From: Unsupervised registration of 3D knee implant components to biplanar X-ray images
Rotation (degrees) | Translation (mm) | ||||||
---|---|---|---|---|---|---|---|
X | Y | Z | X | Y | Z | ||
Femoral components | Hybrid | 1.047 | 0.767 | 1.062 | 0.046 | 0.047 | 0.086 |
± 0.771 | ± 0.591 | ± 0.672 | ± 0.03 | ± 0.031 | ± 0.043 | ||
Potential field | 1.065 | 0.822 | 0.879 | 0.187 | 0.161 | 0.142 | |
± 0.87 | ± 0.68 | ± 0.771 | ± 0.146 | ± 0.115 | ± 0.127 | ||
Object specificity | 1.02 | 0.869 | 0.977 | 0.075 | 0.065 | 0.101 | |
± 0.767 | ± 0.674 | ± 0.722 | ± 0.049 | ± 0.052 | ± 0.07 | ||
Tibial components | Hybrid | 0.727 | 0.713 | 0.878 | 0.085 | 0.075 | 0.046 |
± 0.567 | ± 0.473 | ± 0.578 | ± 0.046 | ± 0.039 | ± 0.035 | ||
Potential field | 2.635 | 4.854 | 3.241 | 0.829 | 0.777 | 0.561 | |
± 1.9 | ± 3.492 | ± 1.765 | ± 0.368 | ± 0.67 | ± 0.554 | ||
Object specificity | 0.641 | 0.776 | 0.869 | 0.073 | 0.109 | 0.067 | |
± 0.523 | ± 0.595 | ± 0.61 | ± 0.062 | ± 0.068 | ± 0.05 | ||
All components | Hybrid | 0.887 | 0.74 | 0.97 | 0.065 | 0.061 | 0.066 |
± 0.695 | ± 0.536 | ± 0.634 | ± 0.044 | ± 0.038 | ± 0.044 | ||
Potential field | 1.85 | 2.838 | 2.06 | 0.508 | 0.469 | 0.351 | |
± 1.673 | ± 3.224 | ± 1.803 | ± 0.426 | ± 0.571 | ± 0.453 | ||
p-value\(_{Hybrid/PotentialField}\) | \(<0.0001\) | \(<0.0001\) | \(<0.0001\) | \(<0.0001\) | \(<0.0001\) | \(<0.0001\) | |
Object specificity | 0.83 | 0.822 | 0.923 | 0.074 | 0.087 | 0.084 | |
± 0.684 | ± 0.637 | ± 0.67 | ± 0.056 | ± 0.064 | ± 0.063 | ||
p-value\(_{Hybrid/ObjectSpecificity}\) | 0.5093 | 0.2225 | 0.5552 | 0.0989 | \(<0.0001\) | 0.0012 |