Table 2 Artifact size of the solely scanned screws

From: Biodegradable magnesium Herbert screw – image quality and artifacts with radiography, CT and MRI

Imaging modality

Mg: mean artifact size [mm]

Ti: mean artifact size [mm]

Mean difference in artifact size [mm]

T-value

P-value

DX

4.5 ± 0.3

5.2 ± 0.2

0.7 ± 0.2

4.1

<0.005

MDCT

5.7 ± 0.4

11.9 ± 0.5

6.2 ± 0.2

27.0

<0.001

FPCT

5.5 ± 0.2

11.4 ± 0.5

5.9 ± 0.2

31.8

<0.001

3 T MRI

 PDw TSE

5.2 ± 0.7

7.9 ± 0.5

2.6 ± 0.3

9.1

<0.001

 T1w TSE

8.4 ± 0.7

12.9 ± 1.0

4.5 ± 0.4

10.6

<0.001

 PDw TSE WARP

5.0 ± 0.7

6.2 ± 0.5

1.1 ± 0.3

3.6

<0.005