Skip to main content

Table 1 Quantitative results of 15 procedures in twelve patients, chronological order

From: Magnetic resonance imaging-guided lumbar nerve root infiltrations: optimization of an in-house protocol

No

Needle-nerve root distance (mm)

Needle angulation (°)

MRI room occupation time (min)

Nominal needle diameter (mm)

FWHM needle artifact (mm)

FWHM needle artifact ratio*

Visual artifact (mm)

Visual artifact ratio

1

2.62

25.0

70

0.9

5.16

5.73

3.66

4.07

2

3.56

27.9

55

0.9

4.94

5.49

3.06

3.40

3

0.99

22.5

60

0.9

4.49

4.99

1.35

1.50

4

1.95

22.7

65

0.9

3.98

4.42

2.68

2.98

5

3.89

16.6

55

0.9

4.73

5.26

1.49

1.66

6

2.69

13.0

54

0.9

4.73

5.26

2.34

2.60

7

1.41

8.2

50

0.9

3.14

3.49

2.99

3.32

8

4.73

21.2

64

0.9

2.57

2.86

1.54

1.71

9

3.20

6.9

49

0.9

3.55

3.94

1.71

1.90

10

0.10

24.6

40

0.9

4.84

5.38

1.83

2.03

11

1.39

33.4

40

0.9

3.75

4.17

2.11

2.34

12

1.89

14.3

39

0.9

3.04

3.38

2.12

2.36

13

1.87

27.2

41

0.9

2.81

3.12

2.02

2.24

14

0.10

12.4

44

0.9

3.05

3.39

1.49

1.66

15

0.83

11.2

40

0.9

2.81

3.12

1.42

1.58

Mean

2.08

19.1

51

 

3.8

4.27

2.12

2.36

SD

1.36

7.9

11.9

 

0.91

1.01

0.69

0.78

  1. SD standard deviation, FWHM full width at half maximum