Table 4 Per-vessel diagnostic performance of CCTA and CCTA plus resting DE-CTP for detection of flow-limiting stenoses

From: Clinical value of resting cardiac dual-energy CT in patients suspected of coronary artery disease

 

CCTA

CCTA plus resting DE-CTP

Accuracy

76.16 (125/162) [69.92–83.38]

95.68 (155/162) [89.72–97.43]

Sensitivity

100.00 (77/77) [95.32–100.00]

96.10 (74/77) [89.03–99.19]

Specificity

56.47 (48/85) [45.28–67.20]

95.29 (81/85) [88.39–98.70]

PPV

67.54 (77/114) [58.14–76.01]

94.87 (74/78) [87.39–98.59]

NPV

100.00 (48/48) [92.60–100.00]

96.43 (81/84) [89.92–99.26]

AUC

0.777 [0.705–0.839]

0.956 [0.912–0.982]

  1. Data are % (raw data) [95% confidence interval] for accuracy, sensitivity, specificity, PPV, and NPV. Data are value [95% confidence interval] for AUC
  2. CCTA = coronary CT angiography; DE-CTP = dual-energy CT perfusion; PPV = positive predictive value; NPV = negative predictive value; AUC = area under the receiver operating characteristic curve