Skip to main content

Table 1 Pulse sequence parameters used in the study

From: Quantitative and qualitative assessment of structural magnetic resonance imaging data in a two-center study

Dataset

Protocol name

Protocol type

TR (ms)

TE (ms)

Flip angle

No. Slices

Scan time (sec)

in-plane resolution (mm2)

Phase encoding direction

Slice thickness (mm)

Slice gap (mm)

FCa

SBD

Initial comparing T1-weighted protocols

A

FLASH

25

4.6

30

160

408

1.0x1.0

Right-left

2.0

−1

-

-

 

B

MPRAGE

9.8

4.6

8

120

279

1.16x1.1

Right-left

1.2

0

-

-

 

C

MPRAGE

7.6

3.5

8

160

614

1.0x1.0

Anterior-posterior

1.0

0

-

-

 

D

MPRAGE

7.6

3.5

8

160

614

1.0x1.0

Anterior-posterior

1.05

0

-

-

 

E

MPRAGE

9.8

4.6

8

120

279

1.16x1.1

Right-left

1.2

0

-

-

 

F

MPRAGE

7.1

3.3

8

145

557

1.0x1.0

Anterior-posterior

1.1

0

-

-

Optimizing T1-weighted protocols

F

MPRAGE

7.1

3.3

8

145

557

1.0x1.0

Anterior-posterior

1.1

0

-

-

 

F1

MPRAGE

7.1

3.3

8

145

557

1.0x1.0

Anterior-posterior

1.1

0

-

✓

 

F2

MPRAGE

7.1

3.3

8

145

557

1.0x1.0

Right-left

1.1

0

-

-

 

F3

MPRAGE

7.1

3.3

8

145

557

1.0x1.0

Right-left

1.1

0

-

✓

 

C1

MPRAGE

9.5

5.3

8

160

614

1.0x1.0

Anterior-posterior

1.0

0

✓

-

 

C2

MPRAGE

7.6

3.5

8

160

614

1.0x1.0

Anterior-posterior

1.0

0

-

✓

 

C3

MPRAGE

10

5.6

8

160

614

1.0x1.0

Right-left

1.0

0

✓

-

  1. FCa: Flow Compensation, manufacturer's default settings.
  2. SBD: Saturation band, manufacturer's default settings.