Table 1 Characteristics of the Discovery MI and Gemini TF PET-CT systems

From: Comparison of conventional and Si-photomultiplier-based PET systems for image quality and diagnostic performance

PET-CT system

Discovery MI

Gemini TF

Crystal material

LYSO

LYSO

Amplifier

SiPM

PM-tubes

Number of rings

4

3

Size of crystals [mm3]

3.95 × 5.3 × 25

4.0 × 4.0 × 22

Axial FOV [mm]

200

180

Bore diameter [mm]

700

700

Overlap [%]

23.9

50

NEMA Sensitivity [cps/kBq]

13.8

7.0

FWHM axial @1 cm [mm]

4.7

5.8

Coincidence window width (ns)

4.9

6

Timing resolution (ps)

382

600

Lower energy threshold (keV)

425

440

Matrix

256 × 256

144 × 144

Pixel size (mm2)

2.7 × 2.7

4.0 × 4.0

Slice thickness [mm]

2.79

4.0

Image planes in the axial FOV

71

45

Number of CT slice

128

16

  1. FOV field of view; FWHM full width at half maximum; LYSO lutetium-yttrium oxyorthosilicate; NEMA National Electrical Manufacturers Association; PM photomultiplier; SiPM silicon photomultiplier