Table 2 Comparison of the amount of artifacts induced by study groups in large and small FOVs

From: Comparison of the amount of artifacts induced by zirconium and titanium implants in cone-beam computed tomography images

Implant’s type

Resolution

FOV

Gray value (mean ± SD)

Mean difference ± SE

P-value*

Zirconium

High (133 µ voxel size)

Small

1566.10 ± 0.77

315.07 ± 18.11

 < 0.001

Large

1251.02 ± 62.74

Low (200 µ voxel size)

Small

1434.46 ± 0.9

105.40 ± 10.52

 < 0.001

Large

1329.05 ± 36.43

Titanium

High (133 µ voxel size)

Small

467.44 ± 143.7

145.28 ± 71.28

0.212

Large

322.16 ± 200.75

Low (200 µ voxel size)

Small

373.28 ± 117.65

60.75 ± 59.02

0.281

Large

312.53 ± 167.23

  1. *Paired t test, FOV field of view, SD standard deviation, SE standard error