Table 4 Comparison of literature reports and present study for scattered dose reduction with the use of gonadal shields during abdominal CT scans (with the testes not in the direct beam)

From: Radiation dose reduction at a price: the effectiveness of a male gonadal shield during helical CT scans

Study

Study Type

Scanning Protocol

Testicular Dose (mGy)

% Reduction

   

Without Shield

With Shield

 

Hidajat et al. 1996 [19]

Phantom, 1 mm capsule

Single-slice sequential, 10 mm slice, 250 mAs/slice, 120 kV

1.46

0.07

95

Price et al. 1995 [18]

Phantom, 1 mm wrap-around lead rubber

Single-slice spiral, 10 mm slice, 220 mA, 120 kV

0.821

0.191

77

Hohl et al. 2005 [7]

Patient, 1 mm capsule

16-slice spiral, 16 × 1.5 mm collimation,150 mAseff,120 kV

2.40

0.32

87

Present study

Phantom, 1 mm wrap-around canvas

16-slice spiral, 16 × 1.25 mm collimation 194 mAseff, 120 kV

0.62

0.26

58

  1. 1Values initially reported in mSv and were converted to mGy for comparison, assuming 1 mSv = 1 mGy.