Table 6 Estimation of the performance of the ABVS in a scenario, where second-reading of the data is performed

From: Diagnostic performance and inter-observer concordance in lesion detection with the automated breast volume scanner (ABVS)

 

At least one examiners judges ABVS as normal (negative)

Both examiners concordantly suspicious ABVS (positive)

Total

Benign cases (conventional ultrasound negative)

64a

18b

82

Malignant cases (conventional ultrasound positive)

3b

15a

18

Total

67

33

100

  1. Sensitivity, specificity and accuracy calculate to 83.3%, 78.0% and 79.0%, respectively. For the evaluation of the second-reading the following rules were applied: If both examiners agreed that a scan was suspicious, the evaluation was considered “positive”. If both examiners agreed that a scan was unsuspicious or disagreed and only one examiner regarded the scan as unsuspicious, the evaluation was considered “negative”.
  2. a“true” evaluation.
  3. b“false” evaluation.