Table 1 Tabulated values for both scanning methods for the scan time, number, location, and size

From: The concordance in lesion detection and characteristics between the Anatomical Intelligence and conventional breast ultrasound Scan method

  C-US A-US ICC (C-A) ICC A1-2
Reader 1 Reader 2
Scan time     0.7642 0.7971
 Nodule < 5 3.99 4.56 4.21 0.8995 0.9549
 Nodule5-10 6.86 6.25 6.33 0.7417 0.7996
 Nodule > 10 10.91 9.26 9.63 0.6664 0.6651
Number 73 73 73 0.7692 0.783
  < 5 22 19 20 0.8015 0.7982
 6–10 36 33 34 0.7652 0.7825
  > 10 15 21 19 0.6913 0.8171
Location (Clockface location; Distance to the nipple)      
 Clockface location     0.8651 0.9283
 Distance to the nipple     0.8436 0.8748
Size (Largest diameter; Mean diameter)      
Largest diameter 6.7092 5.9415 6.2764 0.7502 0.7248
 1–5 mm     0.4241 0.4531
 5–10 mm     0.6538 0.5912
  > 10 mm     0.9614 0.9438
Mean diameter 5.7092 5.9477 6.2437 0.8885 0.8163
 1–5 mm     0.7241 0.7123
 5–10 mm     0.6538 0.6759
  > 10 mm     0.9614 0.9187
  1. Data in bold are almost perfect agreement
  2. CUS represents the conventional US scan method; AIUS represents the AI breast US scan method; C-A represents the ICC between the AI breast US scan method and AI breast US scan method; A1-2 represents the ICC between the two reader of AI breast US scan method