Skip to main content

Table 2 Repeatability of ASOCT graft measurements

From: Correlation of anterior segment optical coherence tomography measurements with graft trephine diameter following descemet stripping automated endothelial keratoplasty

Parameter

Mean (SD)

Mean (SD)

P*

Mean difference (95%CI)

Limits of Agreement

Coefficient of variation (95% CI)

Intraclass Coefficient (95% CI)+

 

Observer 1

Observer 2

Posterior Cornea Arc length (mm)

13.08 (±0.61)

13.06 (±0.59)

0.634

0.02 (−0.06 to 0.10)

−0.32; 0.35

2.56% (1.50, 3.63)

0.959 (0.899 to 0.983)

Anterior chamber Width (mm)

11.24 (±0.50)

11.24 (±0.45)

0.942

0.00 (−0.10 to 0.10)

−0.42; 0.41

3.71% (0.25, 5.25)

0.900 (0.765 to 0.959)

Ant Graft Arc length (mm)

9.77 (±0.51)

9.783 (±0.53)

0.317

−0.02 (−0.06 to 0.02)

−0.18; 0.14

1.61% (0.94, 2.28)

0.988 (0.970 to 0.995)

Ant Graft Cord length (mm)

8.70 (±0.33)

8.73 (±0.34)

0.075

−0.02 (−0.05 to 0.00)

−0.14; 0.09

1.28% (0.75, 1.81)

0.985 (0.963 to 0.994)

Post Graft Arc length (mm)

9.43 (±0.52)

9.45 (±0.51)

0.240

−0.02 (−0.06 to 0.02)

−0.17; 0.13

1.55% (0.91, 2.19)

0.989 (0.973 to 0.996)

Post Graft Cord length (mm)

8.24 (±0.32)

8.26 (±0.34)

0.088

−0.02 (−0.05 to 0.00)

−0.12; 0.08

1.24% (0.73, 1.76)

0.988 (0.969 to 0.995)

  1. *(2-sided).
  2. + all p< 0.001.