Skip to main content

Table 2 Systematic T1 error in inversion-recovery methods MOLLI/ShMOLLI compared to SASHA

From: T1- and ECV-mapping in clinical routine at 3 T: differences between MOLLI, ShMOLLI and SASHA

 

native

post-contrast

p-value

Myocardial T1 error (% lower than SASHA)

 MOLLI

19.5 ± 4.4

16.8 ± 4.6

<  0.05

 ShMOLLI

26.3 ± 6.3

24.8 ± 3.7

<  0.05

Bloodpool T1 error (% lower than SASHA)

 MOLLI

13.3 ± 4.8

5.4 ± 6.1

<  0.05

 ShMOLLI

18.7 ± 3.1

11.5 ± 5.1

<  0.05

  1. Assessed errors in myocardial tissue are comparable between native and post-contrast mapping from MOLLI and ShMOLLI, independently from respective T1 times. Blood T1 error exhibits significant dependence on respectively measured T1 time. Data is expressed mean percental deviation ± SD from T1 time measured with SASHA. p < 0.05 is considered statistically significant