Mapping breast tissue types by miniature radio-frequency near-field spectroscopy sensor in ex-vivo freshly excised specimens

Abstract

Background

Receiving real-time information on tissue properties while performing biopsy procedures has the potential of improving biopsy accuracy. The study goal was to test the ability of a miniature flexible Radio-Frequency (RF) sensor (Dune Medical Devices), designed to be mounted on the surface of surgical tools, in measuring and mapping the various breast tissue types and abnormalities in terms of electrical properties.

Methods

Between January and October 2012, 102 patients undergoing lumpectomy, open-biopsy or mastectomy, in 3 medical centers, were enrolled in this study. The device was applied to freshly excised specimens, with registration between device measurements and histology analysis. Based on histology, the dielectric properties of the various tissue types were derived. Additionally, the ability of the device to differentiate between malignant and non-malignant tissue was assessed.

Results

A total of 4322 measurements from 106 specimens from 102 patients were analyzed. The dielectric properties of 10 tissue types in the low RF-frequency range were measured, showing distinct differences between the various types. Based on the dielectric properties, a score variable was derived, which showed a correlation of 90 % between the RF measurements and the tissue types. Differentiation ability between tissue types was characterized using ROC curve analysis, with AUC of 0.96, and sensitivity and specificity of 90 and 91 % respectively, for tissue feature sizes at or above 0.8 mm.

Conclusions

Using a radio-frequency near-field spectroscopy miniature flexible sensor the dielectric properties of multiple breast tissue types, both normal and abnormal, were evaluated. The results show promise in differentiating between various breast tissue types, and specifically for differentiation between cancer and normal tissues.

Peer Review reports

Background

The response of matter to electromagnetic fields is characterized by the material’s dielectric properties: permittivity and conductivity. The conductivity is a measure of the ease with which free electric charges can migrate through the material; the permittivity reflects the extent to which bound charge distributions can be distorted through polarization by an external field.

Differences in dielectric properties between human tissues and specifically between benign and malignant tissue have been studied and are well established [13]. Tissue dielectric properties are determined by concentration and mobility of intra and extracellular components, cell size, structure and arrangement, amongst other characteristics. Specifically, the tissue dielectric properties have been extensively studied in breast tissue [47], where differences between tissue types, and specifically between normal and malignant were observed over a broad range of frequencies. These properties have been successfully used to differentiate between normal and malignant breast tissue [8, 9] and for intraoperative margin assessment during lumpectomies [10]. Normal breast tissue is heterogeneous, being composed of three different types of tissue (Adipose, Glandular, and Connective), In all the studies to date, the evaluated dielectric properties were based on measurements and comparisons performed on scales larger than the intrinsic scale of tissue heterogeneity within breast tissue. Therefore, the differences observed represent average values of the dielectric properties. Results were generally reported on properties of “normal” and “cancer” types. In some reports [5, 7] an attempt has been made to partition between the three intrinsic tissue types of the normal breast. There has been no reporting on the specific properties of glandular and connective tissue types. Also, cancer of the breast has three major types: Ductal Carcinoma in Situ (DCIS), Invasive Ductal Cancer (IDC), and Invasive Lobular Cancer (ILC). The dielectric properties of these types, to date, have not been characterized separately. Additionally, specifically of importance in breast biopsy procedures, the dielectric characteristics of abnormalities in the breast that may progress to cancer, such as: non-malignant proliferative, non-malignant proliferative with atypia, and LCIS, have not yet been characterized.

The burden of breast cancer is high. Approximately 230,480 American women are diagnosed annually, and 39,520 women die from this disease [11]. Global cancer statistics show that breast cancer is the most frequently diagnosed cancer and the leading cause of cancer death among females, accounting for 23 % of total cancer cases and 14 % of cancer deaths [12]. The majority of breast cancers are diagnosed as a result of an abnormal mammogram or ultrasound, and in selected populations abnormal MRI findings. Some lesions are found by the patient or her physician as a palpable mass. Not all abnormal findings diagnosed by the methods mentioned represent cancer. To determine whether a mass in the breast is a suspicious mass or not the BI-RADS System was developed [13]. All patients with a BI-RADS category of 4 and 5 should undergo a biopsy. Those with category 3 should be followed more frequently. A clinically suspicious mass should also be biopsied, regardless of imaging findings, as about 15 % of such lesions can be mammographically occult [14].

Screening mammography is the most common way to diagnose early breast cancer but carries a high rate of recalls (16.3 % at first and 9.3 % at subsequent mammography) [15]. Biopsy is further recommended in 0.6–1 % of all screened women [16, 17]. Millions of women are screened each year, therefore these figures represent a high number of breast biopsies performed each year, emphasizing the need for accurate biopsy.

During the breast biopsy procedure part or all of a suspicious breast tissue growth is sampled and examined for the presence of cancer, most often in a minimally invasive procedure. Current biopsy techniques have several limitations: First, patients diagnosed with Atypical Ductal Hyperplasia (ADH) are routinely sent for an open surgical biopsy, following which 10–25 % of these patients’ diagnosis will be ‘upgraded’ DCIS, which requires a further lumpectomy or mastectomy [1821]. Second, patients with a diagnosis of DCIS in biopsy will undergo lumpectomy, typically without a sentinel lymph node biopsy. Following the Lumpectomy, about 20 % of these patients’ diagnosis will be upgraded to invasive cancer, requiring a further surgery for node biopsy. Third, studies have shown that up to 10 % of patients endure repeat biopsies [22, 23]. These repeat biopsies reveal carcinoma in up to 25 % of cases [24]. Forth, published data show a 1–7 % false negative rate with current breast core biopsy techniques [25].

The inaccuracies in the biopsy procedure result mainly from the uncertainty in the exact location from which the biopsy sample is taken relative to the image, and from the fact that the features presented on imaging may not be the most abnormal tissue present. Having an in-situ, at the needle tip, tissue characterization ability when performing biopsy procedures has the potential to increase the accuracy of the procedure.

In the presented study, we use a miniature, 0.8 mm in diameter, RF sensor to characterize the dielectric properties of different breast tissue types and abnormities. Based on these characteristics, we tested the potential of this type of sensor in differentiating between normal, abnormal, and malignant breast tissue. The sensor has a coaxial opening that results in a fringing electrical field close to the sensor surface. The sensor is manufactured using flexible printed circuit board technology and can be potentially placed on various devices having different geometries, such as open surgical tools as well as minimally invasive ones, like core biopsy needles. A similar device has been already been used for evaluating freshly excised radical prostatectomy specimens [26].

Material and methods

Ethics, consent and permissions

The study was performed at 3 sites under institutional review board approval and in accordance with the Declaration of Helsinki. All subjects signed an informed consent.

General design

Between January to October 2012, subjects undergoing lumpectomy, open biopsy or mastectomy procedure were enrolled. Inclusion criteria specified subjects over 18 years of age.

Tissue measurements were performed on freshly excised breast specimens. Measurements were compared to histological analysis. All medical staff members were blinded to device output. Specimen handling before and after measurements was performed according to routine procedures. As the device was used on ex-vivo specimens in the pathology lab, safety aspects (adverse effects) of the study and device use were monitored only with regards to specimen handling and analysis.

Device description

The device (Dune Medical Devices Ltd., Israel) used in this study is a near-field Radio-Frequency (RF) spectroscopy-based real-time detection device. It consists of a hand-held, pencil like, probe, connected by cables to a console. The console sends RF waves at several frequencies, transmitted to the tissue through the sensor at the probe’s tip. The frequencies (4–30 MHz) were chosen based on design considerations, and on where differences between breast tissue types are expected to be substantial [5, 6]. The frequency range at which the tissue is probed has no ionizing effects. The amplitude of the transmitted RF field across the sensor opening is low. The voltage across the sensor is ~ 0.1 Volts, much lower than the ionization voltage. The transmitted RF power is ~ 0.1 mW per square mm. Each measurement takes ~ 3 msec. The RF signals are reflected from the tissue through the sensor, and are received by the console. The sensor is designed as a multilayer transmission line structure with a coaxial end opening. It is fabricated from flexible circuit board materials, making the sensor (including the transmission line structure) mechanically flexible and thin. The sensor coaxial end conductive regions are gold–plated. The diameter of the sensor coaxial end is 0.8 mm. The sensor end is attached to the flat tip of the probe. The size of the coaxial sensor is much smaller than the transmitted wavelengths, as the highest frequency of 30Mhz corresponds to a wavelength of 10 m. Under this condition the penetration depth of the field is ~0.078 of the coaxial opening [27]. This translates to a penetration depth of ~0.07 mm.

The sensor has a coaxial opening that results in a fringing electrical field close to the sensors surface, which enables direct calculation of the impedance (dielectric properties - permittivity and conductivity) of the tissue in contact with the sensor. Signals at specific frequencies are transmitted to the sensor. The fringing field interacts with the tissue so that the reflected signals are dependent on the impedance of the tissue close to the sensor. From the reflection amplitude and phase the impedance of the tissue at each frequency measured is extracted using the process described in [28].

Tissue measurements and processing

Following excision, the fresh specimens were directly delivered to the pathology laboratory; the fresh, non-fixed, specimens were then fully inked and sliced into 5–7 transverse sections, approximately 1 cm thick, in a bread-loaf manner.

One or two transverse-cut sections were immediately sampled by the probe. The time between performing the measurements and tissue excision was no longer than 30 min. Specially designed stencils were placed on the slice. These stencils contained a matrix of holes of diameter 3.2 mm, which accommodated the probe’s distal end, enabling it to contact the tissue surface (Fig. 1). The grid spacing was 6 mm. The probe was operated manually to measure all measurement sites of the slice through the stencil (Fig. 1). Good contact between the sensor and tissue was achieved by applying mechanical pressure by hand to the tissue in the stencil holes. Each individual measurement took 1–3 s to complete. Designated software utility helped the user to record the location of the measurements within the stencil. Once a slice was fully measured it was fixed (24–48 h) and further processed for histological analysis. The histological analysis of the slices was performed en-face, without any further sectioning, so that the full surface of the measurement locations was available for histological examination. The relative position of the stencil and the slice was maintained, thus the sampling locations remain fully registered. The Histological slides were prepared for each measured slice. As the diameter of the stencil holes is 3.2 mm and the probe tip diameter is 3.0 mm, it is estimated that the potential registration error between the actual measured site and the analyzed tissue sample was less than 0.1 mm.

Fig. 1
figure 1

Measurement method of the tissue slices with and apparatus and stencil probe. Inset: histology slide of four measurement sites with their respective stencil coordinates

Data analysis

Tissue

Each measurement site was histologically analyzed. The tissue composition and feature sizes of the various tissue types present were recorded. The sites were then classified according to their most advanced abnormality within the 1 mm central area and divided into 6 categories [29]: 1) normal and non-malignant non-proliferative abnormal tissue; 2) non-malignant proliferative abnormal tissue; 3) non-malignant proliferative tissue with atypia; 4) LCIS; 5) DCIS and 6) invasive cancer.

Measurements

The permittivity and conductivity of each measured tissue site were extracted at all measured frequencies.

Dielectric properties

A subset of measurements in which the size of a given tissue type, the feature size, was larger than 0.7 mm was used for assessment of dielectric properties of various tissue types. As the sensor’s size is 0.8 mm, selection of such feature sizes ensures that the sensor came in contact with the specific tissue type being assessed. Analysis was performed by linear regression [30]. Only tissue types presenting in >5 samples were considered. This provided an estimation error of no more than 50 %.

Score variable

The same subset of measurements used to determine the dielectric properties of tissues was used in order to derive a single score variable. The score variable was calculated as a linear combination (using linear discriminant analysis) [31] of the dielectric properties, a linear combination that exhibited differentiation between two selected categories, e.g. normal and malignant tissue. The discriminant analysis was performed for each measurement frequency separately. The analysis presented is based on the results for 10 Mhz. The correlation of the score variable values with tissue type categories was characterized by a linear regression analysis.

ROC curve analysis

A dichotomous device output (positive/negative) for each measurement was derived by placing a threshold for the score variable value. Measurements with a score value lower than the threshold were defined as being negative, and those above the threshold were defined as being positive. Device indication (positive/negative) of qualified measurements was compared to tissue histology. The threshold levels of the score variable were scanned to generate receiver operating characteristics (ROC) curves. Sensitivity and specificity at the optimal cutoff points (points on the ROC curve closest to the upper left corner of the axes) were extracted. This analysis was repeated for normal vs. all abnormal types (malignant and non-malignant). The score variable of normal vs. malignant tissues was also used to estimate the differentiation ability as a function of the malignant tissue feature size.

Results

Out of 104 enrolled subjects, 102 were analyzed (their baseline characteristics are presented in Table 1). Two subjects were excluded from the analysis: one due to neo-adjuvant treatment and one specimen was inserted into formaldehyde prior to device use. Four subjects had bilateral surgery, thus 106 specimens were included in the analysis. Device measurements were found to be non-destructive and had no effect on the specimens or on the ability to inspect them histopathologically.

Table 1 Patient Cohort

In the 106 specimens included in the analysis 5262 measurements were performed. Altogether 940 measurements (18 %) were not analyzed. 388 sites were not reproduced during slide preparation, i.e. no histology to compare to measurement, and in 119 sites the exact registration between measurement site and histology could not be verified. The remaining 433 sites were disqualified due to predefined criteria related to poor quality of the histology slides (fragmented tissue, torn tissue, uneven tissue surface, color covering sample). Altogether, 4322 measurements were analyzed. The distribution of tissue histology of all qualified tissue measurements is shown in Table 2. Note that the numbers present the most advanced abnormality, according to the established 6 category scale described earlier, within each tissue site analyzed. This was recorded irrespective of size of the abnormality within the sensed area. For example, in some of the sites categorized as malignant some non-malignant abnormalities may be present, as well.

Table 2 Distribution of tissue histology

Dielectric properties

The dielectric properties (conductivity, y-axis, and permittivity, x-axis) of 10 tissue types, as assessed from the data at 4, 10 and 30 MHz, are presented in Fig. 2. The assessment was based on a subset of 1253 tissue measurements which had feature sizes above 0.7 mm of a single tissue type. The dielectric properties were calculated using a linear regression with respect to the relative tissue compositions of these samples. Only tissue types that were present within at least 5 samples were analyzed. The estimation error for each tissue type is also presented. The error depends on the actual spread of values as measured, the feature sizes, and the number of the samples at which a given tissue type is present (e.g. the error for ADH is larger as the number of samples containing ADH was only 7). Additional tissue types present in the breast were either not present in the subset or were very rare, rendering the assessment impractical. The results are similar across the frequency range measured, but the distinction between the various breast tissue types is clearer at 4 and 10 MHz.

Fig. 2
figure 2

Dielectric properties, evaluated from measurements performed at 4, 10 and 30 MHz (top to bottom), of various tissue types

Correlation between tissue types and score variable

Figure 3 presents the average (and standard deviation) of the score variable (y-axis), derived based on the tissue dielectric properties, vs. the tissue type categories (x-axis). Although tissue were grouped into 6 categories, in the data used (feature size >0.7 mm) for this analysis only 4 categories were present (after excluding Fibroadenoma). LCIS and proliferative lesions were not present in these samples. The more severe the tissue abnormality, progressing from left to right, the higher the score variable, starting at -0.4 (+/- 1.2) for normal and non-proliferative abnormal tissue, up to 2.6 (+/- 1.4) for invasive cancer. The linear regression coefficient between the score variable and the tissue type grouping for the dataset was 0.59 [95 % CI: 0.11–1.07, P = 0.033]. The need to identify an abnormality such as Fibroadenoma is usually in young women, where these are the majority of the biopsy findings. In these young women the prevalence of cancer (or atypia) is very low. In women who are biopsied with more “suspicion” for cancer, the more relevant differentiation required does not include Fibroademona. Therefore, Fibroademona presents an “isolated” tissue type, with the background tissue being, most always, normal breast tissue.

Fig. 3
figure 3

Correlation of score variable (blue diamond denotes the average value and the bars the standard deviation) and tissue type divided into 6 categories

Differentiation between tissue categories

It is clear from Fig. 2 that various tissue types have different dielectric properties. To further assess whether these differences can serve as a basis for differentiation between various tissue categories, the subset of measurements with feature sizes of >0.7 mm was grouped into binary categories. The resulting ROC curves for two interesting cases, normal vs. malignant (blue line) and normal vs. all abnormal types (red line), are presented in Fig. 4. In both cases the curves follow the left and upper axes, reaching very close to the upper left corner, with the areas under the curve approaching unity; 0.95 (95 % CI: 0.94–0.96) and 0.96 (95 % CI: 0.94–0.97), respectively. The sensitivity and specificity evaluated for each curve at the optimal cut-off point (the point in the curve closest to the upper left corner) are: Sensitivity 90.5 % (95 % CI: 86–94) and Specificity 90.1 (95 % CI: 88–92) for differentiating between normal and malignant tissue, and Sensitivity 88.6 % (95 % CI: 85–92) and Specificity 91.7 (95 % CI: 90–93) for differentiating between normal and all abnormal tissue types, both malignant and non-malignant.

Fig. 4
figure 4

ROC curve analysis of normal vs. malignant measurement sites (blue) and normal vs. abnormal measurement sites (red) in the subset with feature sizes of >0.7 mm

In general, it is expected during a biopsy procedure to encounter lesions of ~1 mm or larger. From analysis of SEER (Surveillance, Epidemiology, and End Results program of the National Cancer Institute) data, more than 97 % of the malignant lesions are larger than 3 mm [32]. Still, it is interesting to analyze the separation ability of the device between normal and malignant tissues for various tissue feature sizes. Figure 5 presents 3 ROC curves for 3 feature size categories - above 0.8 mm (the size of the sensor), 0.5–0.8 mm and below 0.5 mm (including any detectable cancer by pathology, even below 0.1 mm). As seen in the figure, the ability to detect cancer (quantified by the areas under the curves, also presented in Table 3) depends, as expected, on the feature size. Table 3 also presents the sensitivity and specificity for each case at the optimal cutoff point of the score variable.

Fig. 5
figure 5

ROC curve analysis of normal vs. malignant measurement sites as function of cancer feature size, for full dataset

Table 3 Differentiation ability for various cancer feature sizes

Discussion

Our results quantify dielectric properties of 10 different tissue types present within the breast, and show that they all have different dielectric properties. These includes the types related to normal breast tissue (Adipose, Glandular, and Connective), and types associated with various abnormalities, including the differentiation between the dielectric properties of the 3 types of cancer types: IDC, ILC, and DCIS. Prior studies [47] have demonstrated that, generally, there are differences between normal and cancer tissue within the breast, but have not provided the level of differentiation presented in this work. Additionally, the data on the dielectric properties of pre-malignant types is new. These identified differences served as a basis for constructing a score variable that demonstrated correlation with the degree of abnormality, including pre-malignant phase, of the breast tissue. Additionally, using the dielectric properties and the score variable good differentiation between normal and malignant, or non-malignant abnormal, tissue was established. The configuration of the sensor as a 0.8 mm circular sensor can be thought of as a basic sensing unit for use in breast biopsy procedures, with an array of sensors aligned along the biopsy needle.

The goal of the initial biopsy is to obtain sufficient diagnostic material using the least invasive approach and to avoid surgical excision of benign lesions. CNB offers a definitive histologic diagnosis, avoids inadequate samples and may permit the distinction between invasive versus in situ cancer. In most centers, image guided CNB has replaced wire localization and surgical excision as the most common initial biopsy method for nonpalpable abnormalities [3335]. The accuracy of CNB was shown in a series of 952 consecutive breast CNBs performed at one institution (342 without image guidance, 241 with ultrasound guidance, and 369 using a stereotactic vacuum assisted biopsy (VAB)) [36]. The false-negative rate with 11-gauge VAB was 3 %, compared to 13, 5, and 22 % for non-image guided, surgeon-performed ultrasound-guided, and 14-gauge VAB, respectively. In most of these false negative patients (5–22 %) the reason for the false negative result was due to sampling error, meaning that the biopsy was not taken from the lesion as planned.

The false negative rate can be reduced by using very large needles [37]. An alternative approach could be to keep the needle size relatively small, but add local sensors located at the needle’s tissue collection region that provide in-situ information on the tissue about to be sampled. The basic units for these sensors can be sub-millimeter, circular near-field radio-frequency sensing units, as those we have used in this work for characterizing breast tissue properties. These sensors may provide real-time measurements of tissue dielectric properties at the locations of tissue to be biopsied. As the power transmitted by the sensor is very low and the RF radiation is non-ionizing, these type of sensors are well suited for in-vivo use.

As per the current standard of care, imaging will be used to direct the needle to the general location of the suspected abnormality. The in-site sensors will provide indication of the tissue type at the immediate vicinity of the needle and in contact with the sensor, as the sensor is effectively a surface characterization sensor. The penetration depth the 0.8 mm sensor in no more than 0.1 mm. By scanning/moving the needle around the suspected region, the most suspicious tissue abnormality can be identified, and biopsied. The spatial resolution of a sensing device designed using the sensors as the basic building blocks is dictated by the sensor size and by the ability to arrange sensors close together. Arranging 0.8 mm sensors in an array will preserve this resolution, as the sensors can share the same ground plane (the outer conductor of the coaxial aperture).

For a potential set-up for use in a biopsy device configuration, approximately 10 0.8 mm circular sensors will be arranged in a 1D array on the biopsy needle, in a location overlapping the biopsy sampling cavity. As each measurement takes approximately 2.5 msec, a reading from all 10 sensors will take ~ 0.025 s. The full measurement cycle, including displaying the results to the user, will take approximately 0.2 s, thus providing real-time tissue characterization as the needle is progressed through the tissue. Therefore, when the sensors will be integrated with the biopsy needle, it is anticipated that the duration of the biopsy procedure will not be extended.

DCIS is presented mostly as microcalcifications on mammography and diagnosed by stereotactic mammography guided biopsy (mammotomy). The ability of the sensor to distinguish between DCIS and normal breast tissue elements seems promising for using this type of sensor also in mammotomies.

Patients with ADH that were diagnosed on a CNB will be found to have in up to 25 % DCIS or even IDC present at open biopsy. Therefore patients with ADH on CNB are routinely sent for an open surgical biopsy [1821]. A more accurate CNB can reduce the number of unnecessary open biopsies in these patients. The ability of the sensor to differentiate between normal breast tissue, ADH and DCIS, can potentially improve the accuracy of CNB and reduce the number of open biopsies.

The diagnostic capability of the sensor for differentiating between malignant and normal tissue is high, with a Sensitivity of 90.5 % and Specificity of 90.1 %. This is also reflected by the ROC curves, with an area under the curve of 0.95. The ability of the sensor to differentiate any abnormal tissue from normal tissue is also high, Sensitivity 88.6 % and Specificity 91.7 %. With the current false-negative rates of CNB, this level of sensitivity has the potential of reducing the false negative rates in biopsy procedures to below 1 %.

The detection sensitivity of the sensor is dependent on the feature size of the malignant tissue. A 90.2 % Sensitivity features of at least 0.8 m in size, down to 66.8 % sensitivity for features smaller than 0.5 mm. It is anticipated that most biopsied malignancies will be of at least 1 mm in size, as, based on final pathology, most all of the malignant lesions are found to be larger than 3 mm.

There are some limitations to this type of sensor design. The manufacturing process of the sensor has to account for potential chemical modifications (over time) of the senor face, which can affect the reflected signals. Arranging sensors in a tightly packed array provides an additional challenge in isolating the electrical response of the individual sensors from their neighbors. The tissue has to be in direct contact with the sensor. Therefore, a practical device will need good attachment to substrate, as any (air) voids between the sensor and the tissue will skew the impedance measurement results. The very low penetration depth may limit the scope of applications in which this type of sensor will provide benefit. The sensor is gold-plated, with gold known not to be durable with regards to mechanical abrasion. In a biopsy procedure each sensor would be used for a limited duration, typically no more than a few minutes. Also, breast tissue is soft, and therefore it is not expected that the sensor structure will be mechanically effected during use.

Conclusion

By use of a miniature flexible radio-frequency near-field spectroscopy sensor the dielectric properties of 10 tissue types present within the breast were quantified, showing distinct differences between the all these types. Using these differences a good differentiating was achieved between breast tissue states, specifically between cancerous and normal tissue. The sensor’s dimensions and design may enable the use of such sensors in minimally invasive procedures, including breast biopsy.

Abbreviations

ADH:

Atypical Ductal Hyperplasia

CNB:

Core Needle Biopsy

DCIS:

Ductal Carcinoma In Situ

RF:

Radio Frequency

ROC:

Receiver Operating Characteristics

VAB:

Vacuum assisted Biopsy

References

  1. Gabriel S, Lau RW, Gabriel C. The dielectric properties of biological tissue II: Measurements in the frequency range 10 Hz to 20 GHz. Phys Med Biol. 1996;41:2251.

    Article  CAS  PubMed  Google Scholar 

  2. Foster KR, Schepps JL. Dielectric properties of tumor and normal tissues at radio through microwave frequencies. J Micro Power. 1981;16(2):107–19.

    Article  CAS  Google Scholar 

  3. Joines WT, Zhang Y, Li C, Jirtle RL. The measured electrical properties of normal and malignant human tissues from 50 to 900 MHz. Med Phys. 1994;21(4):547–50.

    Article  CAS  PubMed  Google Scholar 

  4. Sha L, Ward ER, Stroy B. A review of dielectric properties of normal and malignant breast tissue. PIEEE Southeast Con. 2002; 457-462.

  5. Surowiec AJ, Stuchly SS, Barr JR, Swarup A. Dielectric properties of breast carcinoma and the surrounding tissue. IEEE Trans Biomed Eng. 1988;35(4):257–63.

    Article  CAS  PubMed  Google Scholar 

  6. Zou Y, Guo Z. A review of electrical impedance techniques for breast cancer. Med Eng Phys. 2003;25(2):79–90.

    Article  CAS  PubMed  Google Scholar 

  7. Lazebnik M, et al. A large-scale study of the ultrawideband microwave dielectric properties of normal breast tissue obtained from reduction surgeries. Phys Med Biol. 2007;52:6093–115.

    Article  PubMed  Google Scholar 

  8. Pappo I, Spector R, Schindel A, Morgenstern S, Sandbank J, Leider LT, et al. Diagnostic performance of a novel device for real-time margin assessment in lumpectomy specimens. J Surg Res. 2010;160(2):277–81.

    Article  PubMed  Google Scholar 

  9. Karni T, Pappo I, Sandbank J, Lavon O, Kent V, Spector R, et al. A device for real-time, intraoperative margin assessment in breast-conservation surgery. Am J Surg. 2007;194:467–73.

    Article  PubMed  Google Scholar 

  10. Allweis TM, Kaufman Z, Lelcuk S, Pappo I, Karni T, Schneebaum S, et al. A prospective, randomized, controlled, multicenter study of a real-time, intraoperative probe for positive margin detection in breast-conserving surgery. Am J Surg. 2008;196:483.

    Article  PubMed  Google Scholar 

  11. Siegel R, Ward E, Brawley O, Jemal A. Cancer statistics, 2011: The impact of eliminating socioeconomic and racial disparities on premature cancer deaths. CA Cancer J Clin. 2011;61:212.

    Article  PubMed  Google Scholar 

  12. Jemal A, Bray F, Center MM, et al. Global cancer statistics. CA Cancer J Clin. 2011;61:69.

    Article  PubMed  Google Scholar 

  13. Breast Imaging Reporting and Data System (BI-RADS) Atlas, 4th ed, American College of Radiology, Reston, VA 2003.

  14. Barlow WE, Lehman CD, Zheng Y, et al. Performance of diagnostic mammography for women with signs or symptoms of breast cancer. J Natl Cancer Inst. 2002;94:1.

    Article  Google Scholar 

  15. Hubbard RA, Kerlikowske K, Flowers CI, et al. Cumulative probability of false positive recall or biopsy recommendation after 10 years of screening mammography: a cohort study. Ann Intern Med. 2011;155:481–92.

    Article  PubMed  PubMed Central  Google Scholar 

  16. Ghosh K, Melton 3rd LJ, Suman VJ, Grant CS, Sterioff S, et al. Breastbiopsy utilization: a population-based study. Arch Intern Med. 2005;165:1593–8.

    Article  PubMed  Google Scholar 

  17. Nelson HD, et al. Screening for breast cancer: an update for the U.S. Preventive services task force. Ann Intern Med. 2009;151:727–37.

    Article  PubMed  PubMed Central  Google Scholar 

  18. Chae BJ, Lee A, Song BJ, Jung SS. Predictive factors for breast cancer in patients diagnosed atypical ductal hyperplasia at core needle biopsy. World J Surg Oncol. 2009;7:77.

    Article  PubMed  PubMed Central  Google Scholar 

  19. Elsheikh TM, Silverman JF. Follow-up surgical excision is indicated when breast core needle biopsies show atypical lobular hyperplasia or lobular carcinoma in situ: a correlative study of 33 patients with review of the literature. Am J Surg Pathol. 2005;29:534.

    Article  PubMed  Google Scholar 

  20. Anderson BO, Calhoun KE, Rosen EL. Evolving concepts in the management of lobular neoplasia. J Natl Compr Canc Netw. 2006;4:511.

    PubMed  Google Scholar 

  21. Margenthaler JA, Duke D, Monsees BS, et al. Correlation between core biopsy and excisional biopsy in breast high-risk lesions. Am J Surg. 2006;192:534.

    Article  PubMed  Google Scholar 

  22. Lui, et al. Rebiopsy after stereotactic core-needle breast biopsy: prospective study. J HK Coll Radiol. 2004;7:116–20.

    Google Scholar 

  23. Youk JH, et al. Missed breast cancers at US-guided core needle biopsy: How to reduce them, radio graphics, vol. 27. 2007. p. 79–94.

    Google Scholar 

  24. Brun del Re R. Minimally invasive breast biopsies, recent results in cancer research 173. Berlin Heidelberg: Springer; 2009. p. 149.

    Google Scholar 

  25. Fajardo, et al. Stereotactic and sonographic large-core biopsy of nonpalpable breast lesions: results of the radiologic diagnostic oncology group V study. Acad Radiol. 2004;11(3):293–308.

    Article  PubMed  Google Scholar 

  26. Dotan ZA, Fridman E, Lindner A, Ramon J, Pode D, Bejar J, Kopolovic J, Pizov G, Sandbank J, Katz R, Shapiro A, Shilo Y, Nativ O. Detection of prostate cancer by radio-frequency near-field spectroscopy in radical prostatectomy ex vivo specimens. Prostate Cancer Prostatic Dis. 2013;16(1):73–8.

    Article  CAS  PubMed  Google Scholar 

  27. Meaney PM, Williams BB, Geimer SD, Flood AB, Swartz HM. A coaxial dielectric probe technique for distinguishing tooth enamel from dental resin. Adv Biomed Eng Res. 2015;3:8–17.

    Article  PubMed  PubMed Central  Google Scholar 

  28. Misra DK. A quasi-static analysis of open-ended coaxial lines. IEEE Trans Microwave Theory Tech. 1987;MTT-35:925–8.

    Article  Google Scholar 

  29. Dupont WD, Page DL. Risk factors for breast cancer in women with proliferative breast disease. N Engl J Med. 1985;312:146–51. American Cancer Society, Non-cancerous Breast Conditions.

    Article  CAS  PubMed  Google Scholar 

  30. Rawlings JO, Pantula SG, Dickey DA. Applied regression analysis: a research tool. Springer; 1998.

  31. Estrela da Silva JE, Marques de Sa JP, Jossinet J. Classification of breast tissue by electrical impedance spectroscopy. J Med Biol Eng Comput. 2000;38:26–30.

    Article  Google Scholar 

  32. Ries LAG, Young JL, Keel GE, Eisner MP, Lin YD, Horner M-J, editors. SEER survival monograph: cancer survival among adults: U.S. SEER program, 1988-2001, patient and tumor characteristics. Bethesda: National Cancer Institute; 2007. SEER Program, NIH Pub. No. 07-6215.

    Google Scholar 

  33. Gutwein LG, Ang DN, Liu H, et al. Utilization of minimally invasive breast biopsy for the evaluation of suspicious breast lesions. Am J Surg. 2011;202:127.

    Article  PubMed  Google Scholar 

  34. Verkooijen HM, Core Biopsy After Radiological Localisation (COBRA) Study Group. Diagnostic accuracy of stereotactic large-core needle biopsy for nonpalpable breast disease: results of a multicenter prospective study with 95 % surgical confirmation. Int J Cancer. 2002;99:853.

    Article  CAS  PubMed  Google Scholar 

  35. Stomper PC, Winston PS, Proulx GM, et al. Mammographic detection and staging of ductal carcinoma in situ: mammographic-pathologic correlation. Semin Breast Dis. 2000;3:1.

    Google Scholar 

  36. Shah VI, Raju U, Chitale D, et al. False-negative core needle biopsies of the breast: an analysis of clinical, radiologic, and pathologic findings in 27 consecutive cases of missed breast cancer. Cancer. 2003;97:1824.

    Article  PubMed  Google Scholar 

  37. Jackman RJ, Marzoni Jr FA, Rosenberg J. False-negative diagnoses at stereotactic vacuum-assisted needle breast biopsy: long-term follow-up of 1,280 lesions and review of the literature. AJR Am J Roentgenol. 2009;192:341.

    Article  PubMed  Google Scholar 

Download references

Acknowledgements

None.

Funding

The study was sponsored by Dune Medical devices.

Availability of data and materials

The data supporting the conclusions of this article is available upon request to the corresponding author.

Authors’ contributions

ZK, HP, IH, PM, TK, IP, and TA participated in the design of the study. TZ, JS, and JD performed the pathological evaluation. ZK and TE drafted the manuscript. All authors read and approved the final manuscript.

Competing interests

The authors declare that they have no competing interests.

Consent for publication

Not applicable.

Ethics approval and consent to participate

The study was performed at 3 sites under institutional review board approval and in accordance with the Declaration of Helsinki. The IRBs that granted study approval are: (1) Meir Hospital’s Helsinki committee (2) Assaf Harofeh Medical Center’s Helsinki Committee and (3) Kaplan Medical Center’s Helsinki Committee. All subjects signed an informed consent form.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Zvi Kaufman.

Rights and permissions

Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Kaufman, Z., Paran, H., Haas, I. et al. Mapping breast tissue types by miniature radio-frequency near-field spectroscopy sensor in ex-vivo freshly excised specimens. BMC Med Imaging 16, 57 (2016). https://doi.org/10.1186/s12880-016-0160-x

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s12880-016-0160-x

Keywords