Table 2 Amount of artifacts in different bone types (n = 27)

From: Effect of exposure parameters of cone beam computed tomography on metal artifact reduction around the dental implants in various bone densities

Bone type

Exposure settings

Mean

Standard deviation

Type 1

Small FOV, 4 mA

209.83

61.23145

Small FOV, 10 mA

174.11

89.35054

Large FOV, 4 mA

382.98

239.04508

Large FOV, 10 mA

342.97

229.02314

Types 2 and 3

Small FOV, 4 mA

263.53

77.44921

Small FOV, 10 mA

283.42

101.28344

Large FOV, 4 mA

360.98

224.80996

Large FOV, 10 mA

382.60

223.14333

Type 4

Small FOV, 4 mA

412.19

77.89704

Small FOV, 10 mA

343.11

125.8897

Large FOV, 4 mA

570.24

292.o681

Large FOV, 10 mA

548.77

411.1727