From: Accurate hybrid template–based and MR-based attenuation correction using UTE images for simultaneous PET/MR brain imaging applications

Fig. 2

Attenuation maps generation workflow. Both UTE images are rigidly registered to the UTEtemplate. Segmentation into three tissue classes (soft tissue, air, bone) is performed using first echo UTE image registered to the UTEtemplate. sUTE maps are then generated by assigning attenuation coefficients to all classes with respect to the masks, 3D 2-mm Gaussian smoothing and warping back to the subject space. Attenuation coefficients are fixed for soft tissue and air (0.1 cm− 1 and 0 cm− 1, respectively), whereas for bones can be either fixed – equal 0.151 cm− 1(sUTEfix) or continuous – based on R2* map (sUTEcont)