Skip to main content

Table 2 Clinical settings of abdominal CBCT acquisitions for patients undergoing TACE

From: Investigations of organ and effective doses of abdominal cone-beam computed tomography during transarterial chemoembolization using Monte Carlo simulation

 

Configuration values and acquisition parameters

Acquisition mode

CB-CTAP

Acquisition rate (fps)

30

Radiation time (sec)

12

Rotation speed

20°/sec

Number of frames

Approximate 315 frames

Acquisition range

215°

Focus-to-image distance (FID) (cm)

120

Field of view size (inch)

17

Focus

Large focus

Total filtration (mm)

11.2-mm aluminum equivalent

kV

100

mA

360

Pulse width (msec)

5.6