Table 1 Reproducibility (RMSCV%) calculated from duplicate HRpQCT measurements at the uninstrumented and instrumented radii

From: Distal radius plate of CFR-PEEK has minimal effect compared to titanium plates on bone parameters in high-resolution peripheral quantitative computed tomography: a pilot study

Bone parameter

RMSCV% Radius #1

RMSCV% Radius #2

RMSCV% Radius #3

Un-instrumented

Instrumented (CFR-PEEK)

Un-instrumented

Instrumented (Titanium)

Un-instrumented

Instrumented (Titanium)

vBMD

ā€ƒDtot

0.52

2.18

0.02

1.29

0.00

0.20

ā€ƒDtrab

0.10

1.95

0.34

0.00

0.31

0.62

ā€ƒDcort

0.31

1.48

0.24

1.13

0.01

0.06

Microā€‰āˆ’ā€‰architecture

ā€ƒBV/TV

0.00

1.76

0.47

0.00

1.25

0.00

ā€ƒTb.N

2.24

0.92

0.00

1.19

0.72

0.67

ā€ƒTb.Th

1.84

0.90

0.00

1.66

0.00

1.11

ā€ƒTb.Sp

2.16

0.99

0.16

1.13

0.37

0.08

ā€ƒCt.Th

1.50

1.55

0.00

1.65

0.93

0.00

Biomechanical

ā€ƒScomp

0.42

1.38

0.28

1.71

0.02

0.49

ā€ƒF.Ult

0.28

1.30

0.40

2.23

0.06

0.54

  1. Abbreviations: vBMD volumetric bone mineral density, Dtot total density, Dtrab trabecular density, Dcort cortical density, BV/TV bone to total volume, Tb.N trabecular number, Tb.Th trabecular thickness, Tb.Sp trabecular separation, Ct.Th cortical thickness, Scomp compression stiffness, F.Ult ultimate failure load