Table 4 Differences in corneal thickness measured by AOCT-1000 M and RTVue XR in 8 directions of 2-5 mm diameter area

From: Repeatability and agreement of AOCT-1000 M, RTVue XR and IOL master 500 in measuring corneal thickness mapping and axial length applying principle of optical coherence tomography

2–5 mm diameter area

Corneal thickness(mean ± SD)

Mean difference of two devices

95%CI of the intergroup difference

AOCT-1000 M

RTVue XR

S

543.81 ± 40.33

547.74 ± 41.19

-3.89 ± 10.67

-6.70, -1.09

SN

544.57 ± 38.12

547.14 ± 40.44

-2.57 ± 11.15

-5.51, 0.36

N

533.18 ± 35.84

538.16 ± 38.48

-4.97 ± 8.53

-7.21, -2.73

IN

526.17 ± 35.07

529.82 ± 37.68

-3.66 ± 10.10

-6.31, -1.00

I

521.98 ± 35.19

522.99 ± 37.81

-1.01 ± 9.45

-3.50, 1.47

IT

515.58 ± 35.25

515.66 ± 37.13

-0.07 ± 9.98

-2.70, 2.55

T

518.20 ± 36.39

518.07 ± 37.04

0.12 ± 8.00

-1.98, 2.23

ST

533.56 ± 37.66

534.87 ± 39.48

-1.31 ± 9.04

-3.69, 1.07