Evaluation of patients with painful total hip arthroplasty using combined single photon emission tomography and conventional computerized tomography (SPECT/CT) – a comparison of semi-quantitative versus 3D volumetric quantitative measurements

Abstract

Background

It was the primary purpose of our study to evaluate the inter- and intra-observer reliability of a standardized SPECT/CT algorithm for evaluating patients with painful primary total hip arthroplasty (THA). The secondary purpose was a comparison of semi-quantitative and 3D volumetric quantification method for assessment of bone tracer uptake (BTU) in those patients.

Methods

A novel SPECT/CT localization scheme consisting of 14 femoral and 4 acetabular regions on standardized axial and coronal slices was introduced and evaluated in terms of inter- and intra-observer reliability in 37 consecutive patients with hip pain after THA. BTU for each anatomical region was assessed semi-quantitatively using a color-coded Likert type scale (0-10) and volumetrically quantified using a validated software. Two observers interpreted the SPECT/CT findings in all patients two times with six weeks interval between interpretations in random order. Semi-quantitative and quantitative measurements were compared in terms of reliability. In addition, the values were correlated using Pearson`s correlation. A factorial cluster analysis of BTU was performed to identify clinically relevant regions, which should be grouped and analysed together.

Results

The localization scheme showed high inter- and intra-observer reliabilities for all femoral and acetabular regions independent of the measurement method used (semiquantitative versus 3D volumetric quantitative measurements). A high to moderate correlation between both measurement methods was shown for the distal femur, the proximal femur and the acetabular cup. The factorial cluster analysis showed that the anatomical regions might be summarized into three distinct anatomical regions. These were the proximal femur, the distal femur and the acetabular cup region.

Conclusions

The SPECT/CT algorithm for assessment of patients with pain after THA is highly reliable independent from the measurement method used. Three clinically relevant anatomical regions (proximal femoral, distal femoral, acetabular) were identified.

Peer Review reports

Background

Combined single photon emission computerized tomography and conventional CT (SPECT/CT) promises the combined assessment of anatomical and functional information and hence its value is increasingly recognized in orthopaedics [1,2,3,4,5,6,7,8,9,10,11,12,13,14,15,16,17,18,19,20,21,22,23].

SPECT/CT has been reported to be beneficial in identifying the cause of patients`pain after total knee arthroplasty, patients with chondral or osteochondral lesions, before and after high tibial osteotomy and after ACL reconstruction [1,2,3,4,5,6,7,8,9,10,11,12,13,14,15,16,17,18,19,20,21,22,23]. Although SPECT/CT has also been used in patients with pain after total hip arthroplasty (THA) there is only scarce evidence about the optimal diagnostic algorithm and method of bone tracer uptake (BTU) analysis [24,25,26,27].

Due to its specific characteristics SPECT/CT is more sensitive and specific than SPECT and CT alone. Clearly, it is the accurate anatomical localization of the SPECT-tracer uptake using the CT as reference map that promises improved diagnostic confidence, particularly in patients with pain after joint replacement surgery [17, 20, 23]. Detection of mechanical or septic loosening of THA even in early stages might be facilitated. In addition, it could provide the surgeon with information on the position of THA components.

Recently our group has described and validated a standardized diagnostic algorithm using SPECT/CT in patients with pain after total knee arthroplasty including analysis of bone tracer activity and position and alignment of the TKA [14, 17, 20, 23, 28]. However, no such diagnostic algorithm has been reported or validated in patients with pain after THA.

In addition, in clinical practice it is a pertinent question if quantification of BTU in SPECT/CT offers so much more information than formerly semi-quantitative methods to recommend its daily use. There is no current study comparing semi-quantitative and quantitative measurement of BTU in patients with pain after THA.

Hence, it was the primary purpose of our study to evaluate the inter- and intra-observer reliability of a standardized SPECT/CT algorithm for evaluating patients with painful primary THA. The secondary purpose was a comparison of semi-quantitative and 3D volumetric quantification method for assessment of BTU in those patients.

Methods

Patients

A consecutive series of 37 patients (m:f = 16:21, mean age ± standard deviation 71 ± 11 years) presenting with hip pain after primary THA were prospectively collected and retrospectively included in this study. The data from all this patients were collected using our clinical information system (KIS, Erne, Switzerland). All patients had a primary THA with a maximum interval of 6 months from primary THA.

The study was approved by the local ethical committee (EKNZ 205/10). Written informed consent was obtained from all patients.

SPECT/CT was performed using a hybrid system (Symbia T16, Siemens, Erlangen, Germany) with a dual-head gamma camera and an integrated, 16x0.75-mm slice-thickness CT. All patients received a commercial 500–700 MBq Tc-99m-HDP injection (Malinckrodt, Wollerau, Switzerland). Scintigraphic images in anterior-posterior and lateral projection were taken in the perfusion phase (immediately after injection), the soft tissue phase (3–5 min after injection) and the delayed metabolic phase (2–3 h after injection). SPECT/CT was performed with a matrix size of 128x128, an angle step of 32, and a time per frame of 25 s.

The CT protocol was modified according to the Imperial Knee Protocol, which is a low dose CT protocol that includes high-resolution 0.75 mm slices of the knee and 3 mm slices of the hip and ankle joints [29].

The localization of bone tracer activity was recorded on a standardized localization scheme developed for use in patients after primary THA (Fig. 1). This defines biomechanically relevant regions of the femoral shaft and acetabulum around total hip prosthesis on standardized axial, coronal, and sagittal slices to accurately map areas of increased activity. The anatomical area (femur, acetabulum) is indicated with capital letters (F, A). The femur (F) is divided into fourteen zones with regards to the modified Gruen classification [30].

Fig. 1
figure 1

The localization scheme for the Tc-99m-HDP bone tracer activity in patients with painful hips after primary THA

Each femoral zone is represented with a number (1-14). The acetabulum (A) is divided into four zones (superior anterior, superior posterior, inferior anterior, inferior posterior).

The highest activity grading on SPECT/CT for each area of the localization scheme was recorded semiquantitatively (0–10). In addition, it was noted whether the area of tracer activity extended to the bone prosthesis interface. In that case an additional “c” was added to the tracer activity value.

In addition, BTU was also quantified in 3D using a voxel based measurement method. For BTU analysis (intensity and anatomical distribution pattern) the 3D reconstructed datasets of the delayed SPECT/CT images were used. The anatomical areas represented by a previously validated localization scheme were 3D volumetrically measured in terms of SPECT/CT tracer uptake values (OrthoImagingSolutions Ltd., London, UK) [4, 9]. The tracer activity was quantified in 3D volumetrically as described in Hirschmann et al. (Figs. 1, 2 and 3) [4]. The maximum intensity values were recorded for each anatomical area.

Fig. 2
figure 2

3D volumetric quantification of acetabular cup areas with regards to the localization scheme using the customized software

Fig. 3
figure 3

3D volumetric quantification of femoral areas with regards to the localization scheme using the customized software

Two observers interpreted the SPECT/CT findings in all patients two times with six weeks interval between interpretations in random order. Both were blinded to results from previous observations. The inter- and intra-observer reliability of the localization scheme and grading of the tracer activity was determined. Semi-quantitative and quantitative measurements were compared in terms of reliability. In addition, the values were correlated using Pearson`s correlation.

Finally, a factorial cluster analysis of BTU was performed to identify clinically relevant regions, which should be grouped and analysed together.

Statistical analysis

Data were analyzed using SPSS 16.0 (SPSS, Chicago, U.S.A.). Sample size was calculated according to the reported estimates for reliability studies using intraclass correlation coefficients (ICCs) [31].

The median differences in measurements between the two observers (inter-observer) and within the measurements of the first observer (intra-observer) were calculated. The intraclass correlation coefficients for inter- and intra-observer reliability were also calculated. ICC values range from 0 to 1. A value of 1 indicates perfect reliability, 0.81 to 1 very good reliability and 0.61-0.80 good reliability [31]. For all analysis, p < 0.05 was considered statistically significant.

Results

The localization scheme showed high inter- and intra-observer reliabilities for all femoral and acetabular regions independent of the measurement methods (semi-quantitative versus 3D volumetric quantitative measurements). Inter- and intra-observer reliability (intra class correlation- ICC) of 99mTc-HDP-SPECT/CT tracer activity using the localization and semi-quantitative BTU grading scheme for the acetabular and femoral zones (Likert scale 0–10) are presented in Table 1. In mean the femoral regions showed an ICC of 0.981–0.992 for intra-observer reliability and 0.871 for inter-observer-reliability. In mean the acetabular regions showed an ICC of 0.967–0.975 for intra-observer reliability and 0.877 for inter-observer-reliability. The acetabular regions AI and PI showed moderate agreement for intra- (ICC 0.529–0.963) and inter-observer testing (0.401–0.493).

Table 1 Inter- and intra-observer reliability (intra class correlation- ICC) of 99mTc-HDP-BTU activity using the localization and semiquantitative BTU grading scheme for the acetabular and femoral zones (Likert scale 0-10)

Inter- and intra-observer reliability (intra class correlation- ICC) of 99mTc-HDP-SPECT/CT tracer activity using the localization and 3D voxel based quantitative BTU grading scheme for the acetabular and femoral zones were for all regions neary perfect (ICCs > 0.90).

The measured values of BTU activity in SPECT/CT for each anatomical region using the semi-quantitative versus 3D volumetric quantitative method are shown in Table 2. The Pearson`s correlation of both BTU measurement methods is presented in Table 3. A high correlation between both measurement methods was found for the distal femur. A moderate correlation was found for the proximal femur and the acetabular cup regions.

Table 2 Absolute values (mean, SD, median) of BTU for each anatomical area using 3D volumetric voxel based quantification and semi-quantitative measurement methods
Table 3 Pearson`s correlations of 99mTc-HDP BTU activity for the acetabular and femoral zones using 3D-quantitative and semi-quantitative methods

The factorial cluster analysis showed that the anatomical regions might be summarized into three distinct anatomical regions Table 4. These were the proximal femur, the distal femur and the acetabular cup region.

Table 4 Factorial cluster analysis of 3D voxel based quantified BTU values showing three distinct regions (proximal thigh, distal thigh, acetabular cup)

Discussion

The most important findings of the present study were twofold. Firstly, a high inter-observer and intra-observer reliability was found for grading and localization of the BTU activity independent of the investigated region. The localization scheme and BTU grading was reliable and easily applicable, which would make it understandable by most clinicians. A reliable localization and grading scheme is needed to standardize the evaluation of SPECT/CT data and make those comparable with each other. The Gruen classification is already widely used for assessment of periprosthetic radiolucencies, hence it was decided to adapt this scheme to the biomechanics of the hip reflecting bone remodeling and integration of the prosthetic hip components. It has also been used by others to report BTU findings in SPECT/CT [26, 32].

In a recent pictorial review by Tam et al. dealing with THA the authors reported their SPECT/CT analysis and reporting system, which is in accordance with the one presented in terms of the localization scheme used [26]. However, there only the two-dimensional localization scheme was used. In this study a modified three-dimensional localization scheme was introduced and has proven highly reliable [26, 33]. The localization scheme showed high inter- and intra-observer reliabilities for both femoral and acetabular regions independent of the measurement methods (semi-quantitative versus 3D volumetric quantitative measurements). Clearly, the 3D volumetric quantification has proven to be as reliable as the standard two-dimensional localization and BTU analysis system.

In a retrospective study Jin et al. investigated the periprosthetic bone remodeling of THA using SPECT/CT [34]. SPECT/CT was reviewed as three-dimensional multiplanar reconstructions with a slice thickness of 4.4 mm [34]. Two-dimensional regions of interest (ROIs) were generated and placed in specific standardized locations for each dataset [34]. All ROI placements and measurements were performed by a single reader to match standardized locations and with the assistance of a ROI template guide [34]. In agreement with the present study they also normalized the absolute measured values by building ratios of the measured value and a value measured at a specific reference regions [34].

Secondly, a high correlation between both measurement methods was found for the distal femur. A moderate correlation was found for the proximal femur and the acetabular cup regions.

The factorial cluster analysis showed that the anatomical regions might be summarized into three distinct anatomical regions These were the proximal femur, the distal femur and the acetabular cup region. In the study by Jin et al. the ROI analysis was done at five different locations (the greater trochanter, the femoral calcar, the mid-stem of the femur, the femoral stem tip and one acetabular region) [34]. The authors choose these locations for analysis as these appeared to be clinically relevant and highly reproducible [34].

However, Jin et al. questioned the need for routine quantification of BTU in patients with THA [34]. Based on their findings it is possible to distinguish between clearly normal and clearly abnormal SPECT/CT images [34]. In difficult cases semi-quantification might be helpful.

In contrast, we believe that a better understanding of bone remodelling after THA reflected by typical BTU pattern distribution will help to improve the reporting and diagnosis when using SPECT/CT. However, until analysis of BTU activity could lead to a better diagnosis we need to achieve a more profound knowledge of normal and abnormal BTU distribution and activity in native and arthroplasty patients.

Another limitation to gain wider acceptance for quantification methods among clinicians is the utility, availability and simplicity of these analysis methods. Clearly, these have to be robust, reliable and easy to perform.

The study bears a few limitations to be considered. This is a well sized small pilot study aiming to evaluate the analysis algorithm in patients with THA undergoing SPECT/CT. The clinical value of the algorithm needs to be further evaluated in larger, homogenous cohorts. The standard deviations are high, which is due to inter-patient variability. A typical findings in metabolic imaging.

Conclusions

The SPECT/CT algorithm for assessment of patients with pain after THA is highly reliable independent from the measurement method used. Three clinically relevant anatomical regions (proximal femoral, distal femoral, acetabular) were identified.

Abbreviations

BTU:

Bone tracer uptake

ICC:

Intraclass coefficient

ROI:

Region of interest

SPECT/CT:

Single photon emission computerized tomography and conventional computerized tomography

THA:

Total hip arthroplasty

References

  1. Suter B, Testa E, Stampfli P, Konala P, Rasch H, Friederich NF, Hirschmann MT. A novel standardized algorithm using SPECT/CT evaluating unhappy patients after unicondylar knee arthroplasty--a combined analysis of tracer uptake distribution and component position. BMC Med Imaging. 2015;15:11.

    Article  PubMed  PubMed Central  Google Scholar 

  2. Schon SN, Afifi FK, Rasch H, Amsler F, Friederich NF, Arnold MP, Hirschmann MT. Assessment of in vivo loading history of the patellofemoral joint: a study combining patellar position, tilt, alignment and bone SPECT/CT. Knee Surg Sports Traumatol Arthrosc. 2014;22(12):3039–46.

    Article  PubMed  Google Scholar 

  3. Rieger B, Friederich NF, Rasch H, Hirschmann MT. Non-dislocated osteoporotic insufficiency fracture of the medial femoral neck. SPECT/CT makes the diagnostic difference. Unfallchirurg. 2014;117(4):369–73.

    Article  CAS  PubMed  Google Scholar 

  4. Rasch H, Falkowski AL, Forrer F, Henckel J, Hirschmann MT. 4D-SPECT/CT in orthopaedics: a new method of combined quantitative volumetric 3D analysis of SPECT/CT tracer uptake and component position measurements in patients after total knee arthroplasty. Skeletal Radiol. 2013;42(9):1215–23.

    Article  PubMed  Google Scholar 

  5. Mucha A, Dordevic M, Testa EA, Rasch H, Hirschmann MT. Assessment of the loading history of patients after high tibial osteotomy using SPECT/CT--a new diagnostic tool and algorithm. J Orthop Surg Res. 2013;8:46.

    Article  PubMed  PubMed Central  Google Scholar 

  6. Mucha A, Dordevic M, Hirschmann A, Rasch H, Amsler F, Arnold MP, Hirschmann MT. Effect of high tibial osteotomy on joint loading in symptomatic patients with varus aligned knees: a study using SPECT/CT. Knee Surg Sports Traumatol Arthrosc. 2015;23(8):2315–23.

    Article  PubMed  Google Scholar 

  7. Mathis DT, Rasch H, Hirschmann MT. In vivo bone tunnel remodeling in symptomatic patients after ACL reconstruction: a retrospective comparison of articular and extra-articular fixation. Muscles Ligaments Tendons J. 2015;5(4):316–24.

    PubMed  Google Scholar 

  8. Konala P, Iranpour F, Kerner A, Rasch H, Friederich NF, Hirschmann MT. Clinical benefit of SPECT/CT for follow-up of surgical treatment of osteochondritis dissecans. Ann Nucl Med. 2010;24(8):621–4.

    Article  PubMed  Google Scholar 

  9. Hirschmann MT, Wagner CR, Rasch H, Henckel J. Standardized volumetric 3D-analysis of SPECT/CT imaging in orthopaedics: overcoming the limitations of qualitative 2D analysis. BMC Med Imaging. 2012;12:5.

    Article  PubMed  PubMed Central  Google Scholar 

  10. Hirschmann MT, Schon S, Afifi FK, Amsler F, Rasch H, Friederich NF, Arnold MP. Assessment of loading history of compartments in the knee using bone SPECT/CT: a study combining alignment and 99mTc-HDP tracer uptake/distribution patterns. J Orthop Res. 2013;31(2):268–74.

    Article  CAS  PubMed  Google Scholar 

  11. Hirschmann MT, Schmid R, Dhawan R, Skarvan J, Rasch H, Friederich NF, Emery R. Combined single photon emission computerized tomography and conventional computerized tomography: Clinical value for the shoulder surgeons? Int J Shoulder Surg. 2011;5(3):72–6.

    Article  PubMed  PubMed Central  Google Scholar 

  12. Hirschmann MT, Mathis D, Rasch H, Amsler F, Friederich NF, Arnold MP. SPECT/CT tracer uptake is influenced by tunnel orientation and position of the femoral and tibial ACL graft insertion site. Int Orthop. 2013;37(2):301–9.

    Article  PubMed  Google Scholar 

  13. Hirschmann MT, Mathis D, Afifi FK, Rasch H, Henckel J, Amsler F, Wagner CR, Friederich NF, Arnold MP. Single photon emission computerized tomography and conventional computerized tomography (SPECT/CT) for evaluation of patients after anterior cruciate ligament reconstruction: a novel standardized algorithm combining mechanical and metabolic information. Knee Surg Sports Traumatol Arthrosc. 2013;21(4):965–74.

    Article  PubMed  Google Scholar 

  14. Hirschmann MT, Konala P, Iranpour F, Kerner A, Rasch H, Friederich NF. Clinical value of SPECT/CT for evaluation of patients with painful knees after total knee arthroplasty--a new dimension of diagnostics? BMC Musculoskelet Disord. 2011;12:36.

    Article  PubMed  PubMed Central  Google Scholar 

  15. Hirschmann MT, Iranpour F, Konala P, Kerner A, Rasch H, Cobb JP, Friederich NF. A novel standardized algorithm for evaluating patients with painful total knee arthroplasty using combined single photon emission tomography and conventional computerized tomography. Knee Surg Sports Traumatol Arthrosc. 2010;18(7):939–44.

    Article  PubMed  Google Scholar 

  16. Hirschmann MT, Iranpour F, Davda K, Rasch H, Hugli R, Friederich NF. Combined single-photon emission computerized tomography and conventional computerized tomography (SPECT/CT): clinical value for the knee surgeons? Knee Surg Sports Traumatol Arthrosc. 2010;18(3):341–5.

    Article  PubMed  Google Scholar 

  17. Hirschmann MT, Henckel J, Rasch H. SPECT/CT in patients with painful knee arthroplasty-what is the evidence? Skeletal Radiol. 2013;42(9):1201–7.

    Article  PubMed  Google Scholar 

  18. Hirschmann MT, Davda K, Rasch H, Arnold MP, Friederich NF. Clinical value of combined single photon emission computerized tomography and conventional computer tomography (SPECT/CT) in sports medicine. Sports Med Arthrosc. 2011;19(2):174–81.

    Article  PubMed  Google Scholar 

  19. Hirschmann MT, Davda K, Iranpour F, Rasch H, Friederich NF. Combined single photon emission computerised tomography and conventional computerised tomography (SPECT/CT) in patellofemoral disorders: a clinical review. Int Orthop. 2011;35(5):675–80.

    Article  PubMed  Google Scholar 

  20. Hirschmann MT, Amsler F, Rasch H. Clinical value of SPECT/CT in the painful total knee arthroplasty (TKA): a prospective study in a consecutive series of 100 TKA. Eur J Nucl Med Mol Imaging. 2015;42(12):1869–82.

    Article  PubMed  Google Scholar 

  21. Hassink G, Testa EA, Leumann A, Hugle T, Rasch H, Hirschmann MT. Intra- and inter-observer reliability of a new standardized diagnostic method using SPECT/CT in patients with osteochondral lesions of the ankle joint. BMC Med Imaging. 2016;16(1):67.

    Article  PubMed  PubMed Central  Google Scholar 

  22. Dordevic M, Hirschmann MT, Rechsteiner J, Falkowski A, Testa E, Hirschmann A. Do Chondral Lesions of the Knee Correlate with Bone Tracer Uptake by Using SPECT/CT? Radiology. 2016;278(1):223–31.

    Article  PubMed  Google Scholar 

  23. Awengen R, Rasch H, Amsler F, Hirschmann MT. Symptomatic versus asymptomatic knees after bilateral total knee arthroplasty: what is the difference in SPECT/CT? Eur J Nucl Med Mol Imaging. 2016;43(4):762–72.

    Article  CAS  PubMed  Google Scholar 

  24. Strobel K, Steurer-Dober I, Huellner MW, Veit-Haibach P, Allgayer B. Importance of SPECT/CT for knee and hip joint Prostheses. Radiologe. 2012;52(7):629–37.

    Article  CAS  PubMed  Google Scholar 

  25. Chew CG, Lewis P, Middleton F, van den Wijngaard R, Deshaies A. Radionuclide arthrogram with SPECT/CT for the evaluation of mechanical loosening of hip and knee prostheses. Ann Nucl Med. 2010;24(10):735–43.

    Article  PubMed  Google Scholar 

  26. Tam HH, Bhaludin B, Rahman F, Weller A, Ejindu V, Parthipun A. SPECT-CT in total hip arthroplasty. Clin Radiol. 2014;69(1):82–95.

    Article  CAS  PubMed  Google Scholar 

  27. Berber R, Henckel J, Khoo M, Wan S, Hua J, Skinner J, Hart A. Clinical Usefulness of SPECT-CT in Patients with an Unexplained Pain in Metal on Metal (MOM) Total Hip Arthroplasty. J Arthroplasty. 2015;30(4):687–94.

    Article  PubMed  Google Scholar 

  28. Slevin O, Amsler F, Hirschmann MT. No correlation between coronal alignment of total knee arthroplasty and clinical outcomes: a prospective clinical study using 3D-CT. Knee Surg Sports Traumatol Arthrosc. 2016. doi: 10.1007/s00167-016-4400-y.

  29. Henckel J, Richards R, Lozhkin K, Harris S, Baena FM, Barrett AR, Cobb JP. Very low-dose computed tomography for planning and outcome measurement in knee replacement. The imperial knee protocol. J Bone Joint Surg (Br). 2006;88(11):1513–8.

    Article  CAS  Google Scholar 

  30. Gruen TA, McNeice GM, Amstutz HC. "Modes of failure" of cemented stem-type femoral components: a radiographic analysis of loosening. Clin Orthop Relat Res. 1979;141:17–27.

    Google Scholar 

  31. Walter SD, Eliasziw M, Donner A. Sample size and optimal designs for reliability studies. Stat Med. 1998;17(1):101–10.

    Article  CAS  PubMed  Google Scholar 

  32. Vaz S, Ferreira TC, Salgado L, Paycha F. Bone scan usefulness in patients with painful hip or knee prosthesis: 10 situations that can cause pain, other than loosening and infection. Eur J Orthop Surg Traumatol. 2016;27(2):147-56.

  33. Dobrindt O, Amthauer H, Krueger A, Ruf J, Wissel H, Grosser OS, Seidensticker M, Lohmann CH. Hybrid SPECT/CT for the assessment of a painful hip after uncemented total hip arthroplasty. BMC Med Imaging. 2015;15:18.

    Article  PubMed  PubMed Central  Google Scholar 

  34. Jin S, Boktor RR, Kc Man K, Pitman AG. Normal range of osteoblastic activity in total hip arthroplasties on single-photon emission computed tomography bone scintigraphy. Nucl Med Commun. 2016;37(9):924–34.

    Article  CAS  PubMed  Google Scholar 

Download references

Acknowledgements

Not applicable.

Funding

There has been no specific grant funding for this study.

Availability of data and materials

The datasets used and/or analysed during the current study available from the corresponding author on reasonable request.

Authors’ contributions

EB and FA set-up the study, collected the data and helped with drafting of the manuscript. HR and MTH designed the study and performed the data analysis. JH, PK and HR helped with data acquisition and drafting of manuscript. All authors read and approved the final manuscript.

Competing interests

The authors declare that they have no competing interests.

Consent for publication

Not applicable.

Ethics approval and consent to participate

'The study was approved by the local ethical committee (EKNZ 205/10). Written informed consent was obtained from all patients.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Michael T. Hirschmann.

Rights and permissions

Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Barthassat, E., Afifi, F., Konala, P. et al. Evaluation of patients with painful total hip arthroplasty using combined single photon emission tomography and conventional computerized tomography (SPECT/CT) – a comparison of semi-quantitative versus 3D volumetric quantitative measurements. BMC Med Imaging 17, 31 (2017). https://doi.org/10.1186/s12880-017-0204-x

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s12880-017-0204-x

Keywords