Skip to main content

Table 4 Differences in mean extent values between each physician and EXINI, without and with suggested delineation of the defect made by EXINI

From: Area of ischemia assessed by physicians and software packages from myocardial perfusion scintigrams

Physician

First evaluation

P-value

Evaluation with suggestion

P-value

A

4.2

0.025

3.3

0.018

B

-0.1

0.95

0.8

0.54

C

7.4

<0.0001

3.9

0.0049

D

7.2

<0.0001

0.4

0.79

E

11.3

<0.0001

0.9

0.52

F

12.6

<0.0001

3.1

0.027

G

-2.2

0.25

0.1

0.95

H

-3.5

0.063

-3.5

0.011

I

12.6

<0.0001

4.4

0.0016

J

19.1

<0.0001

12.5

<0.0001

K

-7.0

0.0002

-3.0

0.031

  1. Mean extent value for EXINI was 17.0%.