Skip to main content

Table 2 Quantitative Image Quality Assessment Results

From: 75% radiation dose reduction using deep learning reconstruction on low-dose chest CT

 

QLD-DLIR,

(3-mm thickness, standard kernel)

LD-IR

(3-mm thickness, standard kernel)

P-value*

QLD-DLIR

(1-mm thickness, sharp kernel)

LD-IR

(1-mm thickness, sharp kernel)

P-value*

Noise

      

Lung

11.5 ± 3.3

16.0 ± 3.8

< 0.001

37.1 ± 6.7

66.5 ± 10.9

< 0.001

Trachea

16.0 ± 3.9

13.4 ± 2.0

< 0.001

44.2 ± 7.7

57.5 ± 8.4

< 0.001

Aorta

8.6 ± 2.3

17.6 ± 2.2

< 0.001

81.8 ± 6.1

104.9 ± 7.2

< 0.001

Muscle

12.2 ± 3.3

22.8 ± 3.3

< 0.001

81.0 ± 9.5

111.8 ± 9.9

< 0.001

Axillary fat

9.2 ± 2.0

17.4 ± 3.3

< 0.001

82.1 ± 11.6

108.3 ± 16.6

< 0.001

SNR

      

Lung

84.4 ± 24.3

58.9 ± 14.1

< 0.001

23.0 ± 4.6

13.7 ± 2.5

< 0.001

Trachea

64.3 ± 14.6

75.8 ± 11.7

< 0.001

20.1 ± 3.6

16.6 ± 2.7

< 0.001

Aorta

5.6 ± 1.3

2.6 ± 0.4

< 0.001

0.58 ± 0.06

0.43 ± 0.06

< 0.001

Muscle

5.0 ± 1.4

2.5 ± 0.5

< 0.001

0.71 ± 0.12

0.49 ± 0.08

< 0.001

Axillary fat

11.3 ± 2.8

6.0 ± 1.2

< 0.001

1.2 ± 0.2

0.95 ± 0.16

< 0.001

Edge-rise-distance (mm)

1.52 ± 0.08

1.80 ± 0.11

< 0.001

1.15 ± 0.07

1.06 ± 0.08

< 0.001

  1. Note.—Data are presented as means ± standard deviations. Italicized data indicate that the values are lower (for noise and edge-rise-distance) or higher (for SNR) than the compared counterpart. DLIR = deep-learning image reconstruction, IR = iterative reconstruction, LD = low dose, QLD = quarter of the low dose, SNR = signal-to-noise ratio
  2. *P-values were calculated using the paired t-test