Table 3 Reproducibility of ADC Values in repeated MR exams

From: Reproducibility and feasibility of optic nerve diffusion MRI techniques: single-shot echo-planar imaging (EPI), readout-segmented EPI, and reduced field-of-view diffusion-weighted imaging

 

rs-EPI

rFOV-DWI

ss-EPI

ROI1

   

 ICCs

0.916

0.850

0.810

 BA-LA

− 20 to 22

− 18 to 28.1

− 33.5 to 22.2

ROI2

   

 ICCs

0.797

0.595

0.442a

 BA-LA

− 31 to 32.2

− 26.5 to 29

− 46.2 to 51.1

ROI3

   

 ICCs

0.781

0.750

0.379a

 BA-LA

− 40.6 to 18.1

− 19.6 to 25.3

− 41.5 to 41.3

  1. BA-LA (in %), the 95% Bland–Altman limits of agreements; ICCs, intraclass correlation coefficients (two-way random); rs-EPI DWI, readout-segmented echo-planar imaging diffusion-weighted imaging; rFOV, reduced field-of-view; ss-EPI, single-shot echo-planar imaging; ROI, region of interest
  2. aP > 0.05