Figure 4From: Low tube voltage CT for improved detection of pancreatic cancer: detection threshold for small, simulated lesions The radiation dose. The radiation dose measured as the CTDIvol at 80 kV and 120 kV is shown for different tube currents. From the regression equations one can see that the tube current must be increased from 160 mA to 500 mA when changing the tube voltage from 120 kV to 80 kV to attain the same CTDIvol of 15 mGy.Back to article page