Table 1 MRI sequences and parameters

From: Multimodel magnetic resonance imaging of mass-forming autoimmune pancreatitis: differential diagnosis with pancreatic ductal adenocarcinoma

 

T2WI

T1WI

Thick slab single slice MRCP

Thin slab multislice MRCP

3D VIBE

DWI

HASTE

RT-TSE

TR/TE (ms)

1600/95

3000/79

4/1.29

4500/979

1150/252

3.55/1.27

4700/66

FA (°)

150°

120°

180°

121°

13°

Section thickness (mm)

5

5

3

45

4

2.5

6

Matrix

320 × 256

320 × 256

320 × 256

512 × 320

512 × 409

320 × 256

192 × 153

FOV (cm)

32–38

32–38

32–38

30

32

32–38

32–38

Acquisition time (s)

60

150

16

5

20

18

198

  1. T2WI T2-weighted images, T1WI T1-weighted images, MRCP magnetic resonance cholangiopancreatography, 3D VIBE three-dimensional volumetric interpolated breath-hold examination, DWI diffusion-weighted imaging, HASTE half-Fourier acquisition single-shot turbo spin-echo, RT-TSE respiratory-triggered turbo spin-echo, TR repetition time, TE echo time, FA flip angle, FOV field of view