Table 2 Measurements obtained from the imaging parameters

From: Low-dose whole-spine imaging using slot-scan digital radiography: a phantom study

BH filter (mm)

Tube voltage (kV)

Tube current (mA)

Exposure time (ms)

ESD (µGy)

CNR

FOM

0.0

80

400

4.5

151.01

0.012476

10.3073 × 10−7

90

400

2.8

128.51

0.011805

10.8441 × 10−7

100

400

2.0

110.87

0.010008

9.0340 × 10−7

110

400

1.4

104.02

0.009008

7.8011 × 10−7

120

400

1.0

93.11

0.008111

7.0657 × 10−7

0.1

80

400

6.3

118.60

0.009911

8.2829 × 10−7

90

400

3.2

88.99

0.008880

8.8610 × 10−7

100

400

2.0

69.76

0.008497

10.3496 × 10−7

110

400

1.6

67.13

0.006588

6.4653 × 10−7

120

400

1.2

62.50

0.005658

5.1221 × 10−7

0.2

80

400

7.1

101.04

0.008222

6.6906 × 10−7

90

400

4.5

82.42

0.007534

6.8868 × 10−7

100

400

2.8

66.39

0.006834

7.0347 × 10−7

110

400

1.8

60.82

0.005984

5.8876 × 10−7

120

400

1.2

55.50

0.005572

5.5941 × 10−7

0.3

80

400

9.0

85.55

0.006754

5.3322 × 10−7

90

400

5.6

72.08

0.006356

5.6047 × 10−7

100

400

3.2

66.01

0.006141

5.7131 × 10−7

110

400

2.2

60.04

0.006026

6.0481 × 10−7

120

400

1.4

55.99

0.005711

5.8252 × 10−7

  1. The tube current is constant at 400 mA on parameters
  2. BH beam hardening, ESD entrance surface dose, CNR contrast-to-noise ratio, FOM figure of merit