From: Performance of TGSE BLADE DWI compared with RESOLVE DWI in the diagnosis of cholesteatoma
Parameters | TGSE | RESOLVE | P value |
---|---|---|---|
Qualitative parameters | |||
 1.Imaging sharpness | 2.85 ± 0.35 | 2.73 ± 0.49 | 0.23 |
 2.Geometric distortion | 3.97 ± 0.15(whole) 3.95 ± 0.21(lesion) | 3.26 ± 0.76(whole) 3.64 ± 0.48(lesion) | < 0.01 < 0.01 |
 3.Ghosting artifacts | 3.92 ± 0.26 | 3.02 ± 0.55 | < 0.01 |
 4.Overall imaging quality | 4.85 ± 0.35 | 4.16 ± 0.69 | < 0.01 |
Quantitative parameters | |||
 1.SNR (b = 1000s/mm2) | |||
  Lesion | 102.3 ± 32.2 | 493.7 ± 241.6 | < 0.01 |
  Brainstem | 59.1 ± 15.5 | 289.8 ± 140.9 | < 0.01 |
 2.Contrast | 1.79 ± 0.35 | 1.62 ± 0.44 | =0.005 |
 3.CNR (b = 1000s/mm2) | 4.33 ± 2.8 | 2.7 ± 2.6 | < 0.01 |