Table 1 3DE volumetric measurements and 2DE calculated volumes of left atrial human cadaveric casts in comparison to their true volumes

From: Validation of 3D echocardiographic volume detection of left atrium by human cadaveric casts

  

3DE volume: GE 4D LVQ

3DE volume: TomTec LV

2DE volume

cast no.

true volume (ml)

measured volume (ml)

difference to true volume (ml)

measured volume (ml)

difference to true volume (ml)

calculated volume (ml)

difference to true volume (ml)

1

38

36

−2

36

-2

35

−3

2

94

94

0

103

9

106

12

3

51

43

−8

50

−1

55

4

4

43

43

0

42

− 1

36

−7

5

44

45

1

50

6

40

−4

6

58

68

10

70

12

68

10

7

68

62

−6

68

0

76

8

mean (95% CI)

56.6

55.9

−0.7 (−6.1–4.6)

59.9

3.3 (−1.9–8.5)

59.5

2.9 (−4.0–9.9)

SD

19.4

20.3

5.8

22.8

5.6

26.0

7.5

range

38–94

36–94

−8–10

36–103

−2–12

35–106

−7–12

LOA

  

−12.1–10.6

 

−7.8–14.4

 

−11.9–17.6

r

 

0.959 (p = 0.001)

0.977 (p = 0.0002)

0.988 (p = 0.00003)

  1. 95% CI – 95% confidence interval, 2DE volume – volume calculated by 2D echocardiographic biplane area-length method, 3DE volume: GE 4D LVQ and 3DE volume: TomTec LV – 3D echocardiographic volume measured by GE 4D LVQ and TomTec 4D LV-Function softwares, respectively, LOA – limits of agreement defined as mean ± 1.96 SD, r – Pearson correlation coefficient, SD – standard deviation