Skip to main content

Table 1 Ocelot and Dragonfly OCT system technical specifications

From: Diagnostic imaging capabilities of the Ocelot -Optical Coherence Tomography System, ex-vivo evaluation and clinical relevance

Technical specifications

Dragonfly system

Ocelot system

Laser

Swept Source

Swept Source

Domain

Frequency

Frequency

Interferometer

Differential Path

Common Path

Imaging Speed

100 Hz

1.0 Hz

Laser Scan Rate

50,000 axial lines/s

20,000 axial lines/s

Sensitivity

>90 dB

>90 dB

Dynamic Range

>50 dB

>50 dB

Imaging Range

≤5 mm

≤3.3 mm

Resolution

<20 μm (axial) 25-60 μm (lateral)

<20 μm (axial) < 300 μm (lateral)

Tissue Penetration

1–2 mm

1–2 mm