Semi-automatic analysis of standard uptake values in serial PET/CT studies in patients with lung cancer and lymphoma

Abstract

Background

Changes in maximum standardised uptake values (SUVmax) between serial PET/CT studies are used to determine disease progression or regression in oncologic patients. To measure these changes manually can be time consuming in a clinical routine. A semi-automatic method for calculation of SUVmax in serial PET/CT studies was developed and compared to a conventional manual method. The semi-automatic method first aligns the serial PET/CT studies based on the CT images. Thereafter, the reader selects an abnormal lesion in one of the PET studies. After this manual step, the program automatically detects the corresponding lesion in the other PET study, segments the two lesions and calculates the SUVmax in both studies as well as the difference between the SUVmax values. The results of the semi-automatic analysis were compared to that of a manual SUVmax analysis using a Philips PET/CT workstation. Three readers did the SUVmax readings in both methods. Sixteen patients with lung cancer or lymphoma who had undergone two PET/CT studies were included. There were a total of 26 lesions.

Results

Linear regression analysis of changes in SUVmax show that intercepts and slopes are close to the line of identity for all readers (reader 1: intercept = 1.02, R2 = 0.96; reader 2: intercept = 0.97, R2 = 0.98; reader 3: intercept = 0.99, R2 = 0.98). Manual and semi-automatic method agreed in all cases whether SUVmax had increased or decreased between the serial studies. The average time to measure SUVmax changes in two serial PET/CT examinations was four to five times longer for the manual method compared to the semi-automatic method for all readers (reader 1: 53.7 vs. 10.5 s; reader 2: 27.3 vs. 6.9 s; reader 3: 47.5 vs. 9.5 s; p < 0.001 for all).

Conclusions

Good agreement was shown in assessment of SUVmax changes between manual and semi-automatic method. The semi-automatic analysis was four to five times faster to perform than the manual analysis. These findings show the feasibility of using semi-automatic methods for calculation of SUVmax in clinical routine and encourage further development of programs using this type of methods.

Peer Review reports

Background

Changes in maximum standard uptake values (SUVmax) and computed tomography (CT) tumour size in 18FDG positron emission tomography (PET)/CT follow-up studies in oncologic patients are often assessed in tandem to determine disease progression or regression. RECIST criteria based on CT [1] are most frequently used, but criteria based on the uptake of 18FDG are receiving increasing attention [2, 3]. With this, the demand for quantitative monitoring of SUV has grown as the indications for performing follow-up 18FDG PET/CT are expanding [4]. SUVmax measurements are more readily reproducible than CT sizes in both pre- and post-treatment studies [5]. Currently available PET/CT software enables the use of tools that can determine SUVmax by defining a ROI/VOI both manually and automatically within one examination. To our knowledge, none of the PET/CT manufacturers offer a software suite that can measure two matching uptakes in serial examinations automatically.

There are several studies on tumour delineation and quantification in PET/CT [612]. New quantitative approaches to evaluate treatment response have been suggested in some papers [6, 11]. But differences in imaging parameters across PET centres have a negative effect on semi-automatic methods' performance [13, 14]. Due to the increasing amount of examinations and the quantification that can be done, standardization of PET image acquisition and effective computer-aided quantification is warranted.

A common clinical setting for PET/CT is the evaluation of follow-up examinations after treatment or expectancy. Manual measurement of SUVmax in single studies or changes in serial studies can be time consuming, making it difficult for clinicians and PET reviewers to follow-up all SUVmax changes in serial studies in a comprehensive and effective way. Studies with semi-automatic quantification of SUVmax changes in serial examinations suggest registration of images and SUV measurements are accurate and reliable [6, 11]. Time saving qualities, or investigation of the reproducibility of using a semi-automatic approach, compared to manual reading of SUVmax has not been elicited. The purpose of this study is to demonstrate these qualities using a semi-automatic method for calculation of SUVmax of an abnormal lesion in serial studies when a region of interest (ROI) has been semi-automatically defined in the PET image.

Materials and methods

Patients

Patients with lung cancer or lymphoma who had undergone two PET/CT studies between July 2008 and January 2010 at the Skåne University Hospital in Malmö, Sweden were included retrospectively.

There were ten lung cancer patients and six lymphoma patients. Diagnoses were verified with biopsy in all cases except for one patient on which the diagnosis was based on clinical findings. The patients had not been treated with chemotherapy or radiation therapy for at least 6 months prior to the baseline examination. The time from baseline examination to the follow-up examination varied between 31 and 150 days. Follow-up examinations were performed to evaluate tumour progression in five cases and therapy response in eleven cases.

Pathological lesions with sharp contrast to surrounding areas, no formation of a large conglomerate mass and presence in both studies were selected. 16 patients (69% men, age 56 ± 15 (mean ± SD)) with 26 pathological lesions were included. Each patient had one to four pathological lesions for manual and semi-automatic measurement. The study was approved by the Research Ethics Committee at Lund University.

Scan data

After 4-6 hours fast 4 MBq/kg of 18FDG was given intravenously. The effective dose was 5-7 mSv for PET, 2-3 mSv for low-dose CT and 10-15 for diagnostic CT. Data were acquired 60 min after injection and the patients were scanned from head to the upper thigh with the use of an integrated PET/CT system (Philips Gemini TF). A diagnostic/low dose CT scan obtained with the use of a standard protocol, 150 mAs/slice, 120 kV, a tube rotation time of 0.75 s per CT rotation, a pitch of 0.9, and a slice thickness of 5 mm preceded the PET scan covering the identical transverse field of view (a 2 min emission scan per table position and 6-10 bed positions per patient).

The PET image data sets were reconstructed iteratively with segmented correction for attenuation with use of the CT data. CT images, PET images and co-registered images were displayed by means of Philips workstation and program. Reconstructed PET and CT images were also transferred to a customized software developed for semi-automatic measurements.

SUVmax measurements

Three readers did manual and semi-automatic analysis of SUVmax on 26 pathological lesions. One of the readers marked the 26 lesions in screenshots of coronal and transaxial PET images and these images were used during the study to secure that all three readers were measuring the same lesions with the manual and the semi-automatic method. The time to measure SUVmax manually and semi-automatically, after complete loading of the serial studies, was recorded on all pathological lesions individually.

Manual method

Manual measurements were done on a Philips PET/CT workstation (Philips Extended Brilliance Workspace, PET/CT Application Suite v1.5 K). The readers identified the lesions in both PET studies and used their most preferred software tool to determine the SUVmax. Readers reported that they used the free-floating SUV search tool and ROIs.

Semi-automatic method

A method to measure SUVmax of a pathological lesion present in two studies from the same patient was developed. Transaxial PET and CT images from the two studies were used as input to the program. The two CT studies were automatically aligned by projecting the skeletons in the CT images on a 2D space. A rigid registration algorithm was used. CT images were reduced in size, blurred, thresholded and then overlaid to compare pixels with the formula: Σ(i, j)erf(i, j) = |IM1(i, j)-IM2(i, j)|

One of the images is repositioned and the formula is applied again to calculate the pixel error. The process is reiterated with less size reduction and blurring each time.

The reader marked an arbitrary pixel of a pathological lesion in one of the two PET studies. After this manual step the program automatically detects the corresponding lesion in the second PET study. The reader's mark in the first PET study has a corresponding location in the first CT study, provided that PET and CT images were correctly aligned from the PET/CT camera. The corresponding location in the second CT study was defined using the matching by the software and the corresponding location in the second PET study was defined.

The volumetric segmentation of the pathological lesions was made in the PET images. The reader's mark in the first PET study and the corresponding location in the second PET study were used as seed points for the segmentation using the graph-cut algorithm [15]. SUVmax in the entire lesion was then calculated automatically and the segmentations were presented to the reader. Semi-automatic analysis was done on a standard laptop computer (Figure 1).

Figure 1
figure 1

Flowchart of the semi-automatic process.

Statistical analyses

Intraobserver reproducibility was analysed with Intraclass correlation (ICC). A Bland-Altman analysis was used to assess the level of agreement between the two methods.

Results

The Bland-Altman analysis indicates that the 95% limits of agreement between manual and semi-automatic assessment ranged from -2.67 to 1.29 (reader 1); -3.39 to 2.29 (reader 2); -2.21 to 1.18 (reader 3). The two methods consistently provide similar measures for most of the tumours. Limits of agreement ranged from ± 0.99 to ± 1.41 across the readers. Manual and semi-automatic method agreed in all cases whether SUVmax had increased or decreased between the serial studies. The range of SUVmax values for the manual method was 2.4 - 19.8 and for the semi-automatic method 2.6 - 21.4.

ICCs of SUVmax readings between readers using the semi-automatic method in the first and second study were 1.00 (95% CI, 1.00-1.00) and 0.94 (95% CI, 0.89-0.97) respectively. ICC of SUVmax readings between readers using the manual method were 1.00 (95% CI, 1.00-1.00) and 0.95 (95% CI, 0.90-0.97) for the first and second study respectively. Almost-perfect reproducibility was thus obtained with both methods.

The average time to measure SUVmax changes in two serial PET/CT examinations was significantly longer for the manual method compared to the semi-automatic method for all readers (Table 1). The semi-automatic method is up to 4 - 5 times faster compared to the manual method (Figure 2).

Table 1 Average time to measure changes in SUVmax between two serial examinations for manual and semi-automatic methods and the difference between them
Figure 2
figure 2

Bland-Altman analysis of SUV max differences between both methods. Limits of agreement for A: -2.67 to 1.29 (reader 1); B: -3.39 to 2.29 (reader 2); C: -2.21 to 1.18 (reader 3)

Discussion

Good agreement between the two methods was showed with the Bland-Altman analysis. The somewhat wider range of limits of agreement in reader 2 (± 1.41) seems to be due to not finding the highest SUVmax with the manual method within some large tumours with high SUVmax. This resulted in a greater difference comparing to the semi-automatic method. Both methods agreed whether the SUVmax had increased or decreased in the follow-up examination. The results of this study indicate the feasibility of using semi-automatic method in serial examinations.

Like other studies [5, 6, 9, 10], the selection of tumours was limited to small-medium sized tumours, which were round, had a clear delineation on the CT scan and were not part of large conglomerates, which produce a "bleeding" effect on the PET image. The almost-perfect reproducibility obtained with both methods is likely dependent on the properties of the pathological sites. The finding supports the findings in previous studies [5, 6, 9, 10]. Clinical reality provides many obstacles for semi-automatic segmentation and it remains to be seen how pathological sites can be best segmented by algorithms no matter the form, size or location it has.

Readers were not instructed whether they should perform the semi-automatic segmentation in the first or second study. In the feedback after measuring with both methods, readers agreed it was easier to segment the correct tumour of both examinations by selecting the tumour on the examination, which had the smallest size. Should the readers have been instructed only to segment in the first or second scan regardless, the results may have shown poorer segmentation quality.

Segmentation in the PET image has more advantages than segmentation in CT. In the study it enabled segmentation of lymphoma in the liver and in locations where the lesion in CT was adjacent to soft tissue. Also, the segmentation tool used did not adequately segment the lesions in CT.

Manual delineation capabilities were intentionally left out of the program in order to test the accuracy of semi-automatic segmentation and in order to keep the time of analysis to a minimum.

SUVmax was chosen over SUVmean due to clinical praxis at our hospital when assessing PET/CT examinations. Furthermore, research has shown SUVmax measurements are readily reproducible between readers [5]. The program shows systematically higher SUVmax compared with the manual method. This may be due to different filter settings between the PET/CT manufacturer and the program. It is unlikely that all readers have systematically missed the "true" SUVmax with the program.

Conclusions

Good agreement was shown in absolute SUVmax measurements between both methods. Almost-perfect reproducibility was seen between three readers using both semi-automatic and manual methods. Using semi-automatic method reduces time to calculate SUVmax by up to 5 times. The findings show feasibility of using semi-automatic calculation of SUVmax in serial studies and encourage further development of programs that accurately segments more complex pathological sites.

Funding

This research is supported by grants from ALF (all authors).

References

  1. Eisenhauer EA, et al: New response evaluation criteria in solid tumours: revised RECIST guideline (version 1.1). Eur J Cancer. 2009, 45 (2): 228-247. 10.1016/j.ejca.2008.10.026.

    Article  CAS  PubMed  Google Scholar 

  2. Ben-Haim S, et al: 18F-FDG PET and PET/CT in the evaluation of cancer treatment response. J Nucl Med. 2009, 50 (1): 88-99. Review

    Article  PubMed  Google Scholar 

  3. Hicks RJ: Role of 18F-FDG PET in assessment of response in non-small cell lung cancer. J Nucl Med. 2009, 50 (Suppl 1): 31S-42S.

    Article  CAS  PubMed  Google Scholar 

  4. Blodgett TM, et al: PET/CT: Form and Function. Radiology. 2007, 242 (2): 360-385. 10.1148/radiol.2422051113. Review

    Article  PubMed  Google Scholar 

  5. Jacene HA, et al: Assessment of interobserver reproducibility in quantitative 18-F-FDG PET and CT measurements of tumor response to therapy. J Nucl Med. 2009, 50: 1760-1769. 10.2967/jnumed.109.063321.

    Article  PubMed  Google Scholar 

  6. Gao X, et al: Computer-assisted quantitative evaluation of therapeutic responses for lymphoma using serial PET/CT imaging. Acad Radiol. 2010, 17: 479-488. 10.1016/j.acra.2009.10.026.

    Article  PubMed  PubMed Central  Google Scholar 

  7. Tylski , et al: Comparative assessment of methods for estimating tumor volume and standardized uptake value in 18F-FDG PET. J Nucl Med February. 2010, 51 (2): 268-276. 10.2967/jnumed.109.066241.

    Article  Google Scholar 

  8. Wang , et al: Computer-assisted quantification of lung tumors in respiratory gated PET/CT images: phantom study. Med Biol Eng Comput. 2010, 48: 49-58. 10.1007/s11517-009-0549-6.

    Article  CAS  PubMed  Google Scholar 

  9. Yan J, et al: Marker-controlled watershed for lymphoma segmentation in sequential CT images. Med Phys. 2006, 33 (7): 2452-2460. 10.1118/1.2207133.

    Article  PubMed  Google Scholar 

  10. Yan J, et al: Automated matching and segmentation of lymphoma on serial CT examinations. Med Phys. 2007, 34 (1): 55-62. 10.1118/1.2404617.

    Article  PubMed  Google Scholar 

  11. Fox , et al: Practical approach for comparative analysis of multilesion molecular imaging using a semiautomated program for PET/CT. J Nucl Med. 2011, 52: 1727-1732. 10.2967/jnumed.111.089326.

    Article  PubMed  PubMed Central  Google Scholar 

  12. Hatt , et al: PET functional volume delineation: a robustness and repeatability study. Eur J Nucl Med Mol Imaging. 2011, 38: 663-672. 10.1007/s00259-010-1688-6.

    Article  PubMed  Google Scholar 

  13. Cheebsumon , et al: Impact of [18F]FDG PET imaging parameters on automatic tumour delineation: need for improved tumour delineation methodology. Eur J Nucl Med Mol Imaging. 2011, 38 (12): 2136-2144. 10.1007/s00259-011-1899-5. ISSN 1619-7070, 12/2011

    Article  PubMed  PubMed Central  Google Scholar 

  14. Cheebsumon NO, et al: Effects of image characteristics on performance of tumor delineation methods: a test-retest assessment. J Nucl Med. 2011, 52 (10): 1550-1558. 10.2967/jnumed.111.088914.

    Article  CAS  PubMed  Google Scholar 

  15. Boykov Yuri, et al: Interactive graph cuts for optimal boundary & region segmentation of objects in N- D images. Computer Vision. 2001, 1: 105-112.

    Google Scholar 

Pre-publication history

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Per Wollmer.

Additional information

Competing interests

The authors declare that they have no competing interests.

Authors' contributions

JL contributed to patient recruitment, did manual reading, took part in data analysis and wrote the first version of the manuscript. SG contributed to patient recruitment and did manual reading. PH contributed to the development of the semi-automatic method and to data analysis. EJ contributed to the development of the semi-automatic method and performed the programming. SV contributed to development of the semi-automatic method, patient recruitment and did manual reading. LE conceived of the study and contributed to the design of the study, to development of the semi-automatic method and to data analysis. PW contributed to the design of the study and to data analysis. All authors read and approved the final manuscript.

Authors’ original submitted files for images

Below are the links to the authors’ original submitted files for images.

Authors’ original file for figure 1

Authors’ original file for figure 2

Rights and permissions

Open Access This article is published under license to BioMed Central Ltd. This is an Open Access article is distributed under the terms of the Creative Commons Attribution License ( https://creativecommons.org/licenses/by/2.0 ), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Reprints and permissions

About this article

Cite this article

Ly, J., Garpered, S., Höglund, P. et al. Semi-automatic analysis of standard uptake values in serial PET/CT studies in patients with lung cancer and lymphoma. BMC Med Imaging 12, 6 (2012). https://doi.org/10.1186/1471-2342-12-6

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/1471-2342-12-6

Keywords