From: T1- and ECV-mapping in clinical routine at 3 T: differences between MOLLI, ShMOLLI and SASHA

Fig. 2

Percental T1 deviation of pre- and post-contrast MOLLI and ShMOLLI in myocardium and blood. T1 error in myocardium is less prominent in MOLLI (a) than ShMOLLI (b) but no significant difference could be observed between pre- and post-contrast T1 error. T1 error was comparable in pathologic (red) and remote segments (grey). T1 deviation in blood was highly dependent on T1 time and significantly different pre- and post-contrast T1 deviation were obtained with MOLLI (c) and ShMOLLI (d). Data is expressed as percental deviation of respective T1 time from SASHA in %. Mean ± SD are indicated

Back to article page