Skip to main content

Table 2 Exposure protocols used for assessing subjective image quality, CNR, MTF and effective dose

From: Assessment of subjective image quality, contrast to noise ratio and modulation transfer function in the middle ear using a novel full body cone beam computed tomography device

Protocol nr

1

2

3

4

5

6

7

8

9

10

Tube voltage (kVp)

80

90

100

110

120

90

90

100

110

120

Q (mAs)

450

280

200

140

100

100

450

450

300

300

Tube curr. (mA)

32

40

40

40

32

36

36

32

28

22

Exp. time (s)

14.1

7.0

5.0

3.5

3.1

2.8

12.5

14.1

10.7

13.6

Voxel (µm)

75

75

75

75

75

75

75

75

75

75

Scan angle

210º

210º

210º

210º

210º

210º

210º

210º

210º

210º

Pulsed

yes

yes

yes

yes

yes

yes

yes

yes

yes

yes

Frame number

500

500

500

500

500

500

500

501

501

502

FOV height (mm)

50

50

50

50

50

50

50

50

50

50

FOV diameter (mm)

50

50

50

50

50

50

50

50

50

50

CNR

10.5

10.3

10.1

8.6

7.9

5.3

12.4

14.2

13.5

14.3

MTF

1.49

1.54

1.60

1.60

1.62

1.51

1.50

1.41

1.55

1.57

FOM

2.1

2.0

1.8

1.4

1.2

1.5

1.8

1.6

1.6

1.3

Effective dose (µSv)

52.8

53.8

56.0

52.7

52.7

19.2

86.5

125.9

112.9

158.0

  1. CNR = contrast to noise ratio; MTF = modulation transfer function; Q = tube current exposure time product; FOV = field of view; FOM = figure of merit = CNR2/Effective dose