Table 3 Per-patient diagnostic performance of CCTA and CCTA plus resting DE-CTP for detection of flow-limiting stenoses

From: Clinical value of resting cardiac dual-energy CT in patients suspected of coronary artery disease

 

CCTA

CCTA plus resting DE-CTP

Accuracy

83.33 (45/54) [70.71–92.08]

96.30 (52/54) [87.25–99.55]

Sensitivity

100.00 (37/37) [90.51–100.00]

97.30 (36/37) [85.84–99.93]

Specificity

47.06 (8/17) [22.98–72.19]

94.12 (16/17) [71.31–99.85]

PPV

80.43 (37/46) [66.09–90.64]

97.30 (36/37) [85.84–99.93]

NPV

100 .00 (8/8) [63.06–100.00]

94.12 (16/17) [70.28–99.88]

AUC

0.735 [0.598–0.846]

0.957 [0.864–0.939]

  1. Data are % (raw data) [95% confidence interval] for accuracy, sensitivity, specificity, PPV, and NPV; Data are value [95% confidence interval] for AUC
  2. CCTA = coronary CT angiography; DE-CTP = dual-energy CT perfusion; PPV = positive predictive value; NPV = negative predictive value; AUC = area under the receiver operating characteristic curve