Skip to main content

Table 3 Concordance between examiner 1 and examiner 2 concerning the distinct BI-RADS® category in the ABVS examination

From: Diagnostic performance and inter-observer concordance in lesion detection with the automated breast volume scanner (ABVS)

  

BI-RADS® ABVS

BI-RADS® ABVS

BI-RADS® ABVS

BI-RADS® ABVS

Total

1

2

0, 3 or 4

5

(No finding)

(Benign finding)

(Unclear finding)

(Malignant finding)

Examiner 2 (AF)

BI-RADS® ABVS

20a

2

4

2

28

1

(No Finding)

BI-RADS® ABVS

4

8a

2

0

14

2

(Benign finding)

BI-RADS® ABVS

14

7

13a

9

43

0, 3 or 4

(Unclear finding)

BI-RADS® ABVS

4

0

5

6a

15

5

(Malignant finding)

 

Total

42

17

24

17

100

  1. Distribution of cases by rater and by detailed BI-RADS® ABVS category. Concerning this detailed evaluation, the inter-rater reliability coefficient calculates to κ = 0.27 (0.14-0.40).
  2. aconcordance between examiner 1 and examiner 2.