Skip to main content

Table 2 Typical MRI sequence parameters

From: Validation of T1 and T2 algorithms for quantitative MRI: performance by a vendor-independent software

 

TE (ms)

TR (ms)/delay between contrast preparations pulses

FA (°)

FOV (mm)

Matrix

Preparation pulse delays (ms)

iPAT/SENSE factor

Receiver BW (kHz)

ACQ time (hh:mm:ss)

T1 spin-echo IR (magnitude images)

5.8

10,000/10,000

90

241 × 241

128 × 128

[21, 60, 100, 200, 300, 500, 660, 900, 1050, 1300, 1600, 2000, 2250, 2500, 3000, 3500, 4300]

off

64

06:45:20

T1 PSIR

1.11

2.4/40,000

35

360 × 270

192 × 144

[150, 300, 400, 500, 700, 900, 1100, 1300, 1800, 2000, 2500, 3000, 3700, 4300, 5000, 6300] + 1 reference image without IR-preparation

2

208.32

00:11:20

T1 saturation recovery balanced SSFP

1.11

2.4/40,000

35

360 × 270

192 × 144

[150, 300, 400, 500, 700, 900, 1100, 1300, 1800, 2000, 2500, 3000, 3700, 4300, 5000, 5600, 6300, 8000] + 1 reference image without SR preparation

2

208.32

00:12:40

T1 MOLLI/T1 Look-Locker correction (magnitude and PSIR images)

1.1

2.4/7920

35

360 × 270

192 × 144

[130, 210, 1130, 1210, 2130, 2210, 3130, 4130]

2

208.32

00:00:11

T2 spin echo (single-echo)

[6, 12, 20, 30, 40, 50, 70, 90, 120, 140, 180, 300, 400, 600, 1000]

10,000/10,000

90

241 × 241

128 × 128

n/a

off

64

05:20:00

T2 spin echo (multi-echo)

[9, 18, 27, 36, 45, 54, 63, 72, 81, 90, 99, 108, 117, 126, 135, 144]

1500/n/a

90

160 × 160

256 × 256

n/a

off

62.5

00:06:24

T2-prepared balanced SSFP

1.11

2.4/20,000

35

360 × 270

192 × 144

T2p range = 25 − 200; ΔT2p = 5;

2

208.32

00:12:20

  1. IR inversion recovery, PSIR phase sensitive inversion recovery, MOLLI modified Look-Locker inversion recovery, SSFP steady-state free precession