Table 3 Comparison of PET-dependent parameters between NIM PET/CT phantom and NEMA IEC Body phantom

From: A NIM PET/CT phantom for evaluating the PET image quality of micro-lesions and the performance parameters of CT

 

Sphere size (mm)

Contrast (%)

Background variability (%)

Residual error (%)

NIM PET/CT phantom

37

82.11

2.05

8.98

28

75.81

2.67

22

73.11

3.27

17

71.43

3.86

13

61.09

4.64

10

45.08

5.46

7

20.52

6.45

4

NEMA IEC Body phantom

37

82.21

1.99

9.10

28

74.84

2.53

22

72.38

3.04

17

69.01

3.59

13

58.41

4.32

10

42.52

5.33

  1. Since the device could not detect 4 mm spheres, no measurement data of those spheres in the measurement module could be achieved. The maximum deviations are highlighted in bold