Skip to main content

Table 3 Difference in interobserver mean area calculation

From: Reproducibility of Corneal Graft Thickness measurements with COLGATE in patients who have undergone DSAEK (Descemet Stripping Automated Endothelial Keratoplasty)

    

Bias (95% CI)

 

Mean area (um2) ±SD

P value

R value ICC value

LoA Mean

Upper limit

Lower limit

M1

151619.4 ± 40912.0

0.164

0.913 0.911 (95% CI 0.849-0.949)

3364.4 (-1423.0 to 8151.7)

36381.0 (28144.5 to 44617.4)

−29652.2 (-37888.7 to -21415.7)

B1

148255.0 ± 39651.0

M2

151843.2 ± 39172.1

0.940

0.969 0.968 (95% CI 0.945-0.982)

110.6 (-2802.0 to 3023.2)

20197.8 (15186.7 to 25208.8)

−19976.5 (-24987.6 to-14965.5)

B2

151732.6 ± 41381.0