not routinely reported although previous reports suggest that low SNR causes a bias in FA which may vary with numerous other technical factors such as the region of brain being studied, field strength, hardware and software [68, 72, 73]. As a first step to evaluate the influence of Rician denoising, we have measured SNR in various anatomical areas of the brain and the spinal cord with and without Rician denoising. Our results show that Rician NLM filtering yields the significant increase of SNR in the brain DKI and the spinal cord DTI data (Table 1 and 2). The ROI-based SNR values with noise correction in our study are consistent with the previous study by Seo et al. [74] which has reported SNR thresholds 20 in the CC and 70 in the PUT for bias-free estimation of tensor metrics. It should be noted that those SNR values from the literatures were obtained from different acquisition protocols. Thus, when comparing those values with literature values, care should be taken to ensure the similarity of protocol chosen for comparison. For instance, the eddy current and off-resonance effects in a DWI sequence may substantially vary with b-value and diffusion gradient direction. Additionally, the spatial noise distribution can be varied by coil geometry, phase-encoding direction and acceleration factor of parallel imaging [38, 75, 76] which also need to be taken into account for. In our study, it is also observed that the degree of SNR improvement relates to the underlying structures. For instance, the CC has the smallest SNR increase among other regions, which might be attributed to the previous findings that the noise within accelerated images is nonhomogeneous with higher signal peripherally and noise centrally when parallel imaging is used [77,78,79]. As the levels and spatial distributions of noise are not an equal across the DW image, it is expected that various anatomical areas with different SNR requirements have diverse range of SNR increase rate.

While denoising DTI with low SNR addressing the strong influence of Rician bias in the brain has been well presented by different studies [48, 80, 81], Rician NLM denoising has not been established in the spinal cord DTI where most of previous studies have focused on data acquisition or motion correction methods to improve SNR [25, 27, 58, 82]. The mean SNR values without Rician denoising in C1-C3 in our study are within a range of values from a multi-centre study reported by Samson et al. (6.74 – 10.9) [27]. However, our mean SNR values significantly increased in both lateral and dorsal areas of the spinal cord levels C1-C6 after a Rician denoising (Table 2). Our results indicate that a simple noise correction method in the spinal cord such as Rician denoising being used in our study increases SNR (Table 2) and thereby can reduce error estimates (Fig. 4b). This is not surprising given the improved quantification of dMRI in the brain, however worth reporting for the growing need of the spinal cord DTI in the clinical research despite more intrinsic challenges in the spinal cord as compared to the brain.

Estimation errors

Significantly reduced number of erroneous pixels (black holes) was observed in AK, RK and MK using CLLS-R as compared to CLLS in the brain (Fig. 5). In particular, clear difference is observed in cortical GM, where the voxel values with the erroneously estimated tensor are lower using CLLS as compared to CLLS-R (Fig. 3a). Note that voxels with extreme (negative or zero tensor) values were excluded from the computation. Even after all of extremely erroneous voxels were removed, the ROI-based mean χ2 values significantly decreased over all ROIs when estimated using CLLS-R compared to CLLS in all examined anatomical regions (Fig. 3b), suggesting the impact of Rician denoising step towards KT estimation in the brain. Additionally, the mean χ2 values in the spinal cord show that Rician NLM filtering significantly reduced error estimates in both lateral and dorsal columns (Fig. 4b). Reduced χ2 values both in the brain and the spinal cord, imply that accuracy of tensor estimation is significantly improved with Rician NLM filtering (Figs. 3b and 4b).

Regional values of DT- and KT-derived metrics

Currently, a few reports in the literature are available for comparison with our results [54,55,56, 83, 84]. Table 5 provides an overview of those works in the literature that include values of parameter estimates. While CLLS estimation led to overestimated DT-derived metrics, especially in FA (Additional File 1: Supporting Information Table S1), which is a common problem in human DTI studies [68, 85, 90], the DT- and KT-derived metrics using CLLS-R are consistent with some of the previously reported values (Table 3). Nevertheless, it is observed that the discrepancy in diffusional metrics values exists amongst various studies in the literature. There are a few factors that may explain discrepancy between studies, from data acquisition to postprocessing perspective. In DTI, it has been well known that low SNR causes a bias in DT-derived metrics, leading to overestimation of FA [68, 72, 73]. When SNR is low, Rician noise does not only cause random fluctuations but also a signal dependent bias to the data, which may lead to difficulty in postprocessing such as tensor calculation. Therefore bias-free measurements require adequate SNR [85] and DTI studies often report SNR values to assure the reliability of the estimated metrics. However, SNR levels are not routinely reported in DKI, resulting in the challenges of comparing results between studies. Therefore it is desirable to ensure the DT- and KT-derived indices are estimated with adequate SNR levels before the comparison between studies. Our results show that the mean MK varied from 0.70 (PUT) to 1.27 (IC) while AK and RK varied from 0.58 (CC) to 0.92 (Cg) and from 0.70 (PUT) to 1.98 (CC), respectively, with a range of SNR levels (from 25.25 in CC to 60.75 in PUT) in various anatomical structures of the brain. Additionally, inter-subject differences substantially contribute the within-group variability [86], and may partially explain discrepancy between studies. In particular, various choice of ROIs among studies (i.e. selection of structure from only contiguous voxels with the highest values to the entire structure) hampers comparisons between studies. For instance, the difference of FA in the genu of the CC between 0.44 in [55] and 0.80 in [54], may be largely due to the ROI used for measurement. Therefore it is important to ensure the DT- and KT-derived indices are reliable across raters within-group, as ROI-based measurement often required rater decisions which may have impacted ROI placement between observers. Our results show that ICC values are near 1 for various ROIs over all metrics, indicating high reliability of ROI-based measurement performed in the brain. It should be also noted that regional variability of DTI values between publications may relate to age differences within-group aside from selection of ROIs, acquisition parameters and SNR. Considerable inter-subject variability of DTI parameters has been shown in previous studies that mostly reported the age dependence of DTI metrics [87,88,89]. Our subjects were young adults (mean age 24 ± 2) and this may contribute on variability of diffusion metrics as compared to those in the literature.

Table 5 Regional values in the healthy brain from the literature

In the spinal cord, there are numerous reports in the literature available for comparison with our results. In order to ensure the similarity of the acquisition sequences and parameters, here we focus on comparing our DT-derived values with those by Qian et al. [64] (a study that our raw data were originally obtained) and Samson et al. [27] (a multi-centre study with the compatible acquisition protocols as ours). Our FA values with Rician denoising are found to be significantly lower than those previous reported by Qian et al. (0.81–0.84; p < 0.05) while our MD, AD and RD values with Rician denoising are not significantly different from those from the same study [64]. This suggests that Rician NLM filtering might reduce overestimation of FA in the spinal cord, which is common observed in DTI of the brain [68, 85, 90], by decreasing estimation error. Our column specific measurements of MD are at the lower end of the previously reported range of 0.93–1.29 mm2/ms by Samson et al. [27] while FA measurements are higher than the range measured by the same study (0.59–0.63). Our RD values are lower than the range measured by Samson et al. (0.68–0.84 mm2/ms) [27] while AD values are at the higher end of the previously reported range of 1.43–2.22 mm2/ms by the same study. Overall, DTI metrics are likely to be within similar ranges, however difference among protocols and the associated SNR in each study must be taken into account when comparing between those DTI metrics.

Consideration for clinical applications

It is worth noting that the DKI data acquisition took around 20 min per person, because we obtained whole brain DKI data with 2 averages in order to evaluate quality of different processing methods through a whole brain with reasonable SNR. Our results showed significant decrease of χ2 values, indicating quality of DKI-derived maps improved through a whole brain within reasonable scan time when CLLS-R estimation was performed. Thus, by reducing the number of slices carefully selected for areas of interest, neurology- or neuroscience-related application studies should be feasible, which would last a clinically acceptable time frame (less than 10 min). Additionally, Rician NLM denoising in combination with CLLS can be readily implemented as it is based on a LSE algorithm available through existing commercial programs. Therefore, practical use of the combined Rician denoising method is widely expected in characterizing microarchitectural integrity of normal and pathological states.

Conclusions

We have demonstrated that a Rician denoising filter incorporated with CLLS (CLLS-R) significantly increases SNR while reducing estimation errors. Our results suggest that the combined postprocessing method may provide the capability to more reliably quantify tissue properties both in the brain and the spinal cord with adequate SNR levels while it is easy to implement. Future studies are warranted towards investigating clinical and neuroscientific applications using this method.

Availability of data and materials

All data are fully available upon reasonable request. The corresponding author should be contacted if someone wants to request the data.

Abbreviations

dMRI:

Diffusion magnetic resonance imaging

DTI:

Diffusion tensor imaging

WM:

White matter

DKI:

Diffusion kurtosis imaging

GM:

Gray matter

FA:

Fractional anisotropy

MD:

Mean diffusivity

RD:

Radial diffusivity

AD:

Axial diffusivity

MK:

Mean kurtosis

RK:

Radial kurtosis

AK:

Axial kurtosis

SNR:

Signal-to-noise ratio

NLM:

Nonlocal means

CLLS:

Constrained linear least squares

CLLS-R:

Constrained linear least squares with Rician denoising

LS:

Least-squares

EPI:

Echo-planar-imaging

AIR:

Automated image registration

ROIs:

Regions of interest

PUT:

Putamen

GP:

Globus pallidus

CC:

Corpus callosum

IC:

Internal capsule

EC:

External capsule

Cg:

Cingulum

χ 2 :

Chi-square

CSF:

Cerebrospinal fluid

ICC:

Intraclass correlation coefficient

References

  1. Basser PJ, Mattiello J, LeBihan D. MR diffusion tensor spectroscopy and imaging. Biophys J. 1994;66(1):259–67.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  2. Pierpaoli C, Jezzard P, Basser PJ, Barnett A, Di Chiro G. Diffusion tensor MR imaging of the human brain. Radiology. 1996;201(3):637–48.

    Article  CAS  PubMed  Google Scholar 

  3. Le Bihan D, Mangin JF, Poupon C, Clark CA, Pappata S, Molko N, Chabriat H. Diffusion tensor imaging: concepts and applications. J Magn Reson Imaging. 2001;13(4):534–46.

    Article  PubMed  Google Scholar 

  4. Poupon C, Mangin J, Clark CA, Frouin V, Regis J, Le Bihan D, Bloch I. Towards inference of human brain connectivity from MR diffusion tensor data. Med Image Anal. 2001;5(1):1–15.

    Article  CAS  PubMed  Google Scholar 

  5. Mori S, Zhang J. Principles of diffusion tensor imaging and its applications to basic neuroscience research. Neuron. 2006;51(5):527–39.

    Article  CAS  PubMed  Google Scholar 

  6. Mori S, Barker PB. Diffusion magnetic resonance imaging: its principle and applications. Anat Rec. 1999;257(3):102–9.

    Article  CAS  PubMed  Google Scholar 

  7. Tae WS, Ham BJ, Pyun SB, Kang SH, Kim BJ. Current clinical applications of diffusion-tensor imaging in neurological disorders. J Clin Neurol. 2018;14(2):129–40.

    Article  PubMed  PubMed Central  Google Scholar 

  8. Moura LM, Luccas R, de Paiva JPQ, Amaro E Jr, Leemans A, Leite CDC, Otaduy MCG, Conforto AB. Diffusion tensor imaging biomarkers to predict motor outcomes in stroke: a narrative review. Front Neurol. 2019;10:445.

    Article  PubMed  PubMed Central  Google Scholar 

  9. Alexander AL, Lee JE, Lazar M, Field AS. Diffusion tensor imaging of the brain. Neurotherapeutics. 2007;4(3):316–29.

    Article  PubMed  PubMed Central  Google Scholar 

  10. Basser PJ, Pierpaoli C. Microstructural and physiological features of tissues elucidated by quantitative-diffusion-tensor MRI. J Magn Reson B. 1996;111(3):209–19.

    Article  CAS  PubMed  Google Scholar 

  11. Pierpaoli C, Basser PJ. Toward a quantitative assessment of diffusion anisotropy. Magn Reson Med. 1996;36(6):893–906.

    Article  CAS  PubMed  Google Scholar 

  12. Jensen JH, Helpern JA, Ramani A, Lu H, Kaczynski K. Diffusional kurtosis imaging: the quantification of non-gaussian water diffusion by means of magnetic resonance imaging. Magn Reson Med. 2005;53(6):1432–40.

    Article  PubMed  Google Scholar 

  13. Lu H, Jensen JH, Ramani A, Helpern JA. Three-dimensional characterization of non-gaussian water diffusion in humans using diffusion kurtosis imaging. NMR Biomed. 2006;19(2):236–47.

    Article  PubMed  Google Scholar 

  14. Hui ES, Cheung MM, Qi L, Wu EX. Towards better MR characterization of neural tissues using directional diffusion kurtosis analysis. Neuroimage. 2008;42(1):122–34.

    Article  PubMed  Google Scholar 

  15. Jensen JH, Falangola MF, Hu C, Tabesh A, Rapalino O, Lo C, Helpern JA. Preliminary observations of increased diffusional kurtosis in human brain following recent cerebral infarction. NMR Biomed. 2010;24:452.

    Article  PubMed  PubMed Central  Google Scholar 

  16. Helpern JA, Adisetiyo V, Falangola MF, Hu C, Di Martino A, Williams K, Castellanos FX, Jensen JH. Preliminary evidence of altered gray and white matter microstructural development in the frontal lobe of adolescents with attention-deficit hyperactivity disorder: a diffusional kurtosis imaging study. J Magn Reson Imaging. 2011;33(1):17–23.

    Article  PubMed  PubMed Central  Google Scholar 

  17. Falangola MFJJ, Babb JS, Hu C, Castellanos FX, Di Martino A, Ferris SH, Helpern JA. Age-related non-Gaussian diffusion patterns in the prefrontal brain. J Magn Reson Imaging. 2008;28(6):1345–50.

    Article  PubMed  PubMed Central  Google Scholar 

  18. Cheung MM, Hui ES, Chan KC, Helpern JA, Qi L, Wu EX. Does diffusion kurtosis imaging lead to better neural tissue characterization? A rodent brain maturation study. Neuroimage. 2009;45(2):386–92.

    Article  PubMed  Google Scholar 

  19. Van Cauter S, Veraart J, Sijbers J, Peeters RR, Himmelreich U, De Keyzer F, Van Gool SW, Van Calenbergh F, De Vleeschouwer S, Van Hecke W, Sunaert S. Gliomas: diffusion kurtosis MR imaging in grading. Radiology. 2012;263(2):492–501.

    Article  PubMed  Google Scholar 

  20. Manjon JV, Carbonell-Caballero J, Lull JJ, Garcia-Marti G, Marti-Bonmati L, Robles M. MRI denoising using non-local means. Med Image Anal. 2008;12(4):514–23.

    Article  PubMed  Google Scholar 

  21. Aja-Fernandez S, Niethammer M, Kubicki M, Shenton ME, Westin CF. Restoration of DWI data using a Rician LMMSE estimator. IEEE Trans Med Imaging. 2008;27(10):1389–403.

    Article  PubMed  PubMed Central  Google Scholar 

  22. Skare S, Li T, Nordell B, Ingvar M. Noise considerations in the determination of diffusion tensor anisotropy. Magn Reson Imaging. 2000;18(6):659–69.

    Article  CAS  PubMed  Google Scholar 

  23. Gudbjartsson H, Patz S. The Rician distribution of noisy MRI data. Magn Reson Med. 1995;34(6):910–4.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  24. Rossi C, Boss A, Steidle G, Martirosian P, Klose U, Capuani S, Maraviglia B, Claussen CD, Schick F. Water diffusion anisotropy in white and gray matter of the human spinal cord. J Magn Reson Imaging. 2008;27(3):476–82.

    Article  PubMed  Google Scholar 

  25. Zhang BT, Li M, Yu LL, Dai YM, Yu SN, Jiang JL. Diffusion tensor imaging of spinal microstructure in healthy adults: improved resolution with the readout segmentation of long variable echo-trains. Neural Regen Res. 2017;12(12):2067–70.

    Article  PubMed  PubMed Central  Google Scholar 

  26. Ries M, Jones RA, Dousset V, Moonen CT. Diffusion tensor MRI of the spinal cord. Magn Reson Med. 2000;44(6):884–92.

    Article  CAS  PubMed  Google Scholar 

  27. Samson RS, Levy S, Schneider T, Smith AK, Smith SA, Cohen-Adad J, Gandini Wheeler-Kingshott CA. ZOOM or Non-ZOOM? Assessing spinal cord diffusion tensor imaging protocols for multi-centre studies. PLoS ONE. 2016;11(5):e0155557.

    Article  PubMed  PubMed Central  CAS  Google Scholar 

  28. Vedantam A, Jirjis MB, Schmit BD, Wang MC, Ulmer JL, Kurpad SN. Characterization and limitations of diffusion tensor imaging metrics in the cervical spinal cord in neurologically intact subjects. J Magn Reson Imaging. 2013;38(4):861–7.

    Article  PubMed  Google Scholar 

  29. Hendrix P, Griessenauer CJ, Cohen-Adad J, Rajasekaran S, Cauley KA, Shoja MM, Pezeshk P, Tubbs RS. Spinal diffusion tensor imaging: a comprehensive review with emphasis on spinal cord anatomy and clinical applications. Clin Anat. 2015;28(1):88–95.

    Article  PubMed  Google Scholar 

  30. Falangola MF, Jensen JH, Babb JS, Hu C, Castellanos FX, Di Martino A, Ferris SH, Helpern JA. Age-related non-Gaussian diffusion patterns in the prefrontal brain. J Magn Reson Imaging. 2008;28(6):1345–50.

    Article  PubMed  PubMed Central  Google Scholar 

  31. Brion V, Poupon C, Riff O, Aja-Fernandez S, Tristan-Vega A, Mangin JF, Le Bihan D, Poupon F. Parallel MRI noise correction: an extension of the LMMSE to non central chi distributions. Med Image Comput Comput Assist Interv. 2011;14(Pt 2):226–33.

    PubMed  Google Scholar 

  32. Parker GJ, Schnabel JA, Symms MR, Werring DJ, Barker GJ. Nonlinear smoothing for reduction of systematic and random errors in diffusion tensor imaging. J Magn Reson Imaging. 2000;11(6):702–10.

    Article  CAS  PubMed  Google Scholar 

  33. Tabelow K, Polzehl J, Spokoiny V, Voss HU. Diffusion tensor imaging: structural adaptive smoothing. Neuroimage. 2008;39(4):1763–73.

    Article  PubMed  Google Scholar 

  34. Ding Z, Gore JC, Anderson AW. Reduction of noise in diffusion tensor images using anisotropic smoothing. Magn Reson Med. 2005;53(2):485–90.

    Article  PubMed  Google Scholar 

  35. Veraart J, Novikov DS, Christiaens D, Ades-Aron B, Sijbers J, Fieremans E. Denoising of diffusion MRI using random matrix theory. Neuroimage. 2016;142:394–406.

    Article  PubMed  Google Scholar 

  36. Becker SM, Tabelow K, Voss HU, Anwander A, Heidemann RM, Polzehl J. Position-orientation adaptive smoothing of diffusion weighted magnetic resonance data (POAS). Med Image Anal. 2012;16(6):1142–55.

    Article  CAS  PubMed  Google Scholar 

  37. Wiest-Daessle N, Prima S, Coupe P, Morrissey SP, Barillot C. Non-local means variants for denoising of diffusion-weighted and diffusion tensor MRI. Med Image Comput Comput Assist Interv. 2007;10(Pt 2):344–51.

    PubMed  PubMed Central  Google Scholar 

  38. Manjon JV, Coupe P, Marti-Bonmati L, Collins DL, Robles M. Adaptive non-local means denoising of MR images with spatially varying noise levels. J Magn Reson Imaging. 2010;31(1):192–203.

    Article  PubMed  Google Scholar 

  39. Coupe P, Yger P, Prima S, Hellier P, Kervrann C, Barillot C. An optimized blockwise nonlocal means denoising filter for 3-D magnetic resonance images. IEEE Trans Med Imaging. 2008;27(4):425–41.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  40. Wiest-Daessle N, Prima S, Coupe P, Morrissey SP, Barillot C. Rician noise removal by non-Local Means filtering for low signal-to-noise ratio MRI: applications to DT-MRI. Medical image computing and computer-assisted intervention : MICCAI International Conference on Medical Image Computing and Computer-Assisted Intervention 2008;11(Pt 2):171–179.

  41. Coupe PYP, Prima S, Hellier P, Kervrann C, Barillot C. An optimized blockwise nonlocal means denoising filter for 3-D magnetic resonance images. IEEE Trans Med Imaging. 2008;27(4):425–41.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  42. Wu X, Liu S, Wu M, Sun H, Zhou J, Gong Q, Ding Z. Nonlocal denoising using anisotropic structure tensor for 3D MRI. Med Phys. 2013;40(10):101904.

    Article  PubMed  Google Scholar 

  43. Manjon JV, Coupe P, Concha L, Buades A, Collins DL, Robles M. Diffusion weighted image denoising using overcomplete local PCA. PLoS ONE. 2013;8(9):e73021.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  44. Wiest-Daesslé NPS, Coupé P, Morrissey SP, Barillot C. Rician noise removal by non-Local Means filtering for low signal-to-noise ratio MRI: applications to DT-MRI. Med Image Comput Comput Assist Interv. 2008;11(pt 2):171–9.

    PubMed  PubMed Central  Google Scholar 

  45. Zhou MX, Yan X, Xie HB, Zheng H, Xu D, Yang G. Evaluation of non-local means based denoising filters for diffusion kurtosis imaging using a new phantom. PLoS ONE. 2015;10(2):e0116986.

    Article  PubMed  PubMed Central  CAS  Google Scholar 

  46. Veraart J, Van Hecke W, Sijbers J. Constrained maximum likelihood estimation of the diffusion kurtosis tensor using a Rician noise model. Magn Reson Med. 2011;66(3):678–86.

    Article  PubMed  Google Scholar 

  47. Andersson JL. Maximum a posteriori estimation of diffusion tensor parameters using a Rician noise model: why, how and but. Neuroimage. 2008;42(4):1340–56.

    Article  PubMed  Google Scholar 

  48. Basu S, Fletcher T, Whitaker R. Rician noise removal in diffusion tensor MRI. Med Image Comput Comput Assist Interv. 2006;9(Pt 1):117–25.

    PubMed  Google Scholar 

  49. Descoteaux M, Wiest-Daessle N, Prima S, Barillot C, Deriche R. Impact of Rician adapted Non-Local Means filtering on HARDI. Medical image computing and computer-assisted intervention : MICCAI International Conference on Medical Image Computing and Computer-Assisted Intervention 2008;11(Pt 2):122–130

  50. Reischauer C, Gutzeit A. Image denoising substantially improves accuracy and precision of intravoxel incoherent motion parameter estimates. PLoS ONE. 2017;12(4):e0175106.

    Article  PubMed  PubMed Central  CAS  Google Scholar 

  51. Gonzalez JE, Thompson PM, Zhao A, Tu Z. Modeling diffusion-weighted MRI as a spatially variant gaussian mixture: application to image denoising. Med Phys. 2011;38(7):4350–64.

    Article  PubMed  PubMed Central  Google Scholar 

  52. Tabesh A, Jensen JH, Ardekani BA, Helpern JA. Estimation of tensors and tensor-derived measures in diffusional kurtosis imaging. Magn Reson Med. 2011;65(3):823–36.

    Article  PubMed  Google Scholar 

  53. Su B, Liu Q, Chen J, Wu X. Non-local mean denoising in diffusion tensor space. Exp Ther Med. 2014;8(2):447–53.

    Article  PubMed  PubMed Central  Google Scholar 

  54. Latt J, Nilsson M, Wirestam R, Stahlberg F, Karlsson N, Johansson M, Sundgren P, van Westen D. Regional values of diffusion kurtosis estimates in the healthy brain. J Magn Reson Imaging. 2013;37:610–8.

    Article  PubMed  Google Scholar 

  55. Das SK, Wang JL, Bing L, Bhetuwal A, Yang HF. Regional Values of Diffusional Kurtosis Estimates in the Healthy Brain during Normal Aging. Clin Neuroradiol. 2017;27(3):283–98.

    Article  CAS  PubMed  Google Scholar 

  56. Andre ED, Grinberg F, Farrher E, Maximov II, Shah NJ, Meyer C, Jaspar M, Muto V, Phillips C, Balteau E. Influence of noise correction on intra- and inter-subject variability of quantitative metrics in diffusion kurtosis imaging. PLoS ONE. 2014;9(4):e94531.

    Article  PubMed  PubMed Central  CAS  Google Scholar 

  57. Tu TW, Budde MD, Xie M, Chen YJ, Wang Q, Quirk JD, Song SK. Phase-aligned multiple spin-echo averaging: a simple way to improve signal-to-noise ratio of in vivo mouse spinal cord diffusion tensor image. Magn Reson Imaging. 2014;32(10):1335–43.

    Article  PubMed  PubMed Central  Google Scholar 

  58. Wheeler-Kingshott CA, Hickman SJ, Parker GJ, Ciccarelli O, Symms MR, Miller DH, Barker GJ. Investigating cervical spinal cord structure using axial diffusion tensor imaging. Neuroimage. 2002;16(1):93–102.

    Article  PubMed  Google Scholar 

  59. Saritas EU, Cunningham CH, Lee JH, Han ET, Nishimura DG. DWI of the spinal cord with reduced FOV single-shot EPI. Magn Reson Med. 2008;60(2):468–73.

    Article  PubMed  Google Scholar 

  60. Cohen-Adad J, El Mendili MM, Lehericy S, Pradat PF, Blancho S, Rossignol S, Benali H. Demyelination and degeneration in the injured human spinal cord detected with diffusion and magnetization transfer MRI. Neuroimage. 2011;55(3):1024–33.

    Article  CAS  PubMed  Google Scholar 

  61. Xu J, Shimony JS, Klawiter EC, Snyder AZ, Trinkaus K, Naismith RT, Benzinger TL, Cross AH, Song SK. Improved in vivo diffusion tensor imaging of human cervical spinal cord. Neuroimage. 2013;67:64–76.

    Article  PubMed  Google Scholar 

  62. Liu C, Bammer R, Acar B, Moseley ME. Characterizing non-Gaussian diffusion by using generalized diffusion tensors. Magn Reson Med. 2004;51(5):924–37.

    Article  PubMed  Google Scholar 

  63. Veraart J, Sijbers J, Sunaert S, Leemans A, Jeurissen B. Weighted linear least squares estimation of diffusion MRI parameters: strengths, limitations, and pitfalls. Neuroimage. 2013;81:335–46.

    Article  PubMed  Google Scholar 

  64. Qian W, Chan Q, Mak H, Zhang Z, Anthony MP, Yau KK, Khong PL, Chan KH, Kim M. Quantitative assessment of the cervical spinal cord damage in neuromyelitis optica using diffusion tensor imaging at 3 Tesla. J Magn Reson Imaging. 2011;33(6):1312–20.

    Article  PubMed  Google Scholar 

  65. Glenn GR, Tabesh A, Jensen JH. A simple noise correction scheme for diffusional kurtosis imaging. Magn Reson Imaging. 2015;33(1):124–33.

    Article  PubMed  Google Scholar 

  66. Jensen JH, Falangola MF, Hu C, Tabesh A, Rapalino O, Lo C, Helpern JA. Preliminary observations of increased diffusional kurtosis in human brain following recent cerebral infarction. NMR Biomed. 2011;24(5):452–7.

    Article  PubMed  Google Scholar 

  67. Woods RP, Grafton ST, Holmes CJ, Cherry SR, Mazziotta JC. Automated image registration: I. General methods and intrasubject, intramodality validation. J Comput Assist Tomogr. 1998;22(1):139–52.

    Article  CAS  PubMed  Google Scholar 

  68. Farrell JA, Landman BA, Jones CK, Smith SA, Prince JL, van Zijl PC, Mori S. Effects of signal-to-noise ratio on the accuracy and reproducibility of diffusion tensor imaging-derived fractional anisotropy, mean diffusivity, and principal eigenvector measurements at 1.5 T. J Magn Reson Imaging. 2007;26(3):756–67.

    Article  PubMed  PubMed Central  Google Scholar 

  69. Tamraz JC, Comair Y. Atlas of Regional Anatomy of the Brain Using MRI: With Functional Correlations. Berlin: Springer; 2000.

    Google Scholar 

  70. Shrout PE, Fleiss JL. Intraclass correlations: uses in assessing rater reliability. Psychol Bull. 1979;86(2):420–8.

    Article  CAS  PubMed  Google Scholar 

  71. Wang ZJ, Chia JM, Ahmed S, Rollins NK. Signal-to-noise assessment for diffusion tensor imaging with single data set and validation using a difference image method with data from a multicenter study. Med Phys. 2014;41(9):092302.

    Article  PubMed  PubMed Central  Google Scholar 

  72. Anderson AW. Theoretical analysis of the effects of noise on diffusion tensor imaging. Magn Reson Med. 2001;46(6):1174–88.

    Article  CAS  PubMed  Google Scholar 

  73. Koay CG, Carew JD, Alexander AL, Basser PJ, Meyerand ME. Investigation of anomalous estimates of tensor-derived quantities in diffusion tensor imaging. Magn Reson Med. 2006;55(4):930–6.

    Article  PubMed  Google Scholar 

  74. Seo Y, Wang ZJ, Morriss MC, Rollins NK. Minimum SNR and acquisition for bias-free estimation of fractional anisotropy in diffusion tensor imaging: a comparison of two analytical techniques and field strengths. Magn Reson Imaging. 2012;30(8):1123–33.

    Article  PubMed  Google Scholar 

  75. Dietrich O, Raya JG, Reeder SB, Reiser MF, Schoenberg SO. Measurement of signal-to-noise ratios in MR images: influence of multichannel coils, parallel imaging, and reconstruction filters. J Magn Reson Imaging. 2007;26(2):375–85.

    Article  PubMed  Google Scholar 

  76. Goerner FL, Clarke GD. Measuring signal-to-noise ratio in partially parallel imaging MRI. Med Phys. 2011;38(9):5049–57.

    Article  PubMed  PubMed Central  Google Scholar 

  77. Lin FH, Chen YJ, Belliveau JW, Wald LL. A wavelet-based approximation of surface coil sensitivity profiles for correction of image intensity inhomogeneity and parallel imaging reconstruction. Hum Brain Mapp. 2003;19(2):96–111.

    Article  PubMed  PubMed Central  Google Scholar 

  78. Hiratsuka Y, Miki H, Kikuchi K, Kiriyama I, Mochizuki T, Takahashi S, Sadamoto K. Sensitivity of an eight-element phased array coil in 3 Tesla MR imaging: a basic analysis. Magn Reson Med Sci. 2007;6(3):177–81.

    Article  PubMed  Google Scholar 

  79. Roemer PB, Edelstein WA, Hayes CE, Souza SP, Mueller OM. The NMR phased array. Magn Reson Med. 1990;16(2):192–225.

    Article  CAS  PubMed  Google Scholar 

  80. Fillard P, Pennec X, Arsigny V, Ayache N. Clinical DT-MRI estimation, smoothing, and fiber tracking with log-Euclidean metrics. IEEE Trans Med Imaging. 2007;26(11):1472–82.

    Article  PubMed  Google Scholar 

  81. Martin-Fernandez M, Munoz-Moreno E, Cammoun L, Thiran JP, Westin CF, Alberola-Lopez C. Sequential anisotropic multichannel Wiener filtering with Rician bias correction applied to 3D regularization of DWI data. Med Image Anal. 2009;13(1):19–35.

    Article  CAS  PubMed  Google Scholar 

  82. Jeong EK, Kim SE, Guo J, Kholmovski EG, Parker DL. High-resolution DTI with 2D interleaved multislice reduced FOV single-shot diffusion-weighted EPI (2D ss-rFOV-DWEPI). Magn Reson Med. 2005;54(6):1575–9.

    Article  PubMed  Google Scholar 

  83. Benitez A, Jensen JH, Falangola MF, Nietert PJ, Helpern JA. Modeling white matter tract integrity in aging with diffusional kurtosis imaging. Neurobiol Aging. 2018;70:265–75.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  84. Jensen JH, Helpern JA. MRI quantification of non-Gaussian water diffusion by kurtosis analysis. NMR Biomed. 2010;23(7):698–710.

    Article  PubMed  PubMed Central  Google Scholar 

  85. Seo Y, Rollins NK, Wang ZJ. Reduction of bias in the evaluation of fractional anisotropy and mean diffusivity in magnetic resonance diffusion tensor imaging using region-of-interest methodology. Sci Rep. 2019;9(1):13095.

    Article  PubMed  PubMed Central  CAS  Google Scholar 

  86. Szczepankiewicz F, Latt J, Wirestam R, Leemans A, Sundgren P, van Westen D, Stahlberg F, Nilsson M. Variability in diffusion kurtosis imaging: impact on study design, statistical power and interpretation. Neuroimage. 2013;76:145–54.

    Article  PubMed  Google Scholar 

  87. Lebel C, Walker L, Leemans A, Phillips L, Beaulieu C. Microstructural maturation of the human brain from childhood to adulthood. Neuroimage. 2008;40(3):1044–55.

    Article  CAS  PubMed  Google Scholar 

  88. Schneiderman JS, Buchsbaum MS, Haznedar MM, Hazlett EA, Brickman AM, Shihabuddin L, Brand JG, Torosjan Y, Newmark RE, Tang C, Aronowitz J, Paul-Odouard R, Byne W, Hof PR. Diffusion tensor anisotropy in adolescents and adults. Neuropsychobiology. 2007;55(2):96–111.

    Article  PubMed  Google Scholar 

  89. Lebel C, Gee M, Camicioli R, Wieler M, Martin W, Beaulieu C. Diffusion tensor imaging of white matter tract evolution over the lifespan. Neuroimage. 2012;60(1):340–52.

    Article  CAS  PubMed  Google Scholar 

  90. Polders DL, Leemans A, Hendrikse J, Donahue MJ, Luijten PR, Hoogduin JM. Signal to noise ratio and uncertainty in diffusion tensor imaging at 1.5, 3.0, and 7.0 Tesla. J Magn Reson Imaging. 2011;33(6):1456–63.

    Article  PubMed  Google Scholar 

Download references

Acknowledgements

The authors would like to thank Dr. Marco Battiston for constructive comments.

Funding

This work was supported by a grant from the University of Hong Kong funding program for basic research. Funding was utilized for data acquisition.

Author information

Authors and Affiliations

Authors

Contributions

Study concepts and design: ZZ and MK; literature research and data collection: ZZ, DV, WQ and MK; statistical analysis: ZZ and DV; manuscript preparation: ZZ and DV; manuscript editing: WQ and MK. All authors have read and approved the manuscript.

Corresponding author

Correspondence to Mina Kim.

Ethics declarations

Ethics approval and consent to participate

This study was approved by the University of Hong Kong Institutional Review Board (IRB) and all subjects provided the signed consent.

Consent for publication

Not applicable.

Competing interests

The authors declare no competing interests.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Supplementary Information

Additional file 1.

Supporting Information Table S1 showing comparison for DT- and KT-derived metrics between CLLS and CLLS-R analyses.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Zhang, Z., Vernekar, D., Qian, W. et al. Non-local means based Rician noise filtering for diffusion tensor and kurtosis imaging in human brain and spinal cord. BMC Med Imaging 21, 16 (2021). https://doi.org/10.1186/s12880-021-00549-9

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s12880-021-00549-9

Keywords